Performance analysis of the OneTouch UltraVue blood glucose monitoring system.

Anna Chang1, Alice Orth, Bryan Le, Perla Menchavez, Lupe Miller.   

Abstract

BACKGROUND: OneTouch UltraVue is a new meter for self-monitoring of blood glucose that includes a color display, used-strip ejector, and no-button interface. The system uses an electrochemical biosensor technology based on glucose oxidase chemistry to detect glucose concentrations from 20 to 600 mg/dl (1.1 to 33.3 mmol/liter).
METHODS: Accuracy and reproducibility were evaluated over a wide range of glucose concentrations according to standard criteria. Clinical accuracy was assessed by health care providers (HCPs) in two studies and by diabetes patients in the second study. Reference glucose levels were determined by a YSI 2300 analyzer. Same-day reproducibility and day-to-day reproducibility were also evaluated.
RESULTS: In the accuracy studies, 99.7% and 98.7% of tests by HCPs and 97.0% of tests by patients were within +/-15 mg/dl (+/-0.8 mmol/liter) of the YSI reference for blood glucose <75 mg/dl (<4.2 mmol/liter), and within +/-20% for blood glucose > or =75 mg/dl (> or =4.2 mmol/liter), respectively. Consensus error grid analysis showed that 99.7% and 95.3% of tests by HCPs and 97.0% of tests by patients fell within zone A (i.e., has no effect on clinical action); all other results were in zone B (i.e., altered clinical action, little or no effect on clinical outcome). In the reproducibility studies, the standard deviation was <1.5 mg/dl (<0.1 mmol/liter) for glucose concentrations <100 mg/dl (<5.6 mmol/liter), and the coefficient of variation was <2% for concentrations > or = 100 mg/dl (> or =5.6 mmol/liter).
CONCLUSIONS: OneTouch UltraVue meets standard acceptability criteria for accuracy and reproducibility across a wide range of glucose concentrations. Its simple interface and lack of contact with used strips make it a viable option for older patients and their caregivers. 2009 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 20144431      PMCID: PMC2769918          DOI: 10.1177/193229680900300521

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  9 in total

1.  A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose.

Authors:  J L Parkes; S L Slatin; S Pardo; B H Ginsberg
Journal:  Diabetes Care       Date:  2000-08       Impact factor: 19.112

2.  Point-of-care glucose testing: effects of critical care variables, influence of reference instruments, and a modular glucose meter design.

Authors:  R F Louie; Z Tang; D V Sutton; J H Lee; G J Kost
Journal:  Arch Pathol Lab Med       Date:  2000-02       Impact factor: 5.534

3.  American Association of Clinical Endocrinologists medical guidelines for clinical practice for the management of diabetes mellitus.

Authors:  Helena W Rodbard; Lawrence Blonde; Susan S Braithwaite; Elise M Brett; Rhoda H Cobin; Yehuda Handelsman; Richard Hellman; Paul S Jellinger; Lois G Jovanovic; Philip Levy; Jeffrey I Mechanick; Farhad Zangeneh
Journal:  Endocr Pract       Date:  2007 May-Jun       Impact factor: 3.443

4.  Self-monitoring of blood glucose in type 2 diabetes: a new look at published trials.

Authors:  K Kempf; W Neukirchen; S Martin; H Kolb
Journal:  Diabetologia       Date:  2008-02-20       Impact factor: 10.122

5.  Standards of medical care in diabetes--2008.

Authors: 
Journal:  Diabetes Care       Date:  2008-01       Impact factor: 19.112

6.  Performance evaluation of a new blood glucose monitor that requires no coding: the OneTouch Vita System.

Authors:  Joanna K Young; John M Ellison; Robert Marshall
Journal:  J Diabetes Sci Technol       Date:  2008-09

Review 7.  The role of self-monitoring of blood glucose in the care of people with diabetes: report of a global consensus conference.

Authors:  Richard M Bergenstal; James R Gavin
Journal:  Am J Med       Date:  2005-09       Impact factor: 4.965

Review 8.  Assessing glucose meter accuracy.

Authors:  Darius Arabadjief; James H Nichols
Journal:  Curr Med Res Opin       Date:  2006-11       Impact factor: 2.580

9.  Global prevalence of diabetes: estimates for the year 2000 and projections for 2030.

Authors:  Sarah Wild; Gojka Roglic; Anders Green; Richard Sicree; Hilary King
Journal:  Diabetes Care       Date:  2004-05       Impact factor: 19.112

  9 in total
  3 in total

1.  Analysis article on the performance analysis of the OneTouch UltraVue blood glucose monitoring system.

Authors:  Bogdan Solnica
Journal:  J Diabetes Sci Technol       Date:  2009-09-01

2.  Fabrication of functionalized carbon nanotube buckypaper electrodes for application in glucose biosensors.

Authors:  Henry Papa; Melissa Gaillard; Leon Gonzalez; Jhunu Chatterjee
Journal:  Biosensors (Basel)       Date:  2014-11-11

3.  Human umbilical cord Wharton jelly cells promote extra-pancreatic insulin formation and repair of renal damage in STZ-induced diabetic mice.

Authors:  Martin Maldonado; Tianhua Huang; Lujun Yang; Lan Xu; Lian Ma
Journal:  Cell Commun Signal       Date:  2017-10-17       Impact factor: 5.712

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.