Literature DB >> 3587140

The polarity effect for commercially available plane-parallel ionization chambers.

B J Gerbi, F M Khan.   

Abstract

The polarity effect was investigated for three different commercially available plane-parallel ionization chambers: the Memorial Pipe chamber, the Victoreen/Nuclear Associates model 30-329 chamber manufactured by PTW, Frieburg, and the Capintec PS-033 thin-window ionization chamber. The primary study was the polarity effect versus depth below the phantom surface for 6-, 10-, 18-, and 24-MV x-ray beams, and 9- and 22-MeV electron beams. The polarity effect in the region of nonelectronic equilibrium that exists at the interface of two dissimilar materials, polystyrene and aluminum, was investigated as well as the effects of field size. For the group of plane-parallel ionization chambers that we studied, we found a polarity effect of only 1%-2% for electron beams at the depth of dmax. At depths greater than dmax, the polarity effect for electrons increased and was as high as 4.5% for some chambers. When used in the buildup region of high-energy photon beams, these same chambers exhibited up to a 30% difference in collected charge between one polarity and the other. This effect and its relationship to physical chamber characteristics is discussed.

Entities:  

Mesh:

Year:  1987        PMID: 3587140     DOI: 10.1118/1.596072

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  Effect of a carbon fiber tabletop on the surface dose and attenuation for high-energy photon beams.

Authors:  Tülay P Meydanci; Gönül Kemikler
Journal:  Radiat Med       Date:  2008-11-22

2.  Surface dose and build-up region measurements with wedge filters for 6 and 18 MV photon beams.

Authors:  Hatice Bilge; Nurdan Ozbek; Murat Okutan; Aydin Cakir; Hilal Acar
Journal:  Jpn J Radiol       Date:  2010-02-26       Impact factor: 2.374

3.  Polarity effect in commercial ionization chambers used in photon beams with small fields.

Authors:  Tetsunori Shimono; Kichiro Koshida; Hidekazu Nambu; Kosuke Matsubara; Hiroshi Takahashi; Hiroshi Okuda
Journal:  Radiol Phys Technol       Date:  2008-12-18

4.  Effect of various physical parameters on surface and build-up dose for 15-MV X-rays.

Authors:  Girigesh Yadav; R S Yadav; Alok Kumar
Journal:  J Med Phys       Date:  2010-10

5.  Skin dose estimation for various beam modifiers and source-to-surface distances for 6MV photons.

Authors:  Girigesh Yadav; R S Yadav; Alok Kumar
Journal:  J Med Phys       Date:  2009-04

6.  Surface dose variations in 6 and 10 MV flattened and flattening filter-free (FFF) photon beams.

Authors:  Jason Cashmore
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

  6 in total