Literature DB >> 20182845

Surface dose and build-up region measurements with wedge filters for 6 and 18 MV photon beams.

Hatice Bilge1, Nurdan Ozbek, Murat Okutan, Aydin Cakir, Hilal Acar.   

Abstract

PURPOSE: High-energy photons are most commonly used in radiotherapy to treat cancer. Wedge filters are required to obtain homogeneous dose distribution in the patient. Different wedge filter types create different surface doses. In this study, the effect of the virtual and physical wedge filters on the surface and build-up region doses was examined for 6- and 18-MV high-energy photon beams.
MATERIALS AND METHODS: The measurements were made in a water equivalent phantom in the build-up region at a 100-cm source-to-surface distance for various field sizes using virtual and physical wedge filters having different angles. A parallel-plate ion chamber was used to measure the percent depth doses.
RESULTS: The percentage dose at the surface increased as the field size increased for open, virtual, and physical wedged beams. For open, physical, and virtual wedged beams, the surface doses were found to be 15.4%, 9.9%, and 15.9% with 6-MV photons and 10.6%, 8.8%, 11.9% with 18-MV photons, respectively, at 10 x 10 cm(2) field size.
CONCLUSION: Build-up doses of virtual wedged beams were similar to those of open beams. Surface and buildup doses of physical wedged beams were lower than those of open and virtual wedged beams.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20182845     DOI: 10.1007/s11604-009-0393-5

Source DB:  PubMed          Journal:  Jpn J Radiol        ISSN: 1867-1071            Impact factor:   2.374


  19 in total

1.  Dose measurements in the build-up region for the photon beams from Clinac-1800 dual energy medical linear accelerator.

Authors:  M Ravikumar; R Ravichandran
Journal:  Strahlenther Onkol       Date:  2000-05       Impact factor: 3.621

2.  Comparison of dosimetric characteristics of Siemens virtual and physical wedges.

Authors:  X R Zhu; M T Gillin; P A Jursinic; F Lopez; D F Grimm; J J Rownd
Journal:  Med Phys       Date:  2000-10       Impact factor: 4.071

3.  Measurement of dose in the buildup region using fixed-separation plane-parallel ionization chambers.

Authors:  B J Gerbi; F M Khan
Journal:  Med Phys       Date:  1990 Jan-Feb       Impact factor: 4.071

4.  The effect of a dynamic wedge in the medial tangential field upon the contralateral breast dose.

Authors:  B J McParland
Journal:  Int J Radiat Oncol Biol Phys       Date:  1990-12       Impact factor: 7.038

5.  Surface and peripheral doses of dynamic and physical wedges.

Authors:  Z Li; E E Klein
Journal:  Int J Radiat Oncol Biol Phys       Date:  1997-03-01       Impact factor: 7.038

6.  The polarity effect for commercially available plane-parallel ionization chambers.

Authors:  B J Gerbi; F M Khan
Journal:  Med Phys       Date:  1987 Mar-Apr       Impact factor: 4.071

7.  Entrance and exit dose regions for a Clinac-2100C.

Authors:  E E Klein; J A Purdy
Journal:  Int J Radiat Oncol Biol Phys       Date:  1993-09-30       Impact factor: 7.038

8.  Effects of beam modifiers and immobilization devices on the dose in the build-up region.

Authors:  D P Fontenla; J J Napoli; M Hunt; D Fass; B McCormick; G J Kutcher
Journal:  Int J Radiat Oncol Biol Phys       Date:  1994-08-30       Impact factor: 7.038

9.  A measurement and analysis of buildup region dose for open field photon beams (cobalt-60 through 24 MV).

Authors:  E C McCullough
Journal:  Med Dosim       Date:  1994       Impact factor: 1.482

10.  "Measurements of surface dose using build-up curves" obtained with an extrapolation chamber.

Authors:  D J Manson; D Velkley; J A Purdy; G D Oliver
Journal:  Radiology       Date:  1975-05       Impact factor: 11.105

View more
  8 in total

1.  Superficial dosimetry imaging based on Čerenkov emission for external beam radiotherapy with megavoltage x-ray beam.

Authors:  Rongxiao Zhang; Adam K Glaser; David J Gladstone; Colleen J Fox; Brian W Pogue
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071

2.  Superficial dosimetry imaging of Čerenkov emission in electron beam radiotherapy of phantoms.

Authors:  Rongxiao Zhang; Colleen J Fox; Adam K Glaser; David J Gladstone; Brian W Pogue
Journal:  Phys Med Biol       Date:  2013-07-24       Impact factor: 3.609

3.  Beam and tissue factors affecting Cherenkov image intensity for quantitative entrance and exit dosimetry on human tissue.

Authors:  Rongxiao Zhang; Adam K Glaser; Jacqueline Andreozzi; Shudong Jiang; Lesley A Jarvis; David J Gladstone; Brian W Pogue
Journal:  J Biophotonics       Date:  2016-08-10       Impact factor: 3.207

4.  The influence of the bolus-surface distance on the dose distribution in the build-up region.

Authors:  Martyna Sroka; Jakub Reguła; Włodzimierz Lobodziec
Journal:  Rep Pract Oncol Radiother       Date:  2010-11-04

5.  Characterization of GafChromic EBT2 film dose measurements using a tissue-equivalent water phantom for a Theratron® Equinox Cobalt-60 teletherapy machine.

Authors:  Daniel Akwei Addo; Elsie Effah Kaufmann; Samuel Nii Tagoe; Augustine Kwame Kyere
Journal:  PLoS One       Date:  2022-08-19       Impact factor: 3.752

6.  An investigation of the depth dose in the build-up region, and surface dose for a 6-MV therapeutic photon beam: Monte Carlo simulation and measurements.

Authors:  Lukkana Apipunyasopon; Somyot Srisatit; Nakorn Phaisangittisakul
Journal:  J Radiat Res       Date:  2012-10-26       Impact factor: 2.724

7.  Equivalent square formula for determining the surface dose of rectangular field from 6 MV therapeutic photon beam.

Authors:  Lukkana Apipunyasopon; Somyot Srisatit; Nakorn Phaisangittisakul
Journal:  J Appl Clin Med Phys       Date:  2013-09-06       Impact factor: 2.102

8.  The Measurement of the Surface Dose in Regular and Small Radiation Therapy Fields Using Cherenkov Imaging.

Authors:  Yi Li; HongJun Liu; Nan Huang; Zhaolu Wang; Chunmin Zhang
Journal:  Technol Cancer Res Treat       Date:  2022 Jan-Dec
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.