Concept of minimal heart rate for each pitch value to avoid interpolation artifact when using dual-source CT: a phantom study.

Joon-Won Kang1, Kyung-Hyun Do, Jae-Yeon Chung, Hyun Jong Cho, Joon Beom Seo, Tae-Hwan Lim.   

Abstract

Interpolation artifact is known to occur when the heart rate is decreased lower than the critical value for the specific pitch. The purpose of our study is to determine the minimum heart rate (minHR) for the specific pitch that provides images without interpolation artifact when using dual-source computed tomography (DSCT). We scanned the 'thin slice thickness block' of the CT performance phantom provided by the American Association of Physicists in Medicine using DSCT for variable pitches. Change in heart rate was simulated through ECG editing by changing R-R interval. Axial, sagittal, and coronal image sets were reconstructed and assessed for the presence and extent of interpolation artifact. MinHR at which no interpolation artifact was detected for each pitch value was determined. Length of interpolation artifact (LOA) on sagittal view was also measured when the heart rate was simulated at 10 bpm lower than the minHR on each pitch setting. MinHRs for each pitch value were 9-10 bpm from the estimated heart rate. However, minHR for the lowest pitch value 0.2, estimated heart value 40 bpm was 37 bpm. LOA was larger in the low heart rate condition. Measured values of minHR were correlated exactly with the calculated values. MinHRs that provide images without interpolation artifact for each pitch value when using DSCT were determined. The concept of minHR is important for obtaining high quality images of cardiac CT angiography when using DSCT.

Mesh:

Year:  2010        PMID: 20119848     DOI: 10.1007/s10554-010-9586-3

Source DB:  PubMed          Journal:  Int J Cardiovasc Imaging        ISSN: 1569-5794            Impact factor:   2.357


  13 in total

1.  Subsecond multi-slice computed tomography: basics and applications.

Authors:  K Klingenbeck-Regn; S Schaller; T Flohr; B Ohnesorge; A F Kopp; U Baum
Journal:  Eur J Radiol       Date:  1999-08       Impact factor: 3.528

2.  Dual-source CT cardiac imaging: initial experience.

Authors:  Thorsten R C Johnson; Konstantin Nikolaou; Bernd J Wintersperger; Alexander W Leber; Franz von Ziegler; Carsten Rist; Sonja Buhmann; Andreas Knez; Maximilian F Reiser; Christoph R Becker
Journal:  Eur Radiol       Date:  2006-05-13       Impact factor: 5.315

3.  Contrast-enhanced coronary artery visualization by dual-source computed tomography--initial experience.

Authors:  Stephan Achenbach; Dieter Ropers; Axel Kuettner; Thomas Flohr; Bernd Ohnesorge; Herbert Bruder; Heike Theessen; Meri Karakaya; Werner G Daniel; Werner Bautz; Willi A Kalender; Katharina Anders
Journal:  Eur J Radiol       Date:  2006-01-19       Impact factor: 3.528

Review 4.  Relationship between noise, dose, and pitch in cardiac multi-detector row CT.

Authors:  Andrew N Primak; Cynthia H McCollough; Michael R Bruesewitz; Jie Zhang; Joel G Fletcher
Journal:  Radiographics       Date:  2006 Nov-Dec       Impact factor: 5.333

Review 5.  Artifacts in ECG-synchronized MDCT coronary angiography.

Authors:  L J M Kroft; A de Roos; J Geleijns
Journal:  AJR Am J Roentgenol       Date:  2007-09       Impact factor: 3.959

Review 6.  Physics of cardiac imaging with multiple-row detector CT.

Authors:  Mahadevappa Mahesh; Dianna D Cody
Journal:  Radiographics       Date:  2007 Sep-Oct       Impact factor: 5.333

7.  Noninvasive coronary angiography with 64-section CT: effect of average heart rate and heart rate variability on image quality.

Authors:  Sebastian Leschka; Simon Wildermuth; Thomas Boehm; Lotus Desbiolles; Lars Husmann; André Plass; Pascal Koepfli; Tiziano Schepis; Borut Marincek; Philipp A Kaufmann; Hatem Alkadhi
Journal:  Radiology       Date:  2006-09-11       Impact factor: 11.105

8.  Noninvasive imaging of the coronary arteries using a 64-row multidetector CT scanner: initial clinical experience and radiation dose concerns.

Authors:  M Francone; A Napoli; I Carbone; M Cavacece; P G Nardis; K Lanciotti; S Visconti; L Bertoletti; E Di Castro; C Catalano; R Passariello
Journal:  Radiol Med       Date:  2007-02-22       Impact factor: 3.469

9.  Dual-source CT: effect of heart rate, heart rate variability, and calcification on image quality and diagnostic accuracy.

Authors:  Harald Brodoefel; Christof Burgstahler; Ilias Tsiflikas; Anja Reimann; Stephen Schroeder; Claus D Claussen; Martin Heuschmid; Andreas F Kopp
Journal:  Radiology       Date:  2008-03-27       Impact factor: 11.105

Review 10.  Strategies for reduction of radiation dose in cardiac multislice CT.

Authors:  Jean-François Paul; Hicham T Abada
Journal:  Eur Radiol       Date:  2007-02-22       Impact factor: 7.034

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.