| Literature DB >> 33955161 |
Hirofumi Honda1,2, Masahide Tominaga3, Motoharu Sasaki3, Masataka Oita4, Hiromitsu Kanzaki5, Yasushi Hamamoto5, Yoshiaki Ishii1, Ryuji Yamamoto1, Teruhito Mochizuki6, Teruhito Kido6, Yoshihiro Uto7.
Abstract
Volumetric-modulated arc therapy (VMAT) requires highly accurate control of multileaf collimator (MLC) movement, rotation speed of linear accelerator gantry, and monitor units during irradiation. Pretreatment validation and monitoring of these factors during irradiation are necessary for appropriate VMAT treatment. Recently, a gantry mounted transmission detector "Delta4 Discover® (D4D)" was developed to detect errors in delivering doses and dose distribution immediately after treatment. In this study, the performance of D4D was evaluated. Simulation plans, in which the MLC position was displaced by 0.5, 1.0, 1.5, 2.0, 2.5, and 3.0 mm from the clinically used original plans, were created for ten patients who received VMAT treatment for prostate cancer. Dose deviation (DD), distance-to-agreement (DTA), and gamma index analysis (GA) for each plan were evaluated by D4D. These results were compared to the results (DD, DTA and GA) measured by Delta4 Phantom + (D4P). We compared the deviations between the planned and measured values of the MLC stop positions A-side and B-side in five clinical cases of prostate VMAT during treatment and measured the GA values. For D4D, when the acceptable errors for DD, DTA, and GA were determined to be ≤3%, ≤2 mm, and ≤3%/2 mm, respectively, the minimum detectable errors in the MLC position were 2.0, 1.5, and 1.5 mm based on DD, DTA, and GA respectively. The corresponding minimum detectable MLC position errors were 2.0, 1.0, and 1.5 mm, respectively, for D4P. The deviation between the planned and measured position of MLC stopping point of prostate VMAT during treatment was stable at an average of -0.09 ± 0.05 mm, and all GA values were above 99.86%. In terms of delivering doses and dose distribution of VMAT, error detectability of D4D was comparable to that of D4P. The transmission-type detector "D4D" is thus suitable for detecting delivery errors during irradiation.Entities:
Keywords: prostate; transmission detector; volumetric-modulated arc therapy
Mesh:
Year: 2021 PMID: 33955161 PMCID: PMC8292707 DOI: 10.1002/acm2.13260
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Fig. 1Schematics of MLC same‐direction sifts. (a) Schematic of the original MLC position. (b) Schematic of the stop position of A‐side and B‐side MLC displaced 3.0 mm from B‐side to A‐side direction.
Fig. 2Procedure of D4D measurement. Combine D4P and D4D to measure the reference measurement using the "synthetic mode". Next, the fluences are converted into doses by directly measuring using D4D. * Reference measurement of this process should be performed for all dose verification cases.
Fig. 3Evaluation of D4D reproducibility in the same setup (D4D mounting reproducibility). (a) DD: the horizontal axis shows the D4P pass‐ratio, and the vertical axis shows the D4D pass‐ratio. The line connecting the values where D4P and D4D are equal is defined as the reference line. (b) DTA: the horizontal axis shows the D4P pass‐ratio, and the vertical axis shows the D4D pass‐ratio. (c) GA: the horizontal axis shows the D4P pass‐ratio, and the vertical axis shows the D4D pass‐ratio.
Fig. 4Evaluation of reproducibility of D4P setup (Reproducibility of reference measurement). (a) DD: the horizontal axis shows the D4P pass‐ratio, and the vertical axis shows the D4D pass‐ratio. The line connecting the values where D4P and D4D are equal is defined as the reference line. (b) DTA: the horizontal axis shows the D4P pass‐ratio, and the vertical axis shows the D4D pass‐ratio. (c) GA: the horizontal axis shows the D4P pass‐ratio, and the vertical axis shows the D4D pass‐ratio.
Fig. 5Evaluation of VMAT for ten prostate cancer patients. (a) DD: the horizontal axis shows the D4P pass‐ratio, and the vertical axis shows the D4D pass‐ratio. The line connecting the values where D4P and D4D are equal is defined as the reference line. (b) DTA: the horizontal axis shows the D4P pass‐ratio, and the vertical axis shows the D4D pass‐ratio. (c) GA: the horizontal axis shows the D4P pass‐ratio, and the vertical axis shows the D4D pass‐ratio.
Comparison of p‐value that all MLC position error simulation plans can detect in ten plans.
| Evaluation method | Device | MLC error (mm) | ||||||
|---|---|---|---|---|---|---|---|---|
| 0.0 | 0.5 | 1.0 | 1.5 | 2.0 | 2.5 | 3.0 | ||
| DD | D4P | 1.000 | <0.001 | <0.001 | 0.419 | <0.001 | <0.001 | <0.001 |
| D4D | 1.000 | <0.001 | <0.001 | 0.194 | 0.001 | <0.001 | <0.001 | |
| DTA | D4P | 1.000 | 0.528 | 0.003 | <0.001 | <0.001 | <0.001 | <0.001 |
| D4D | 1.000 | 0.397 | 0.175 | 0.001 | <0.001 | <0.001 | <0.001 | |
| GA | D4P | ‐ | 0.168 | 0.018 | 0.001 | <0.001 | <0.001 | <0.001 |
| D4D | ‐ | 0.168 | 0.084 | <0.001 | <0.001 | <0.001 | <0.001 | |
Statistical tests were performed on the significance of the original design and the simulated design (MLC position error 0.5, 1.0, 1.5, 2.0, 2.5, 3.0 mm) using Welch's t‐test. Both D4P and D4D were evaluated. Using a Bonferroni correction to account for multiple comparisons using 0.05, a p‐value was considered to be statistically significant when it was less than 0.007.
Statistically significant (p < 0.007), ( ‐ ) incalculable because D4P and D4D are 100%.
Variation in MLC Stop Position and GA of Prostate VMAT Clinical Data Using D4D.
| Patient | Arc number | Side | MLC error (mm) | Gamma analysis (%) |
|---|---|---|---|---|
| Patient1 | Arc 1 | A side | 0.04 ± 0.04 | 99.98 ± 0.08 |
| B side | −0.25 ± 0.05 | |||
| Arc 2 | A side | −0.07 ± 0.06 | 99.98 ± 0.11 | |
| B side | −0.31 ± 0.03 | |||
| Patient2 | Arc 1 | A side | 0.10 ± 0.05 | 99.86 ± 0.25 |
| B side | −0.29 ± 0.07 | |||
| Arc 2 | A side | −0.09 ± 0.04 | 99.93 ± 0.30 | |
| B side | −0.46 ± 0.03 | |||
| Patient3 | Arc 1 | A side | 0.15 ± 0.06 | 99.98 ± 0.06 |
| B side | −0.07 ± 0.05 | |||
| Arc 2 | A side | 0.17 ± 0.07 | 99.94 ± 0.12 | |
| B side | −0.10 ± 0.05 | |||
| Patient4 | Arc 1 | A side | 0.10 ± 0.05 | 99.94 ± 0.09 |
| B side | −0.08 ± 0.05 | |||
| Arc 2 | A side | 0.02 ± 0.06 | 100.00 ± 0.00 | |
| B side | −0.12 ± 0.07 | |||
| Patient5 | Arc 1 | A side | −0.08 ± 0.03 | 99.93 ± 0.10 |
| B side | −0.26 ± 0.05 | |||
| Arc 2 | A side | −0.03 ± 0.05 | 100.00 ± 0.00 | |
| B side | −0.22 ± 0.04 |
The average and 1SD of MLC stop position variation (A‐Side and B‐Side) of prostate VMAT clinical data by D4D were calculated. The average and 1SD of the GA of each 1arc plan were calculated.