| Literature DB >> 20072631 |
Gregory R Jackson, John G Edwards.
Abstract
Dark adaptometry may be a useful diagnostic test and clinical trial endpoint for age-related maculopathy (ARM) because impaired night vision is a hallmark of early ARM. A novel dark adaptometer, the AdaptDx, was evaluated for the detection of ARM. The AdaptDx incorporates a 20-minute protocol optimized for the detection of ARM. ARM patients (N = 17) exhibited substantial dark adaptation impairment compared with normal adults (N = 17). The diagnostic sensitivity was 88% and the specificity was 100%. The diagnostic test characteristics of the AdaptDx are similar to previously reported studies using 60- to 120-minute protocols.Entities:
Keywords: Age-related macular degeneration; Clinical trial endpoint; Dark adaptation; Diagnostic; Human; Macular degeneration
Year: 2008 PMID: 20072631 PMCID: PMC2802416 DOI: 10.1007/s12177-008-9002-6
Source DB: PubMed Journal: J Ocul Biol Dis Infor ISSN: 1936-8437
Fig. 1The patient’s view of the AdaptDx. The bleaching light (not shown) is presented through an aperture co-localized with the stimulus light
Participant demographics
| Young Normal ( | Old Normal ( | ARM ( | |
|---|---|---|---|
| Age (years), mean (SD) | 32.6 (5.3) | 73.1 (4.8) | 75.1 (6.4) |
| Gender, % ( | |||
| Female | 100 (8) | 66.7 (6) | 52.9 (9) |
| Male | 0 (0) | 33.3 (3) | 47.1 (8) |
| Race, % ( | |||
| White | 62.5 (5) | 88.9 (8) | 100 (17) |
| African American | 25.0 (2) | 11.1 (1) | 0 (0) |
| Asian | 12.5 (1) | 0 (0) | 0 (0) |
| Visual acuity, logMAR mean (SD) | |||
| Tested eye | −.02 (0.4) | .04 (.12) | .14 (.19) |
| Fellow eye | .05 (0.1) | .14 (.21) | .52 (.40) |
Fig. 2Representative dark adaptation curves of a normal adult (closed circles), early ARM patient (open triangles), and intermediate ARM patient (open circles)
Dark adaptation impairment increased with disease severity
| Young Normal | Old Normal | Borderline ARM | Early ARM | Intermediate ARM | Advanced ARM | |
|---|---|---|---|---|---|---|
| Sample Size | 8 | 9 | 5 | 6 | 5 | 1 |
| AREDS Fundus Grade | N/Aa | 1 | 2 | 3,4,5 | 6,7,8,9 | ≥10 |
| Rod Intercept, min mean (SD) | 8.2 (1.4) | 8.7 (1.9) | 13.8 (4.4) | 17.2 (2.7) | 20 (b) | 20 (b) |
aFundus photography and grading was not performed on young normal group.
bAll subjects failed to reach the rod intercept within the 20-minute test duration. Rod intercepts were set at 20 minutes for comparison purposes.