Literature DB >> 17562426

Late fiber metal shedding of the first and second-generation Harris Galante acetabular component. A report of 5 cases.

David J Mayman1, Alejandro González Della Valle, Edward Lambert, John Anderson, Timothy Wright, Bryan Nestor, Thomas P Sculco, Eduardo A Salvati.   

Abstract

Five patients presented with fiber metal mesh shedding of a Harris Galante II acetabular cup detected between 11 and 15 years after implantation. All patients presented with hip pain and 4 demonstrated gross acetabular loosening and fiber metal separation on preoperative radiographs. The remaining patient underwent revision surgery because of a liner dislodgment and had a radiographically well fixed shell. Loosening and fiber metal separation were detected intraoperatively. Scanning electron microscopy of the retrieved shells demonstrated isolated diffusion bonding marks in the areas where the mesh separated from the substrate and no evidence of corrosion. Progressive osteolysis in the iliac bone was evident in 4 of our cases. Progressive iliac osteolysis may lead to loss of bone support in well-fixed cups and excessive stresses transferred to the interface between the fiber metal mesh and the titanium substrate leading to the separation of the 2 layers. Fiber metal separation may contribute to long-term loosening in the Harris Galante acetabular component.

Entities:  

Mesh:

Year:  2007        PMID: 17562426     DOI: 10.1016/j.arth.2006.05.027

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  1 in total

1.  Surface Roughness of CoCr and ZrO(2) Femoral Heads with Metal Transfer: A Retrieval and Wear Simulator Study.

Authors:  Alan W Eberhardt; R Travis McKee; John M Cuckler; Donald W Peterson; Preston R Beck; Jack E Lemons
Journal:  Int J Biomater       Date:  2009-07-01
  1 in total