PURPOSE: To evaluate the safety and effectiveness of the aspheric diffractive Tecnis multifocal ZM900 (TCMF) intraocular lens (IOL; Abbott Medical Optics, Inc, Santa Ana, California, USA).
DESIGN: One-year, nonrandomized, multicenter, evaluator-masked, bilateral, parallel-group comparative clinical evaluation.
METHODS: Subjects underwent bilateral implantation with the TCMF IOL or the CeeOn 911A monofocal (CEMN) IOL (Abbott Medical Optics, Inc) according to subject preference.
RESULTS: One-year results were available for 244 eyes of 125 TCMF IOL subjects and for 245 eyes of 123 CEMN IOL subjects. Mean distance visual acuities (VAs) were statistically and clinically equivalent between the 2 groups. Mean binocular and monocular uncorrected and distance-corrected near VAs were significantly better for the TCMF IOL versus the CEMN IOL group (P < .0001). A greater proportion of TCMF IOL versus CEMN IOL subjects achieved binocular combined VAs of 20/25 distance and 20/32 near (84.2% [96/114] vs 6.2% [7/113]; P < .0001). The TCMF IOL group had excellent depth of focus, maintaining a mean of 20/40 or better for far, intermediate, and near distances. Mean contrast sensitivity scores were lower for the TCMF IOL versus the CEMN IOL group, but the differences were not considered clinically significant. Halos and night glare were more common in the TCMF IOL versus the CEMN IOL groups. Both reading acuity (20/20 vs 20/47; P < .0007) and speed (148 vs 117 words/minute; P < .0001) were significantly better for the TCMF IOL versus the CEMN IOL group, as were the proportions of the TCMF IOL group versus the CEMN IOL group that achieved spectacle independence (84.8% vs 5.2%; P < .0001) and that functioned comfortably at near without glasses (96.4% vs 30.4%; P < .0001).
CONCLUSIONS: The TCMF IOL provided enhanced near VA, reading acuity per speed, depth of focus, and spectacle independence compared with the monofocal IOL. Copyright 2010 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20138609     DOI: 10.1016/j.ajo.2009.10.022

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  24 in total

Review 1.  Premium intraocular lenses use in patients with cataract and concurrent glaucoma: a review.

Authors:  Raluca Iancu; Catalina Corbu
Journal:  Maedica (Buchar)       Date:  2013-09

2.  Visual performance with accommodating and multifocal intraocular lenses.

Authors:  Jie Lan; Yu-Sen Huang; Yun-Hai Dai; Xiao-Ming Wu; Jia-Jun Sun; Li-Xin Xie
Journal:  Int J Ophthalmol       Date:  2017-02-18       Impact factor: 1.779

3.  Visual acuity of pseudophakic patients predicted from in-vitro measurements of intraocular lenses with different design.

Authors:  Fidel Vega; Maria S Millán; Nuria Garzón; Irene Altemir; Francisco Poyales; Jose Manuel Larrosa
Journal:  Biomed Opt Express       Date:  2018-09-17       Impact factor: 3.732

4.  Objective evaluation of through-focus optical performance of presbyopia-correcting intraocular lenses using an optical bench system.

Authors:  Myoung Joon Kim; Len Zheleznyak; Scott Macrae; Hungwon Tchah; Geunyoung Yoon
Journal:  J Cataract Refract Surg       Date:  2011-07       Impact factor: 3.351

Review 5.  Personalized aspheric intraocular lens implantation based on corneal spherical aberration: a review.

Authors:  Wei Du; Wei Lou; Qiang Wu
Journal:  Int J Ophthalmol       Date:  2019-11-18       Impact factor: 1.779

6.  Clinical outcomes of a new diffractive multifocal intraocular lens.

Authors:  Baha Toygar; Ozge Yabas Kiziloglu; Okan Toygar; Ali Murat Hacimustafaoglu
Journal:  Int J Ophthalmol       Date:  2017-12-18       Impact factor: 1.779

7.  Visual outcome and optical quality after bilateral implantation of aspheric diffractive multifocal, aspheric monofocal and spherical monofocal intraocular lenses: a prospective comparison.

Authors:  Pan-Pan Ye; Xia Li; Ke Yao
Journal:  Int J Ophthalmol       Date:  2013-06-18       Impact factor: 1.779

8.  Comparison of through-focus image quality across five presbyopia-correcting intraocular lenses (an American Ophthalmological Society thesis).

Authors:  Jay S Pepose; Daozhi Wang; Griffith E Altmann
Journal:  Trans Am Ophthalmol Soc       Date:  2011-12

9.  Comparison of visual performance of multifocal intraocular lenses with same material monofocal intraocular lenses.

Authors:  Tomofusa Yamauchi; Hitoshi Tabuchi; Kosuke Takase; Hideharu Ohsugi; Zaigen Ohara; Yoshiaki Kiuchi
Journal:  PLoS One       Date:  2013-06-28       Impact factor: 3.240

10.  Intraocular lens multifocality combined with the compensation for corneal spherical aberration: a new concept of presbyopia-correcting intraocular lens.

Authors:  Reiner Friedrich
Journal:  Case Rep Ophthalmol       Date:  2012-10-24
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.