Literature DB >> 20158534

Cyclic fatigue of NiTi rotary instruments in a simulated apical abrupt curvature.

G Plotino1, N M Grande, M C Melo, M G Bahia, L Testarelli, G Gambarini.   

Abstract

AIM: To evaluate the cyclic fatigue resistance of five NiTi rotary systems in an abrupt apical curvature.
METHODOLOGY: Cyclic fatigue testing was performed in stainless-steel artificial canals with a 2-mm radius of curvature and an angle of curvature of 90 degrees constructed to the dimensions of the instruments tested. The middle of the simulated curvature was 2.5 mm from the tip of the instrument that was placed at full working length. All instruments were new and 25 mm in length. Ten ProTaper Universal F2 (Dentsply Maillefer, Ballaigues, Switzerland); FlexMaster (VDW, Munich, Germany) tip size 25, taper 0.06; Mtwo (Sweden & Martina, Padova, Italy) tip size 25, taper 0.06; ProFile tip size 25, taper 0.06 from Dentsply Maillefer (Ballaigues, Switzerland); and ProFile tip size 25, taper 0.06 from Dentsply Tulsa (Tulsa, OK, USA) were rotated passively at 300 rpm until fracture occurred, and the number of cycles to failure (NCF) recorded. Length of the fractured tip was measured. Data were analysed by one-way anova and Tukey HSD test to determine any statistical difference amongst groups.
RESULTS: Mtwo had the highest fatigue resistance compared to the other instruments (NCF 124 +/- 25) (P < 0.001); there was no statistical difference between ProFile from the two different brands, although ProFile from Maillefer had the higher fatigue life (NCF 75 +/- 10) compared to ProFile from Tulsa (NCF 66 +/- 10). No difference was registered between FlexMaster (NCF 53 +/- 5) and ProFile from Tulsa; ProTaper F2 had a significantly (P < 0.001) lower fatigue life compared to the other instruments tested (NCF 29 +/- 5).
CONCLUSIONS: Lifespan registered for the instruments tested in an apical abrupt curvature was Mtwo > ProFile from Maillefer > ProFile from Tulsa > FlexMaster > ProTaper.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20158534     DOI: 10.1111/j.1365-2591.2009.01668.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  15 in total

1.  Fatigue resistance of rotary instruments manufactured using different nickel-titanium alloys: a comparative study.

Authors:  Gianluca Plotino; Luca Testarelli; Dina Al-Sudani; Gianlcarlo Pongione; Nicola M Grande; Gianluca Gambarini
Journal:  Odontology       Date:  2012-10-18       Impact factor: 2.634

2.  Flexibility and resistance to cyclic fatigue of endodontic instruments made with different nickel-titanium alloys: a comparative test.

Authors:  Giancarlo Pongione; Giorgio Pompa; Valerio Milana; Stefano Di Carlo; Alessio Giansiracusa; Emanuele Nicolini; Francesca De Angelis
Journal:  Ann Stomatol (Roma)       Date:  2012-01-14

3.  Cyclic fatigue resistances of Hyflex EDM, WaveOne gold, Reciproc blue and 2shape NiTi rotary files in different artificial canals.

Authors:  Taha Özyürek; Mustafa Gündoğar; Gülşah Uslu; Koray Yılmaz; Simone Staffoli; Grande Nm; Gianluca Plotino; Antonella Polimeni
Journal:  Odontology       Date:  2018-01-30       Impact factor: 2.634

4.  The influence of three different instrumentation techniques on the incidence of postoperative pain after endodontic treatment.

Authors:  Gianluca Gambarini; Luca Testarelli; Massimo De Luca; Valerio Milana; Gianluca Plotino; Nicola Maria Grande; Alessio Giansiracusa Rubini; Dina Al Sudani; Gianpaolo Sannino
Journal:  Ann Stomatol (Roma)       Date:  2013-03-20

5.  Analysis of the Importance of the Motion Used in the Resistance of Different Mechanical Instrumentation Systems in Endodontics: A Comparative Study.

Authors:  Jesús Mena-Álvarez; Manuel Almanzor-López; Norberto Quispe-López; Ana De Pedro-Muñoz; Cristina Rico-Romano
Journal:  Materials (Basel)       Date:  2022-06-24       Impact factor: 3.748

6.  Cyclic fatigue resistance of D-RaCe, ProTaper, and Mtwo nickel-titanium retreatment instruments after immersion in sodium hypochlorite.

Authors:  Hüseyin Sinan Topçuoğlu; Kanşad Pala; Ahmet Aktı; Salih Düzgün; Gamze Topçuoğlu
Journal:  Clin Oral Investig       Date:  2015-09-30       Impact factor: 3.573

7.  Cyclic fatigue resistance of Mtwo NiTi rotary instruments used by experienced and novice operators--an in vivo and in vitro study.

Authors:  Gianluca Plotino; Dina Al-Sudani; Stefano Pulino; Nicola M Grande; Piero Alessandro Marcoli; Silvia Pizzi; Luca Testarelli; Gianluca Gambarini
Journal:  Med Sci Monit       Date:  2012-06

8.  Treatment of Necrotic Teeth Using Two Engine-Driven Systems and Patient's Postoperative Pain: A Double-Blind Clinical Trial.

Authors:  Vahid Zand; Amin Salem Milani; Ayla Hassani Dehkharghani; Mahdi Rahbar; Pardis Tehranchi
Journal:  Iran Endod J       Date:  2016

9.  Mechanical behavior of deep cryogenically treated martensitic shape memory nickel-titanium rotary endodontic instruments.

Authors:  Thilla Sekar Vinothkumar; Deivanayagam Kandaswamy; Gopalakrishnan Prabhakaran; Arunachalam Rajadurai
Journal:  Eur J Dent       Date:  2016 Apr-Jun

10.  Evaluation of the incidence of microcracks caused by Mtwo and ProTaper Next rotary file systems versus the self-adjusting file: A scanning electron microscopic study.

Authors:  Suparna Ganguly Saha; Neelam Vijaywargiya; Divya Saxena; Mainak Kanti Saha; Anuj Bharadwaj; Sandeep Dubey
Journal:  J Conserv Dent       Date:  2017 Sep-Oct
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.