Literature DB >> 20125059

The usefulness of waveform score from the ocular response analyzer.

Andrew K C Lam1, Davie Chen, Jimmy Tse.   

Abstract

PURPOSE: To study the usefulness of the Waveform Score from the ocular response analyzer (ORA).
METHODS: Both eyes of sixty-four normal Chinese adults were measured by the ORA. An experienced practitioner who was masked to the score evaluated the waveforms. Four measurements were obtained from each eye, and the average was considered as the gold standard. Agreement was compared for the mean of different multiple measurements to the gold standard.
RESULTS: There was no significant difference between the gold standard and the mean of different multiple measurements for all ORA parameters. The Waveform Score of 512 signals (four measurements from each eye of 64 subjects) ranged from 1.58 to 9.06. When the best signal value of four measured signals from each eye was considered, the lowest score was 3.60 from 128 eyes. The lower 10th percentile from all signals had Waveform Scores <3.48.
CONCLUSIONS: If the Waveform Score provides information on the reliability of the signals, a score <3.50 may indicate an unreliable signal, and the signal should be discarded. We also recommend taking three measurements with all the signals having a Waveform Score of 3.50 or above to increase the precision.

Entities:  

Mesh:

Year:  2010        PMID: 20125059     DOI: 10.1097/OPX.0b013e3181d1d940

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  20 in total

1.  Age-dependency of ocular parameters: a cross sectional study of young and elderly healthy subjects.

Authors:  G Jóhannesson; P Hallberg; K Ambarki; A Eklund; C Lindén
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-08-13       Impact factor: 3.117

2.  Dynamic contour tonometry over silicone hydrogel contact lens.

Authors:  Andrew K C Lam; Jimmy S H Tse
Journal:  J Optom       Date:  2013-08-28

3.  Corneal biomechanical properties after LASIK, ReLEx flex, and ReLEx smile by Scheimpflug-based dynamic tonometry.

Authors:  Iben Bach Pedersen; Sashia Bak-Nielsen; Anders Højslet Vestergaard; Anders Ivarsen; Jesper Hjortdal
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-05-27       Impact factor: 3.117

4.  Measuring mechanical wave speed, dispersion, and viscoelastic modulus of the cornea using optical coherence elastography.

Authors:  Antoine Ramier; Behrouz Tavakol; Seok-Hyun Yun
Journal:  Opt Express       Date:  2019-06-10       Impact factor: 3.894

5.  Comparison of corneal biomechanical properties in normal tension glaucoma patients with different visual field progression speed.

Authors:  Ying Hong; Nobuyuki Shoji; Tetsuya Morita; Kazunori Hirasawa; Kazuhiro Matsumura; Masayuki Kasahara; Kimiya Shimizu
Journal:  Int J Ophthalmol       Date:  2016-07-18       Impact factor: 1.779

6.  Analyzing biomechanical parameters of the cornea with glaucoma severity in open-angle glaucoma.

Authors:  Karin R Pillunat; Cosima Hermann; Eberhard Spoerl; Lutz E Pillunat
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2016-04-27       Impact factor: 3.117

7.  Corneal biomechanical properties and glaucoma-related quantitative traits in the EPIC-Norfolk Eye Study.

Authors:  Anthony P Khawaja; Michelle P Y Chan; David C Broadway; David F Garway-Heath; Robert Luben; Jennifer L Y Yip; Shabina Hayat; Kay-Tee Khaw; Paul J Foster
Journal:  Invest Ophthalmol Vis Sci       Date:  2014-01-07       Impact factor: 4.799

8.  Intraocular pressure measurement by three different tonometers in primary congenital glaucoma.

Authors:  Athar Zareei; Mohammad Reza Razeghinejad; Mohammad Hosein Nowroozzadeh; Yadollah Mehrabi; Mohammad Aghazadeh-Amiri
Journal:  J Ophthalmic Vis Res       Date:  2015 Jan-Mar

9.  Measuring corneal hysteresis: threshold estimation of the waveform score from the Ocular Response Analyzer.

Authors:  Marcelo Ayala; Enping Chen
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-05-09       Impact factor: 3.117

10.  Corneal biomechanics in iatrogenic ectasia and keratoconus: A review of the literature.

Authors:  Majid Moshirfar; Jason N Edmonds; Nicholas L Behunin; Steven M Christiansen
Journal:  Oman J Ophthalmol       Date:  2013-01
View more