| Literature DB >> 20182549 |
Ge Wang1, Wenxiang Cong, Haiou Shen, Yu Zou.
Abstract
Although x-ray imaging is widely used in biomedical applications, biological soft tissues have small density changes, leading to low contrast resolution for attenuation-based x-ray imaging. Over the past years, x-ray small-angle scattering was studied as a new contrast mechanism to enhance subtle structural variation within the soft tissue. In this paper, we present a detection method to extract this type of x-ray scattering data, which are also referred to as dark-field signals. The key idea is to acquire an x-ray projection multiple times with varying collimation before an x-ray detector array. The projection data acquired with a collimator of a sufficiently high collimation aspect ratio contain mainly the primary beam with little scattering, while the data acquired with an appropriately reduced collimation aspect ratio include both the primary beam and small-angle scattering signals. Then, analysis of these corresponding datasets will produce desirable dark-field signals; for example, via digitally subtraction. In the numerical experiments, the feasibility of our dark-field detection technology is demonstrated in Monte Carlo simulation. The results show that the acquired dark field signals can clearly reveal the structural information of tissues in terms of Rayleigh scattering characteristics.Entities:
Year: 2010 PMID: 20182549 PMCID: PMC2825654 DOI: 10.1155/2009/847537
Source DB: PubMed Journal: Int J Biomed Imaging ISSN: 1687-4188
Figure 1Single pass cone-beam circular scan with a dual collimation detector array for both dark-field tomography and transmission X-ray CT.
Figure 2Geometry for detection of X-ray scattering signals.
Figure 3Simulation setting for imaging based on small-angle scattering.
X-ray absorption and scattering coefficients at 50 KeV of the phantom components.
| Material | Absorption Coefficient (cm−1) | Scattering coefficients ( | ||
|---|---|---|---|---|
|
|
|
| ||
| Water | 0.028 | 0.0140 | 0.1680 | 0.0769 |
| Cube 1 | 0.028 | 0.0035 | 0.1785 | 0.0193 |
| Cube 2 | 0.028 | 0.0070 | 0.1750 | 0.0385 |
| Cube 3 | 0.028 | 0.0260 | 0.1560 | 0.1429 |
| Cube 4 | 0.028 | 0.0520 | 0.1300 | 0.2858 |
Figure 4Numerical results on X-ray small-angle scattering imaging. (a) An image collected with a high collimation aspect and a small object-detector distance (2.5 cm), (b) an image with a low collimation aspect and a small object-detector distance (2.5 cm), (c) an image with a low collimation aspect and a large object-detector distance (17.5 cm), (d) the difference between (a) and (b), and (e) the difference between (b) and (c).
Figure 5Numerical results on X-ray single scattering imaging. (a) A single scattering image from (6)-(7), and (b) a pure small angle scattering image from the Monte Carlo simulation.