Transfer of metallic debris after dislocation of ceramic femoral heads in hip prostheses.

Frank A Müller1, Marcel Hagymási, Peter Greil, Günther Zeiler, Alexander Schuh.   

Abstract

INTRODUCTION: Main reasons for the early revision of total hip arthroplasties are aseptic loosening as a result of wear debris caused by polyethylene particles, and postoperative dislocations. In revision arthroplasty of the hip metallic deposits were frequently found on femoral heads after dislocation or repositioning in total hip arthroplasties with metal back acetabular components.
MATERIALS AND METHODS: In this study we examined ceramic femoral heads from alumina-on-alumina and alumina-on-polymer pairings, respectively. All the heads resulted from revision of total hip replacement with a titanium containing acetabular component and showed metallic deposit on their surface. The metallic deposit was characterized by environmental scanning electron microscopy, energy dispersive X-ray spectroscopy and laser scanning microscopy.
RESULTS: The metallic deposits covered areas between 5 mm(2) and 8 cm(2) and reached a maximum height of 30 mum. They were shown to consist of titanium, and thus resulted from contact between the femoral head and the posterior part of the metal back. In case of twofold dislocated alumina-on-alumina pairings the titanium deposits showed different roughness profiles, whereas no differences were detectable for alumina-on-polymer pairings.
CONCLUSION: From the obtained results it can be concluded, that titanium deposited onto a femoral head during dislocation will be abraded within the harder alumina-on-alumina bearing couple during normal gait and will subsequently be released into the body by synovial fluid. In case of a ceramic-on-polymer bearing the deposited metal seems to be more harmful. The increased surface roughness of the femoral head within the tribological pairing probably initiates damage of the polymeric acetabular inlay and increases the wear of the bearing surface through a third-body wear mechanism resulting in the release of high amounts of PE.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16482451     DOI: 10.1007/s00402-006-0109-6

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  11 in total

1.  Distinctive damage patterns on THA metal bearing surfaces: case studies.

Authors:  Anneliese D Heiner; Nishant M Tikekar; Karen M Kruger; John J Lannutti; Thomas D Brown
Journal:  Iowa Orthop J       Date:  2014

2.  Enhancing damage visibility on metallic bearing surfaces: a simple technique for photography and viewing.

Authors:  Anneliese D Heiner; Karen M Kruger; Thomas E Baer; Thomas D Brown
Journal:  J Arthroplasty       Date:  2013-01-18       Impact factor: 4.757

3.  Surface Roughness of CoCr and ZrO(2) Femoral Heads with Metal Transfer: A Retrieval and Wear Simulator Study.

Authors:  Alan W Eberhardt; R Travis McKee; John M Cuckler; Donald W Peterson; Preston R Beck; Jack E Lemons
Journal:  Int J Biomater       Date:  2009-07-01

4.  Risk of impingement and third-body abrasion with 28-mm metal-on-metal bearings.

Authors:  Ian C Clarke; Jean-Yves Lazennec; Adrien Brusson; Christina Savisaar; John G Bowsher; Michelle Burgett; Thomas K Donaldson
Journal:  Clin Orthop Relat Res       Date:  2014-02       Impact factor: 4.176

5.  Sliding direction dependence of polyethylene wear for metal counterface traverse of severe scratches.

Authors:  Liam P Glennon; Thomas E Baer; James A Martin; William D Lack; Thomas D Brown
Journal:  J Biomech Eng       Date:  2008-10       Impact factor: 2.097

6.  A simulator study of adverse wear with metal and cement debris contamination in metal-on-metal hip bearings.

Authors:  T Halim; I C Clarke; M D Burgett-Moreno; T K Donaldson; C Savisaar; J G Bowsher
Journal:  Bone Joint Res       Date:  2014-03       Impact factor: 5.853

7.  Raman and Photoemission Spectroscopic Analyses of Explanted Biolox® Delta Femoral Heads Showing Metal Transfer.

Authors:  Paola Taddei; Eleonora Pavoni; Saverio Affatato
Journal:  Materials (Basel)       Date:  2017-07-03       Impact factor: 3.623

8.  Does Metal Transfer Differ on Retrieved Ceramic and CoCr Femoral Heads?

Authors:  Eliza K Fredette; Daniel W MacDonald; Richard J Underwood; Antonia F Chen; Michael A Mont; Gwo-Chin Lee; Gregg R Klein; Clare M Rimnac; Steven M Kurtz
Journal:  Biomed Res Int       Date:  2015-10-25       Impact factor: 3.411

9.  Are PEEK-on-Ceramic Bearings an Option for Total Disc Arthroplasty? An In Vitro Tribology Study.

Authors:  Ryan Siskey; Lauren Ciccarelli; Melissa K C Lui; Steven M Kurtz
Journal:  Clin Orthop Relat Res       Date:  2016-11       Impact factor: 4.176

10.  The reasons for ceramic-on-ceramic revisions between the third- and fourth-generation bearings in total hip arthroplasty from multicentric registry data.

Authors:  Sang-Min Kim; Kee Hyung Rhyu; Jeong Joon Yoo; Seung-Jae Lim; Je Hyun Yoo; Suc Hyun Kweon; Kyung-Jae Lee; Seung-Beom Han
Journal:  Sci Rep       Date:  2021-03-10       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.