Fingerstick glucose determination in shock.

S H Atkin1, A Dasmahapatra, M A Jaker, M I Chorost, S Reddy.   

Abstract

OBJECTIVE: To evaluate the accuracy of fingerstick glucose measurements in severely hypotensive patients.
DESIGN: Prospective, nonrandomized comparison study.
SETTING: Emergency department in a university hospital. PATIENTS: Twenty-five severely hypotensive patients (systolic blood pressure less than or equal to 80 mm Hg) and 39 normotensive patients. MEASUREMENTS: Simultaneous fingerstick reagent strip glucose, venous reagent strip glucose, and laboratory glucose values were compared. Data were analyzed using error-grid analysis.
RESULTS: In hypotensive patients, the mean fingerstick glucose values were significantly lower than the values obtained either by venous reagent strip or laboratory glucose measurements. Fingerstick glucose values in the hypotensive group were 67.5% of laboratory glucose values, and were significantly lower than the values obtained in the normotensive group (91.8%, P less than 0.001). Only 36% of the hypotensive patients had fingerstick glucose values within the acceptable range of 20% of the laboratory glucose value. Thirty-two percent of hypotensive patients were incorrectly diagnosed as hypoglycemic (glucose less than 3.89 mmol/L [70 mg/dL]); 2 of these patients were actually hyperglycemic (glucose greater than 11.10 mmol/L [200 mg/dL]). Venous reagent strip measurements accurately reflected laboratory glucose values in both hypotensive and normotensive patients.
CONCLUSIONS: Fingerstick glucose testing does not accurately represent venous glucose levels in severely hypotensive patients. If fingerstick glucose testing is relied on for these patients, errors in clinical management may be made. Venous reagent strip glucose testing correlates well with laboratory glucose measurements and should be the preferred method for rapid assessment of glucose level in critically ill patients with severe hypotension.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 2029097     DOI: 10.7326/0003-4819-114-12-1020

Source DB:  PubMed          Journal:  Ann Intern Med        ISSN: 0003-4819            Impact factor:   25.391


  24 in total

1.  Capillary versus venous bedside blood glucose estimations.

Authors:  R Boyd; B Leigh; P Stuart
Journal:  Emerg Med J       Date:  2005-03       Impact factor: 2.740

Review 2.  Accuracy of point-of-care glucose measurements.

Authors:  Annette Rebel; Mark A Rice; Brenda G Fahy
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

3.  Analysis: New point-of-care blood glucose monitoring system for the hospital demonstrates satisfactory analytical accuracy using blood from critically ill patients--an important step toward improved blood glucose control in the hospital.

Authors:  Jeffrey I Joseph
Journal:  J Diabetes Sci Technol       Date:  2013-09-01

4.  Galactokinase deficiency in a patient with congenital hyperinsulinism.

Authors:  Mashbat Bayarchimeg; Dunia Ismail; Amanda Lam; Derek Burk; Jeremy Kirk; Wolfgang Hogler; Sarah E Flanagan; Sian Ellard; Khalid Hussain
Journal:  JIMD Rep       Date:  2011-12-13

5.  The need for continuous blood glucose monitoring in the intensive care unit.

Authors:  Ram Weiss; Isaac Lazar
Journal:  J Diabetes Sci Technol       Date:  2007-05

6.  Accuracy of point-of-care blood glucose measurements in critically ill patients in shock.

Authors:  Carlo Jan Pati-An Garingarao; Myrna Buenaluz-Sedurante; Cecilia Alegado Jimeno
Journal:  J Diabetes Sci Technol       Date:  2014-06-08

Review 7.  Challenges to glycemic measurement in the perioperative and critically ill patient: a review.

Authors:  Andrew D Pitkin; Mark J Rice
Journal:  J Diabetes Sci Technol       Date:  2009-11-01

8.  Accuracy of bedside capillary blood glucose measurements in critically ill patients.

Authors:  C Dana Critchell; Vincent Savarese; Amy Callahan; Christine Aboud; Serge Jabbour; Paul Marik
Journal:  Intensive Care Med       Date:  2007-09-01       Impact factor: 17.440

9.  Failure of Glucose Monitoring in an Individual with Pseudohypoglycemia.

Authors:  Kimberley T Lee; Peter M Abadir
Journal:  J Am Geriatr Soc       Date:  2015-08       Impact factor: 5.562

10.  Evaluation of the VIA Blood Chemistry Monitor for Glucose in Healthy and Diabetic Volunteers.

Authors:  Arjunan Ganesh; Brian Hipszer; Navdeep Loomba; Barbara Simon; Marc C Torjman; Jeffrey Joseph
Journal:  J Diabetes Sci Technol       Date:  2008-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.