Clinical performance of a transmission detector array for the permanent supervision of IMRT deliveries.

Björn Poppe1, Hui Khee Looe, Ndimofor Chofor, Antje Rühmann, Dietrich Harder, Kay C Willborn.   

Abstract

BACKGROUND AND
PURPOSE: Clinical evaluation of a novel dosimetric accessory serving the permanent supervision of MLC function.
MATERIALS AND METHODS: The DAVID system (PTW-Freiburg, Germany) is a transparent, multi-wire transmission ionization chamber, placed in the accessory holder of the treatment head. Since each of the 37 individual wires is positioned exactly below the associated leaf pair of the MLC, its signal records the opening of this leaf pair during patient treatment.
RESULTS: The DAVID system closes a gap in the quality assurance program, permitting the permanent in-vivo verification of IMRT plans. During dosimetric plan verification with the 2D-ARRAY (PTW-Freiburg, Germany), reference values of the 37 DAVID signals are collected, with which the DAVID readings recorded during daily patient treatment are compared. This comparison is visually displayed in the control room, and warning and alarm levels of any discrepancies can be defined. The properties of the DAVID system as a transmission device, its sensitivity to beam delivery and leaflet errors as well as its stability have been analyzed for clinically relevant examples. In a recent version, the DAVID system has been equipped with 80 wires.
CONCLUSIONS: The DAVID system permits the on-line detection of clinically relevant MLC discrepancies in IMRT deliveries. Copyright 2010 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20138379     DOI: 10.1016/j.radonc.2009.12.041

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  14 in total

1.  Real-time dosimetry in external beam radiation therapy.

Authors:  Ramachandran Prabhakar
Journal:  World J Radiol       Date:  2013-10-28

2.  X-Ray Beam Segment Size and Entrance Location Effects on the Integral Quality Monitor (IQM®) Signal and Usefulness in Predicting Complex Segment Output Signals.

Authors:  Mahuvava C; Du Plessis F C P
Journal:  J Biomed Phys Eng       Date:  2020-08-01

3.  Dosimetric properties and clinical application of an a-Si EPID for dynamic IMRT quality assurance.

Authors:  Kenji Matsumoto; Masahiko Okumura; Yoshiyuki Asai; Kouhei Shimomura; Masaya Tamura; Yasumasa Nishimura
Journal:  Radiol Phys Technol       Date:  2012-12-04

4.  Sensitivity of the IQM and MatriXX detectors in megavolt photon beams.

Authors:  Oluwaseyi Michael Oderinde; Freek du Plessis
Journal:  Rep Pract Oncol Radiother       Date:  2019-08-02

5.  Usability of detecting delivery errors during treatment of prostate VMAT with a gantry-mounted transmission detector.

Authors:  Hirofumi Honda; Masahide Tominaga; Motoharu Sasaki; Masataka Oita; Hiromitsu Kanzaki; Yasushi Hamamoto; Yoshiaki Ishii; Ryuji Yamamoto; Teruhito Mochizuki; Teruhito Kido; Yoshihiro Uto
Journal:  J Appl Clin Med Phys       Date:  2021-05-05       Impact factor: 2.102

6.  Characterization and evaluation of an integrated quality monitoring system for online quality assurance of external beam radiation therapy.

Authors:  David Hoffman; Eunah Chung; Clayton Hess; Robin Stern; Stanley Benedict
Journal:  J Appl Clin Med Phys       Date:  2016-11-29       Impact factor: 2.102

7.  Beam perturbation characteristics of a 2D transmission silicon diode array, Magic Plate.

Authors:  Ziyad A Alrowaili; Michael L F Lerch; Marco Petasecca; Martin G Carolan; Peter E Metcalfe; Anatoly B Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2016-03-08       Impact factor: 2.102

8.  Validation of a method for in vivo 3D dose reconstruction in SBRT using a new transmission detector.

Authors:  Yuji Nakaguchi; Takeshi Ono; Masato Maruyama; Yoshinobu Shimohigashi; Yudai Kai
Journal:  J Appl Clin Med Phys       Date:  2017-06-02       Impact factor: 2.102

9.  Diode-based transmission detector for IMRT delivery monitoring: a validation study.

Authors:  Taoran Li; Q Jackie Wu; Thomas Matzen; Fang-Fang Yin; Jennifer C O'Daniel
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

10.  Evaluation of MLC leaf positioning accuracy for static and dynamic IMRT treatments using DAVID in vivo dosimetric system.

Authors:  Gulay Karagoz; Faruk Zorlu; Mete Yeginer; Demet Yildiz; Gokhan Ozyigit
Journal:  J Appl Clin Med Phys       Date:  2016-03-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.