Comparison of the percutaneous screw placement precision of isocentric C-arm 3-dimensional fluoroscopy-navigated pedicle screw implantation and conventional fluoroscopy method with minimally invasive surgery.

Hiroaki Nakashima1, Kouji Sato, Toshihiro Ando, Hidefumi Inoh, Hiroshi Nakamura.   

Abstract

STUDY
DESIGN: Retrospective clinical study to compare the percutaneous screw placement precision of isocentric C-arm (Iso-C) 3-dimensional fluoroscopy-navigated pedicle screw implantation and conventional fluoroscopy.
OBJECTIVE: The purpose of this study was to evaluate the accuracy of clinical percutaneous pedicle screw placement (PPSP) using Iso-C 3-dimensional fluoroscopy navigation. SUMMARY OF BACKGROUND DATA: The technique of PPSP recently has been applied to internal fixation for segmental lumbar instability as a form of minimally invasive spine surgery. The percutaneous insertion technique requires surgical skill and experience. The 3-dimensional fluoroscopy-based image-guidance navigation system has been proven to be effective in guiding accurate PPSP in the lumbar area of cadavers. However, there have been few clinical reports comparing the accuracy of Iso-C 3-dimensional fluoroscopy-navigated pedicle screw implantation and the conventional fluoroscopy method.
METHODS: Here, 300 percutaneous pedicle screws were evaluated; half of them were inserted under Iso-C 3-dimensional navigation and the others under fluoroscopy. The accuracy of pedicle screw placement was examined postoperatively using an axial 2-mm slice computed tomography. Screw malpositioning was classified either as exposed screw (<50% of the screw outside the pedicle) or perforated pedicle (>50% of the screw outside the pedicle boundaries).
RESULTS: Of the 150 pedicle screws placed with Iso-C 3-dimensional fluoroscopy-based image-guide assistance, 11 (7.3%) were classified as exposed screws and 0 (0%) as perforated pedicle. Of the 150 pedicle screws inserted under fluoroscopy, 18 (12%) were classified as exposed screw and 5 (3.3%) as perforated pedicle. The difference in frequency of screw misplacement between the procedure using Iso-C 3-dimensional fluoroscopy-based image-guide assistance and that using fluoroscopy was statistically significant (P<0.05; chi test).
CONCLUSIONS: This study demonstrates the feasibility of placing percutaneous posterior lumbar pedicle screws with the assistance of Iso-C 3-dimensional navigation.

Entities:  

Mesh:

Year:  2009        PMID: 20075808     DOI: 10.1097/BSD.0b013e31819877c8

Source DB:  PubMed          Journal:  J Spinal Disord Tech        ISSN: 1536-0652


  32 in total

Review 1.  Pedicle screw insertion accuracy with different assisted methods: a systematic review and meta-analysis of comparative studies.

Authors:  Nai-Feng Tian; Qi-Shan Huang; Ping Zhou; Yang Zhou; Rui-Kai Wu; Yi Lou; Hua-Zi Xu
Journal:  Eur Spine J       Date:  2010-09-23       Impact factor: 3.134

2.  Fluoroscopy-guided pedicle screw accuracy with a mini-open approach: a tomographic evaluation of 470 screws in 125 patients.

Authors:  José Antonio Soriano-Sánchez; Luis Alberto Ortega-Porcayo; Carlos Francisco Gutiérrez-Partida; Luis Rodolfo Ramírez-Barrios; Ramses Uriel Ortíz-Leyva; Manuel Rodríguez-García; Oscar Sánchez-Escandón
Journal:  Int J Spine Surg       Date:  2015-10-23

3.  Computer-Assisted Orthopedic and Trauma Surgery.

Authors:  Timo Stübig; Henning Windhagen; Christian Krettek; Max Ettinger
Journal:  Dtsch Arztebl Int       Date:  2020-11-20       Impact factor: 5.594

Review 4.  The evolution of image-guided lumbosacral spine surgery.

Authors:  Austin C Bourgeois; Austin R Faulkner; Alexander S Pasciak; Yong C Bradley
Journal:  Ann Transl Med       Date:  2015-04

5.  Accuracy of percutaneous pedicle screws for thoracic and lumbar spine fractures: a prospective trial.

Authors:  Timo Michael Heintel; Andreas Berglehner; Rainer Meffert
Journal:  Eur Spine J       Date:  2012-08-18       Impact factor: 3.134

6.  Percutaneous pedicle screw placements: accuracy and rates of cranial facet joint violation using conventional fluoroscopy compared with intraoperative three-dimensional computed tomography computer navigation.

Authors:  Tetsuro Ohba; Shigeto Ebata; Koji Fujita; Hironao Sato; Hirotaka Haro
Journal:  Eur Spine J       Date:  2016-03-08       Impact factor: 3.134

7.  Anterior single screw fixation of odontoid fracture with intraoperative Iso-C 3-dimensional imaging.

Authors:  Yong Liang Yang; Bai Sheng Fu; Rachel W Li; Paul N Smith; Wei Dong Mu; Lian Xin Li; Dong Sheng Zhou
Journal:  Eur Spine J       Date:  2011-06-04       Impact factor: 3.134

8.  Robot-assisted spine surgery: feasibility study through a prospective case-matched analysis.

Authors:  Nicolas Lonjon; Emilie Chan-Seng; Vincent Costalat; Benoit Bonnafoux; Matthieu Vassal; Julien Boetto
Journal:  Eur Spine J       Date:  2015-01-10       Impact factor: 3.134

9.  Inserting pedicle screws in lumbar spondylolisthesis - The easy bone conserving way.

Authors:  Hitesh Lal; Lalit Kumar; Ramesh Kumar; Tankeshwar Boruah; Pankaj Kumar Jindal; Vinod Kumar Sabharwal
Journal:  J Clin Orthop Trauma       Date:  2017-03-06

10.  Comparison of superior-level facet joint violations during open and percutaneous pedicle screw placement.

Authors:  Ranjith Babu; Jong G Park; Ankit I Mehta; Tony Shan; Peter M Grossi; Christopher R Brown; William J Richardson; Robert E Isaacs; Carlos A Bagley; Maragatha Kuchibhatla; Oren N Gottfried
Journal:  Neurosurgery       Date:  2012-11       Impact factor: 4.654

View more