Literature DB >> 20187699

Effects of magnetic resonance imaging in patients with implanted deep brain stimulation systems.

Valerie Fraix1, Stephan Chabardes, Alexandre Krainik, Eric Seigneuret, Sylvie Grand, Jean-François Le Bas, Paul Krack, Alim-Louis Benabid, Pierre Pollak.   

Abstract

OBJECT: The aim of this study was to study the effects of MR imaging on the electrical settings of deep brain stimulation (DBS) systems and their clinical consequences.
METHODS: The authors studied the effects of 1.5-T MR imaging on the electrical settings of implanted DBS systems, including 1 or more monopolar or quadripolar leads, extension leads, and single- or dual-channel implantable pulse generators (IPGs). The IPG was switched off during the procedure and the voltage was set to 0. The impedances were checked before and after MR imaging.
RESULTS: Five hundred seventy patients were treated with DBS for movement disorders and underwent brain MR imaging after lead implantation and before IPG implantation. None of the patients experienced any adverse events. Thirty-one of these patients underwent 61 additional MR imaging sessions after the entire DBS system had been implanted. The authors report neither local cutaneous nor neurological disorders during or after the MR imaging session. No change in the IPG settings occurred when the magnet reed switch function remained disabled during the procedure.
CONCLUSIONS: This study demonstrates that 1.5-T MR imaging can be performed safely with continuous monitoring in patients with a DBS system. The ability to disable the magnet reed switch function of the IPG prevents any change in the electrical settings and thus any side effects. The increasing number of DBS indications and the widespread use of MR imaging indicates the need for defining safety guidelines for the use of MR imaging in patients with implanted neurostimulators.

Entities:  

Mesh:

Year:  2010        PMID: 20187699     DOI: 10.3171/2010.1.JNS09951

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  11 in total

Review 1.  High frequency deep brain stimulation of the subthalamic nucleus versus continuous subcutaneous apomorphine infusion therapy: a review.

Authors:  R Carron; V Fraix; C Maineri; E Seigneuret; B Piallat; P Krack; P Pollak; A L Benabid; Stéphan Chabardès
Journal:  J Neural Transm (Vienna)       Date:  2010-12-29       Impact factor: 3.575

Review 2.  Role of radiology in central nervous system stimulation.

Authors:  D P Minks; E A C Pereira; V E L Young; K M Hogarth; G Quaghebeur
Journal:  Br J Radiol       Date:  2015-02-25       Impact factor: 3.039

3.  Optimized, Minimal Specific Absorption Rate MRI for High-Resolution Imaging in Patients with Implanted Deep Brain Stimulation Electrodes.

Authors:  A M Franceschi; G C Wiggins; A Y Mogilner; T Shepherd; S Chung; Y W Lui
Journal:  AJNR Am J Neuroradiol       Date:  2016-07-14       Impact factor: 3.825

Review 4.  Improving Safety of MRI in Patients with Deep Brain Stimulation Devices.

Authors:  Alexandre Boutet; Clement T Chow; Keshav Narang; Gavin J B Elias; Clemens Neudorfer; Jürgen Germann; Manish Ranjan; Aaron Loh; Alastair J Martin; Walter Kucharczyk; Christopher J Steele; Ileana Hancu; Ali R Rezai; Andres M Lozano
Journal:  Radiology       Date:  2020-06-23       Impact factor: 11.105

Review 5.  [Neurological and technical aspects of deep brain stimulation].

Authors:  J Voges; J K Krauss
Journal:  Nervenarzt       Date:  2010-06       Impact factor: 1.214

6.  Measurements of RF heating during 3.0-T MRI of a pig implanted with deep brain stimulator.

Authors:  Krzysztof R Gorny; Michael F Presti; Stephan J Goerss; Sun C Hwang; Dong-Pyo Jang; Inyong Kim; Hoon-Ki Min; Yunhong Shu; Christopher P Favazza; Kendall H Lee; Matt A Bernstein
Journal:  Magn Reson Imaging       Date:  2012-12-07       Impact factor: 2.546

Review 7.  Deep brain stimulation and electromagnetic interference.

Authors:  Shervin Rahimpour; Musa Kiyani; Sarah E Hodges; Dennis A Turner
Journal:  Clin Neurol Neurosurg       Date:  2021-02-25       Impact factor: 1.876

8.  A novel brain stimulation technology provides compatibility with MRI.

Authors:  Peter Serano; Leonardo M Angelone; Husam Katnani; Emad Eskandar; Giorgio Bonmassar
Journal:  Sci Rep       Date:  2015-04-29       Impact factor: 4.379

9.  Pathological alterations and stress responses near DBS electrodes after MRI scans at 7.0T, 3.0T and 1.5T: an in vivo comparative study.

Authors:  Lin Shi; An-Chao Yang; Da-Wei Meng; Shao-Wu Li; Huan-Guang Liu; Jun-Ju Li; Xiu Wang; Xin Zhang; Jian-Guo Zhang
Journal:  PLoS One       Date:  2014-07-02       Impact factor: 3.240

10.  White Matter Changes Along the Electrode Lead in Patients Treated With Deep Brain Stimulation.

Authors:  Richard Erasmi; Oliver Granert; Dmitry Zorenkov; Daniela Falk; Fritz Wodarg; Günther Deuschl; Karsten Witt
Journal:  Front Neurol       Date:  2018-11-21       Impact factor: 4.003