Literature DB >> 17310293

Noninvasive imaging of the coronary arteries using a 64-row multidetector CT scanner: initial clinical experience and radiation dose concerns.

M Francone1, A Napoli, I Carbone, M Cavacece, P G Nardis, K Lanciotti, S Visconti, L Bertoletti, E Di Castro, C Catalano, R Passariello.   

Abstract

PURPOSE: We present our initial clinical experience with a recently introduced 64-detector computed tomography (64-MDCT) scanner that makes use of a periodic motion of the focal spot in the longitudinal direction (z-flying focal spot), which enables it to reach a final spatial resolution of 0.4 x 0.4 x 0.4 mm(3) and a temporal resolution of 83 ms.
MATERIALS AND METHODS: A total of 114 patients (108 men, six women; age range 36-77 years, mean 63.1 years) underwent retrospective electrocardiogram (ECG)-gated examination of the coronary arteries using a 64-MDCT scanner (Somatom Sensation 64, Siemens Medical Solutions, Germany). Acquisition parameters were the following: collimation 64 x 0.6 mm, 800 quality reference milliampere second (mAs), 120 kVp, 0.33-s gantry rotation time and pitch 0.2. Images were acquired in all cases after i.v. administration of 80 ml of contrast agent (Iomeron 400 mgI/dl, Bracco, Italy) + 30 ml of saline at 4 /s and delay time determined using a bolus triggering technique. Oral betablockers were administered to patients with heart rate (HR) >75 bpm. To reduce radiation exposure, an automatic exposure control system was applied in all cases to adapt tube current to patient size and anatomic shape (CARE Dose 4D, Siemens Medical Solutions, Germany). The optimal temporal window for raw data reconstruction was chosen from an initial preview of images reconstructed with different phase settings (range 0%-95% RR interval with 5% gap) at a selected anatomical level in the mid part of the right coronary artery. CT dose index volume and effective dose were quantified in all patients using dedicated software.
RESULTS: Mean HR recorded during image acquisition was 65.6+/-19.2 bmp (range: 44-96 bmp), and beta-blockers were administered to 16/114 patients (14.0%). Technical adequacy was achieved in all patients but two (2/114; 1.7%). In patients with HR <60 bmp, the best reconstruction intervals were identified in the end-systolic (30%-35% of the RR interval) and end-diastolic (60%-65% of the RR interval) phases; with faster HR (>80 bmp), high image quality was observed in end-systole (30%-35% of the RR interval). Mean CT dose index (CTDI) volume was 36.53+/-8.30 mGy per patient. In comparison with a conventional examination with fixed mAs, the use of the CARE Dose 4D system provided a 33.3% CTDI volume reduction (p<0.001). Mean effective dose was 9.5+/-3.4 millisievert (mSv) per patient (range 7.1-17.7).
CONCLUSIONS: The 64-MDCT scanner diagnostic performance for coronary CT angiography is further improved with better spatial and temporal resolution and faster scan times; besides, initial clinical results are promising. The use of dose-reducing acquisition techniques is mandatory to limit radiation exposure to the patient.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17310293     DOI: 10.1007/s11547-007-0118-8

Source DB:  PubMed          Journal:  Radiol Med        ISSN: 0033-8362            Impact factor:   3.469


  12 in total

1.  Concept of minimal heart rate for each pitch value to avoid interpolation artifact when using dual-source CT: a phantom study.

Authors:  Joon-Won Kang; Kyung-Hyun Do; Jae-Yeon Chung; Hyun Jong Cho; Joon Beom Seo; Tae-Hwan Lim
Journal:  Int J Cardiovasc Imaging       Date:  2010-01-30       Impact factor: 2.357

Review 2.  Low dose CT of the heart: a quantum leap into a new era of cardiovascular imaging.

Authors:  E Maffei; C Martini; S De Crescenzo; T Arcadi; A Clemente; E Capuano; A Rossi; R Malagò; N Mollet; A Weustink; C Tedeschi; L La Grutta; S Seitun; A Igoren Guaricci; F Cademartiri
Journal:  Radiol Med       Date:  2010-06-23       Impact factor: 3.469

3.  Effect of kVp on image quality and accuracy in coronary CT angiography according to patient body size: a phantom study.

Authors:  Sang Min Lee; Whal Lee; Jin Wook Chung; Eun-Ah Park; Jae Hyung Park
Journal:  Int J Cardiovasc Imaging       Date:  2013-11-02       Impact factor: 2.357

4.  Dose reduction in spiral CT coronary angiography with dual source equipment. Part II. Dose surplus due to slope-up and slope-down of prospective tube current modulation in a phantom model.

Authors:  C Martini; A Palumbo; E Maffei; A Rossi; M Rengo; R Malagò; M Dijkshoorn; A Weustink; N Mollet; G Krestin; F Cademartiri
Journal:  Radiol Med       Date:  2010-02       Impact factor: 3.469

5.  Dose reduction in spiral CT coronary angiography with dual-source equipment. Part I. A phantom study applying different prospective tube current modulation algorithms.

Authors:  C Martini; A Palumbo; E Maffei; A Rossi; M Rengo; R Malagò; M Dijkshoorn; A Weustink; N Mollet; G Krestin; F Cademartiri
Journal:  Radiol Med       Date:  2009-08-07       Impact factor: 3.469

6.  Dual-source CT coronary angiography: prospective versus retrospective acquisition technique.

Authors:  C N De Cecco; V Buffa; S Fedeli; A Vallone; R Ruopoli; M Luzietti; V Miele; M Maurizi Enrici; F Musumeci; V David
Journal:  Radiol Med       Date:  2010-10-27       Impact factor: 3.469

7.  Assessment of coronary artery lesions in children with Kawasaki disease: evaluation of MSCT in comparison with 2-D echocardiography.

Authors:  Yanlin Xing; Hong Wang; Xianyi Yu; Rui Chen; Yang Hou
Journal:  Pediatr Radiol       Date:  2009-08-11

8.  Preoperative coronary risk assessment with dual-source CT in patients undergoing noncoronary cardiac surgery.

Authors:  V Buffa; C N De Cecco; L Cossu; S Fedeli; A Vallone; R Ruopoli; M Luzietti; G Angelica; V David; F Musumeci
Journal:  Radiol Med       Date:  2010-03-09       Impact factor: 3.469

9.  A mini pig model for visualization of perforator flap by using angiography and MIMICS.

Authors:  Jin Mei; ZhiXun Yin; Ji Zhang; Koonhei W Lui; Siwang Hu; Zhou Peng; Shixin Chen; Maolin Tang
Journal:  Surg Radiol Anat       Date:  2009-11-14       Impact factor: 1.246

10.  Optimal phase for coronary interpretations and correlation of ejection fraction using late-diastole and end-diastole imaging in cardiac computed tomography angiography: implications for prospective triggering.

Authors:  Hussain Isma'eel; Yasmin S Hamirani; Ramona Mehrinfar; Songshuo Mao; Naser Ahmadi; Vahid Larijani; Subu Nair; Matthew J Budoff
Journal:  Int J Cardiovasc Imaging       Date:  2009-07-25       Impact factor: 2.357

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.