Hypoglycemia alarm enhancement using data fusion.

Victor N Skladnev1, Stanislav Tarnavskii, Thomas McGregor, Nejhdeh Ghevondian, Steve Gourlay, Timothy W Jones.   

Abstract

BACKGROUND: The acceptance of closed-loop blood glucose (BG) control using continuous glucose monitoring systems (CGMS) is likely to improve with enhanced performance of their integral hypoglycemia alarms. This article presents an in silico analysis (based on clinical data) of a modeled CGMS alarm system with trained thresholds on type 1 diabetes mellitus (T1DM) patients that is augmented by sensor fusion from a prototype hypoglycemia alarm system (HypoMon). This prototype alarm system is based on largely independent autonomic nervous system (ANS) response features.
METHODS: Alarm performance was modeled using overnight BG profiles recorded previously on 98 T1DM volunteers. These data included the corresponding ANS response features detected by HypoMon (AiMedics Pty. Ltd.) systems. CGMS data and alarms were simulated by applying a probabilistic model to these overnight BG profiles. The probabilistic model developed used a mean response delay of 7.1 minutes, measurement error offsets on each sample of +/- standard deviation (SD) = 4.5 mg/dl (0.25 mmol/liter), and vertical shifts (calibration offsets) of +/- SD = 19.8 mg/dl (1.1 mmol/liter). Modeling produced 90 to 100 simulated measurements per patient. Alarm systems for all analyses were optimized on a training set of 46 patients and evaluated on the test set of 56 patients. The split between the sets was based on enrollment dates. Optimization was based on detection accuracy but not time to detection for these analyses. The contribution of this form of data fusion to hypoglycemia alarm performance was evaluated by comparing the performance of the trained CGMS and fused data algorithms on the test set under the same evaluation conditions.
RESULTS: The simulated addition of HypoMon data produced an improvement in CGMS hypoglycemia alarm performance of 10% at equal specificity. Sensitivity improved from 87% (CGMS as stand-alone measurement) to 97% for the enhanced alarm system. Specificity was maintained constant at 85%. Positive predictive values on the test set improved from 61 to 66% with negative predictive values improving from 96 to 99%. These enhancements were stable within sensitivity analyses. Sensitivity analyses also suggested larger performance increases at lower CGMS alarm performance levels.
CONCLUSION: Autonomic nervous system response features provide complementary information suitable for fusion with CGMS data to enhance nocturnal hypoglycemia alarms. 2010 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20167165      PMCID: PMC2825622          DOI: 10.1177/193229681000400105

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  10 in total

1.  Comparison of a needle-type and a microdialysis continuous glucose monitor in type 1 diabetic patients.

Authors:  Iris M Wentholt; Marit A Vollebregt; Augustus A Hart; Joost B Hoekstra; J Hans DeVries
Journal:  Diabetes Care       Date:  2005-12       Impact factor: 19.112

2.  Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements.

Authors:  Richard L Weinstein; Sherwyn L Schwartz; Ronald L Brazg; Jolyon R Bugler; Thomas A Peyser; Geoffrey V McGarraugh
Journal:  Diabetes Care       Date:  2007-03-02       Impact factor: 19.112

3.  Peculiarities of the continuous glucose monitoring data stream and their impact on developing closed-loop control technology.

Authors:  Boris Kovatchev; William Clarke
Journal:  J Diabetes Sci Technol       Date:  2008-01

4.  Can we really close the loop and how soon? Accelerating the availability of an artificial pancreas: a roadmap to better diabetes outcomes.

Authors:  Aaron J Kowalski
Journal:  Diabetes Technol Ther       Date:  2009-06       Impact factor: 6.118

5.  The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus.

Authors:  D M Nathan; S Genuth; J Lachin; P Cleary; O Crofford; M Davis; L Rand; C Siebert
Journal:  N Engl J Med       Date:  1993-09-30       Impact factor: 91.245

6.  Reducing false alarm rates for critical arrhythmias using the arterial blood pressure waveform.

Authors:  Anton Aboukhalil; Larry Nielsen; Mohammed Saeed; Roger G Mark; Gari D Clifford
Journal:  J Biomed Inform       Date:  2008-03-21       Impact factor: 6.317

7.  Neural-network detection of hypoglycemic episodes in children with type 1 diabetes using physiological parameters.

Authors:  Hung T Nguyen; Nejhdeh Ghevondian; Timothy W Jones
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2006

8.  Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes (UKPDS 33). UK Prospective Diabetes Study (UKPDS) Group.

Authors: 
Journal:  Lancet       Date:  1998-09-12       Impact factor: 79.321

9.  Spurious reporting of nocturnal hypoglycemia by CGMS in patients with tightly controlled type 1 diabetes.

Authors:  Kilty McGowan; William Thomas; Antoinette Moran
Journal:  Diabetes Care       Date:  2002-09       Impact factor: 19.112

10.  Clinical evaluation of a noninvasive alarm system for nocturnal hypoglycemia.

Authors:  Victor N Skladnev; Nejhdeh Ghevondian; Stanislav Tarnavskii; Nirubasini Paramalingam; Timothy W Jones
Journal:  J Diabetes Sci Technol       Date:  2010-01-01
  10 in total
  7 in total

1.  Paper electrocardiograph strips may contain overlooked clinical information in screen-detected type 2 diabetes patients.

Authors:  Jesper Fleischer; Morten Charles; Lise Tarnow; Klaus Skovbo Jensen; Hans Nygaard; Annelli Sandbaek; Niels Ejskjaer
Journal:  J Diabetes Sci Technol       Date:  2012-01-01

2.  Combining information of autonomic modulation and CGM measurements enables prediction and improves detection of spontaneous hypoglycemic events.

Authors:  Simon Lebech Cichosz; Jan Frystyk; Lise Tarnow; Jesper Fleischer
Journal:  J Diabetes Sci Technol       Date:  2014-09-12

Review 3.  Toward Big Data Analytics: Review of Predictive Models in Management of Diabetes and Its Complications.

Authors:  Simon Lebech Cichosz; Mette Dencker Johansen; Ole Hejlesen
Journal:  J Diabetes Sci Technol       Date:  2015-10-14

4.  The degree of autonomic modulation is associated with the severity of microvascular complications in patients with type 1 diabetes.

Authors:  Jesper Fleischer; Simon Lebech Cichosz; Poul Erik Jakobsen; Knud Yderstraede; Elisabeth Gulichsen; Hans Nygaard; Ebbe Eldrup; Hans Henrik Lervang; Lise Tarnow; Niels Ejskjaer
Journal:  J Diabetes Sci Technol       Date:  2015-01-14

Review 5.  Diabetic autonomic imbalance and glycemic variability.

Authors:  Jesper Fleischer
Journal:  J Diabetes Sci Technol       Date:  2012-09-01

6.  A novel algorithm for prediction and detection of hypoglycemia based on continuous glucose monitoring and heart rate variability in patients with type 1 diabetes.

Authors:  Simon Lebech Cichosz; Jan Frystyk; Ole K Hejlesen; Lise Tarnow; Jesper Fleischer
Journal:  J Diabetes Sci Technol       Date:  2014-03-31

7.  "Turn it off!": diabetes device alarm fatigue considerations for the present and the future.

Authors:  Joseph P Shivers; Linda Mackowiak; Henry Anhalt; Howard Zisser
Journal:  J Diabetes Sci Technol       Date:  2013-05-01
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.