Label-free biomarker detection from whole blood.

Eric Stern1, Aleksandar Vacic, Nitin K Rajan, Jason M Criscione, Jason Park, Bojan R Ilic, David J Mooney, Mark A Reed, Tarek M Fahmy.   

Abstract

Label-free nanosensors can detect disease markers to provide point-of-care diagnosis that is low-cost, rapid, specific and sensitive. However, detecting these biomarkers in physiological fluid samples is difficult because of problems such as biofouling and non-specific binding, and the resulting need to use purified buffers greatly reduces the clinical relevance of these sensors. Here, we overcome this limitation by using distinct components within the sensor to perform purification and detection. A microfluidic purification chip simultaneously captures multiple biomarkers from blood samples and releases them, after washing, into purified buffer for sensing by a silicon nanoribbon detector. This two-stage approach isolates the detector from the complex environment of whole blood, and reduces its minimum required sensitivity by effectively pre-concentrating the biomarkers. We show specific and quantitative detection of two model cancer antigens from a 10 microl sample of whole blood in less than 20 min. This study marks the first use of label-free nanosensors with physiological solutions, positioning this technology for rapid translation to clinical settings.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 20010825      PMCID: PMC2818341          DOI: 10.1038/nnano.2009.353

Source DB:  PubMed          Journal:  Nat Nanotechnol        ISSN: 1748-3387            Impact factor:   39.213


Biomarkers have emerged as potentially important diagnostic tools for cancer and many other diseases. Continuing discoveries of such biomarkers and their aggregation into molecular signatures suggests that multiple biomarkers will be necessary to precisely define disease states. Thus, parallel detection of biomarker arrays is essential for translation from benchtop discovery to clinical validation. Such a technique would enable rapid, point-of-care (POC) applications requiring immediate diagnosis from a physiological sample. Critically, such a system must also be capable of detecting very low levels of aberrant genes and proteins, as many biomarkers are present at minute concentrations during early disease phases3-6. Given these requirements, the use of conventional diagnostic assays5,6,14 has been limiting. An approach that is based on rapid, label-free sensing technologies would be ideally suited for clinical applications6-13. Since their introduction in 2001 (Ref. 7), label-free nanosensors have demonstrated great potential to serve as POC detectors capable of ultrasensitive, real-time, multiplexed detection of multiple biomolecular species6,8-13. Despite their appeal, electronic nanosensors continue to be a challenge to implement because fundamental limitations render them incapable of sensing molecules in complex, physiologic solutions6,8-13. Biofouling and nonspecific binding readily degrade the minute active surface areas of such devices (< 0.1 μm2)15 and label-free sensing requires purified, precisely controlled buffers in order to enable measurements to be performed. In the case of nanowire field-effect transistor (FET) sensing, low salt (<∼1 mM) buffers are required to prevent screening of the charge-based electronic signal12,16. To overcome these limitations we have developed a novel in-line microfabricated device that operates upstream of the nanosensors to purify biomarkers of interest. This microfluidic purification chip (MPC) captures cancer biomarkers from physiologic solutions and, after washing, releases the antigens17 into a pure buffer suitable for sensing. The chip design increases nanosensor specificity to that of conventional sandwich assay techniques because it requires two antibodies to bind biomarkers in order for a positive signal to be produced18. Figure 1 schematically illustrates the operation of the MPC chip. The avidin-functionalized chip19 (Fig. 1a) is treated with antibodies to any number of specific biomarkers conjugated to biotinylated, photocleavable crosslinkers containing a specific 19-mer DNA sequence (Fig. 2a)20. The MPC geometry was chosen to optimize biomarker binding (Supplementary Fig. S1)14 and chips were fabricated from 4-inch silicon wafers in a one-step photolithographic process (Supplementary Fig. S2). Completed chips (Fig. 2b) were loaded into a custom-machined flow chamber (inset, Fig. 2b and Supplementary Fig. S3), which enabled fluid handling and maintained a constant 5 μL volume in the system.
Figure 1

Schematic of MPC operation

a, Primary antibodies to multiple biomarkers, here prostate specific antigen (PSA) and carbohydrate antigen-15.3 (CA15.3), are bound with a photocleavable crosslinker to the MPC. The chip is placed in a plastic housing and a valve (pink) directs fluid flow exiting the chip to either a waste receptacle or the nanosensor chip. b, Whole blood is injected into the chip with the valve set to the waste compartment (black arrow shows the direction of fluid flow) and, if present in the sample, biomarkers bind their cognate antibodies. c, Washing steps follow blood flow and the chip volume (5 μL) is filled with sensing buffer prior to UV irradiation (orange arrows). During UV exposure, the photolabile crosslinker cleaves, releasing the antibody-antigen complexes into solution. d, The valve is set to the nanosensor reservoir (black arrow shows the direction of fluid flow) and the 5 μL volume is transferred, enabling label-free sensing to be performed to determine the presence of specific biomarkers.

Figure 2

MPC operation

a, Molecular structure of the photocleavable crosslinker. Primary antibody conjugation was performed with the amino group (right) and binding to chip-bound avidin occurred through the biotin group (left). b, Scanning electron micrograph of a representative w = 4 mm × l = 7 mm × h = 100 μm MPC capture-release chip. The inset is an optical image of MPC operation during washing. c, Schematic representation of PSA and CA15.3 detection using a modified ELISA technique. d, Fluorescence optical micrograph of an anti-OVA functionalized MPC following OVA-FITC-spiked whole blood flow and washing. The inset plots the pixel intensity (gray value, determined by ImageJ) versus position for the red cutline (green dataplot) and similar cutlines from images of post-UV irradiation and transfer (blue) and of an anti-PSA functionalized MPC following OVA-FITC-spiked blood flow and washing. The same exposure times were used for all images. e, Scatter plot showing the concentration of PSA and f, CA15.3 released from the MPC versus the concentration of PSA and CA15.3 introduced in whole blood, respectively. Each datapoint represents the average of three separate MPC runs and error bars represent one standard deviation.

An example operation is illustrated in Fig. 1b-d. First, a blood sample flows through the chip (Fig. 1b) and the chip-bound antibodies bind specific soluble biomarkers, essentially purifying these molecules from whole blood. After this capture step, wash and sensing buffers are perfused through the device. Next, flow is halted and the sensing buffer-filled MPC chip is irradiated with UV light (Fig. 1c), resulting in cleavage of the photolabile group20-23 and release of the bound biomarker-antibody-DNA complexes. The UV photocleavage process was shown not to affect the immunoactivity of the biomarkers (Supplementary Fig. S4). The DNA component was critical for preliminary assay validation experiments (Fig. 2c). As shown in Fig. 1d, after a second valve switching step transfers MPC contents to the nanosensor chip, the complexes bind the secondary antibodies on the nanowire surfaces. The purification/sensing operation thus requires two specific antibody binding events for detection, a significant improvement in selectivity over previous label-free nanosensing schemes6-13. To demonstrate the effectiveness of the capture-release approach, we utilized a readily available fluorescently labeled antigen-antibody pair, fluorescently labeled chicken ovalbumin (OVA-FITC), and its antibody anti-OVA IgG. OVA-FITC was added to heparinized murine blood and flowed through an anti-OVA functionalized chip. After washing and flushing with sensing buffer, fluorescence imaging demonstrated specific OVA-FITC binding to chip-bound antibodies (Fig. 2d). A control chip, to which anti-prostate specific antigen (PSA) was bound, showed a negligible fluorescent signal (inset, Fig. 2d). After UV irradiation and subsequent flushing of the sensing reservoir with fresh buffer, the fluorescence signal from the anti-OVA chip was greatly diminished (inset, Fig. 2d). To demonstrate the generality of MPC technique we used two model cancer antigens, PSA and carbohydrate antigen 15.3 (CA15.3), standard clinical markers for prostate24-25 and breast cancer26-27, respectively. Successful capture and release of PSA and CA15.3 was verified with a modified enzyme-linked immunoassay (ELISA) technique (Fig. 2c)18, in which the first detection step consisted of the hybridization of a complementary, biotinylated 19-mer to the crosslinker DNA sequence. Six increasing concentrations of PSA and CA15.3 were added to heparinized rat blood and samples were flowed through MPCs functionalized with both anti-PSA and anti-CA15.3. The introduced concentrations spanned clinically relevant ranges24-27. The data in Fig. 2e-f demonstrate a monotonic relationship between the concentration of biomarker introduced in whole blood and that released into pure sensing buffer. The absolute yields of these experiments are in agreement with modeling studies (Supplementary Fig. S1c). Biomarker capture by MPCs can be significantly increased by adjusting either the operation conditions such as the flow rate into the device (modeled in Supplementary Fig. S1d) or the device dimensions. A critical feature of this integrated approach is that the MPC-purified biomarker complex concentrations are well above those required for label-free, electronic detection. Although previous studies using nanowire sensors have demonstrated PSA detection as low as 0.9 pg/mL (Refs. 6,10), this exquisite sensitivity is not a critical factor for MPC-nanosensor operation. We thus chose to use “nanoribbons,” devices with nanoscale thicknesses and microscale lateral dimensions28, which are less sensitive but have significant fabrication and cost advantages. These devices, fabricated using conventional lithographic techniques, have been demonstrated to detect streptavidin in the 0.0318–53 ng/mL range28, a sensitivity range ideally suited for MPC-purified cancer antigen detection. We fabricated 25 nm-thin devices according to a similar process (Supplementary Fig. S5)28, but incorporated Ohmic contacts to devices. Device images are given in Fig. 3a. Electrical characterization verified that this approach produced high-quality devices, with on/off ratios of >106 (Fig. 3b) and small hystereses between forward and reverse IDS(VG) sweeps (Fig. 3c). Surface functionalization did not compromise device electrical characteristics (Supplementary Fig. S7) and solution gating (VG, SOLN) demonstrated that VG = -5 V was an optimal operating point for sensing studies (Fig. 3d).
Figure 3

Nanosensor electrical characteristics

a, Optical image of devices outfitted with sensing reservoirs. The inset shows an optical micrograph of a completed device. Only the central region of the device (black arrow) is exposed to the solution. Metal leads contact the device source and drain and fan out to larger contacts (not shown). The 25 nm thick silicon device appears yellow. b, IDS(VDS) plot for VG varied from 0 to -20V (black arrow shows direction of increasing negative VG) for a representative device illustrating p-type accumulation mode behavior. c, IDS(VG) plot (VDS = 1V) for the device used in b. The inset highlights IDS (nA) around the operating point (VG = -5V). d, Plot demonstrating the effect of varying solution gate voltage (VG, SOLN) on device current (IDS; black solid) and device-to-solution leakage current (ILEAK; blue dashed) for VDS = 1V.

As shown in Fig. 2 and detailed in the Supplementary Information, devices are functionalized either with anti-PSA or anti-CA15.3. Antibodies were immobilized to the sensor using NHS/EDC chemistry. To verify that the signal from binding proteins would not be screened by the buffer solution, direct measurements of the amount of the signal that would be unscreened were done by varying buffer salt concentration16. This study indicated that approximately 50% of the signal was not screened by the buffer solution (see Supplementary Information). Next, we applied these devices to sensing the biomarkers from the MPC-purified whole blood samples. The normalized responses of these same devices to MPC-purified, antigen-spiked blood samples containing both 2.5 ng/mL PSA and 30 U/mL CA15.3 (as well as negative controls) are shown in Fig. 4a and 4b, respectively. After the injection transient noise subsides11, device current levels were increased by antigen binding due to the negative charge conferred to the antigens by the basic sensing buffer. Similar signals were obtained with a PSA/CA15.3 spiked sensing buffer positive control, and no device response was observed with an unspiked, MPC-purified blood negative control. To reduce potential transient electrical signals upon injection, buffer salt concentrations of the functionalized devices and the MPC-purifies samples are kept the same. The positive signal is observed to increase linearly in time, following well-known ligand-receptor kinetics,29 in which initial rates at low relative analyte concentrations are directly proportional to the species concentration30. In fact, the asymptotic saturation value of the device response is weakly dependent on the concentration for reversible reactions with a low dissociation constant29 which is the case for the antigen-antibody interactions. Thus, we focus on the initial kinetic reaction rates instead of endpoint detection30.
Figure 4

Label-free sensing

All sensing measurements were performed at VDS = 1V and VG = -5V and all sample introductions occurred at time = 0. Normalizations were performed by dividing device currents by the pre-addition (time < 0) current level average. a, Response of an anti-PSA functionalized sensor to a MPC-purified blood sample initially containing 2.5 ng/mL PSA (and also 30 U/mL CA15.3), or a control sample containing neither. b, Response of an anti-CA15.3 functionalized sensor to a MPC-purified blood sample blood sample initially containing 30 U/mL CA15.3 (and also 2.5 ng/mL PSA), or a control sample containing neither. c, Normalized response of two anti-PSA and d, two anti-CA15.3 functionalized devices to MPC-purified blood containing both PSA and CA 15.3, with concentrations labeled. A least squares fit is represented by a solid black line, over the selected region (line endpoints). The ratio of the the normalized slopes calibrates the ratio of concentrations.

Using these rates, a quantification of analyte concentrations (against a known) can be made, as shown in Fig. 4c and 4d. Whole blood samples spiked with 2 ng/mL PSA and 15 U/mL CA15.3 were MPC purified and sensed with anti-PSA and anti-CA15.3 functionalized devices. Using the slope of the normalized device temporal response, we find slope ratios of both the PSA and CA15.3 responses agree quite well with the initial spiked whole biomarker concentrations. For PSA, the slope ratio is 1.38, compared with a concentration ratio of 1.25; for CA15.3, the slope ratio is 1.94, versus a concentration ratio of 2.0. It should be noted that this quantification occurs in the presence of another species, demonstrating selectivity as well (See Supplementary Information for further repeatability data). The integration of a microfluidic purification step with label-free nanosensor detection represents a paradigm shift in label-free electronic sensing of biomolecules. The technique described here enables biomarker detection from whole blood or any other physiologic fluid without the challenges associated with tailoring sensor operation for the medium of interest or engineering nanosensors that can withstand complex fluid media. Additionally, the need for ultrasensitivity in electronic detection may not be essential with such an integrated platform because of its ability to pre-concentrate molecules of choice prior to sensing. The attractiveness of the method lies in its simplicity, speed, and ability to simultaneously capture multiple biomarkers, enabling multiplexed, highly sensitive downstream detection with label-free sensors. The proof of principle demonstration of the non-integrated individual components here should be easily to integrate into a compact, self-contained system. Furthermore, the low cost of MPC purification renders this system capable of stand-alone use or use in tandem with more expensive sensing methodologies, such as rare circulating tumor cell detectors14 for more complex diagnoses. The portability and versatility of this method represents the crucial next step for label-free sensors and should position these and similar nascent sensing technologies for rapid molecular signature determinations.
  25 in total

1.  Genomic medicine and the future of health care.

Authors:  C Sander
Journal:  Science       Date:  2000-03-17       Impact factor: 47.728

Review 2.  The case for early detection.

Authors:  Ruth Etzioni; Nicole Urban; Scott Ramsey; Martin McIntosh; Stephen Schwartz; Brian Reid; Jerald Radich; Garnet Anderson; Leland Hartwell
Journal:  Nat Rev Cancer       Date:  2003-04       Impact factor: 60.716

Review 3.  Present and future of surface plasmon resonance biosensors.

Authors:  Jirí Homola
Journal:  Anal Bioanal Chem       Date:  2003-07-19       Impact factor: 4.142

Review 4.  Protein microarrays and multiplexed sandwich immunoassays: what beats the beads?

Authors:  Markus F Templin; Dieter Stoll; Jutta Bachmann; Thomas O Joos
Journal:  Comb Chem High Throughput Screen       Date:  2004-05       Impact factor: 1.339

Review 5.  Nanotechnology in clinical laboratory diagnostics.

Authors:  Kewal K Jain
Journal:  Clin Chim Acta       Date:  2005-08       Impact factor: 3.786

6.  Quantitative proteome analysis by solid-phase isotope tagging and mass spectrometry.

Authors:  Huilin Zhou; Jeffrey A Ranish; Julian D Watts; Ruedi Aebersold
Journal:  Nat Biotechnol       Date:  2002-05       Impact factor: 54.908

7.  Design and synthesis of a photocleavable biotinylated nucleotide for DNA analysis by mass spectrometry.

Authors:  Xiaopeng Bai; Sobin Kim; Zengmin Li; Nicholas J Turro; Jingyue Ju
Journal:  Nucleic Acids Res       Date:  2004-01-26       Impact factor: 16.971

Review 8.  Systematic review of pretreatment prostate-specific antigen velocity and doubling time as predictors for prostate cancer.

Authors:  Andrew J Vickers; Caroline Savage; M Frank O'Brien; Hans Lilja
Journal:  J Clin Oncol       Date:  2008-12-08       Impact factor: 44.544

9.  Photocleavable biotin derivatives: a versatile approach for the isolation of biomolecules.

Authors:  J Olejnik; S Sonar; E Krzymañska-Olejnik; K J Rothschild
Journal:  Proc Natl Acad Sci U S A       Date:  1995-08-01       Impact factor: 11.205

Review 10.  Serum CA 15.3, CEA and ESR patterns in breast cancer.

Authors:  M Rubach; J J Szymendera; J Kamińska; M Kowalska
Journal:  Int J Biol Markers       Date:  1997 Oct-Dec       Impact factor: 3.248

View more
  96 in total

Review 1.  Inorganic Complexes and Metal-Based Nanomaterials for Infectious Disease Diagnostics.

Authors:  Christine F Markwalter; Andrew G Kantor; Carson P Moore; Kelly A Richardson; David W Wright
Journal:  Chem Rev       Date:  2018-12-04       Impact factor: 60.622

Review 2.  Molecular analysis of blood with micro-/nanoscale field-effect-transistor biosensors.

Authors:  Matthew S Makowski; Albena Ivanisevic
Journal:  Small       Date:  2011-06-03       Impact factor: 13.281

Review 3.  Measurement of biomarker proteins for point-of-care early detection and monitoring of cancer.

Authors:  James F Rusling; Challa V Kumar; J Silvio Gutkind; Vyomesh Patel
Journal:  Analyst       Date:  2010-07-08       Impact factor: 4.616

Review 4.  Nano-Bioelectronics.

Authors:  Anqi Zhang; Charles M Lieber
Journal:  Chem Rev       Date:  2015-12-21       Impact factor: 60.622

Review 5.  Nanomaterials-based electrochemical immunosensors for proteins.

Authors:  James F Rusling
Journal:  Chem Rec       Date:  2012-01-30       Impact factor: 6.771

6.  Biofunctionalization on alkylated silicon substrate surfaces via "click" chemistry.

Authors:  Guoting Qin; Catherine Santos; Wen Zhang; Yan Li; Amit Kumar; Uriel J Erasquin; Kai Liu; Pavel Muradov; Barbara Wells Trautner; Chengzhi Cai
Journal:  J Am Chem Soc       Date:  2010-10-29       Impact factor: 15.419

7.  General strategy for biodetection in high ionic strength solutions using transistor-based nanoelectronic sensors.

Authors:  Ning Gao; Wei Zhou; Xiaocheng Jiang; Guosong Hong; Tian-Ming Fu; Charles M Lieber
Journal:  Nano Lett       Date:  2015-02-16       Impact factor: 11.189

8.  Analysis of circulating non-coding RNAs in a non-invasive and cost-effective manner.

Authors:  Yu-Min Wang; Michael Patrick Trinh; Yongzan Zheng; Kaizhu Guo; Luis A Jimenez; Wenwan Zhong
Journal:  Trends Analyt Chem       Date:  2019-07-05       Impact factor: 12.296

Review 9.  Point-of-care technologies for molecular diagnostics using a drop of blood.

Authors:  Yujun Song; Yu-Yen Huang; Xuewu Liu; Xiaojing Zhang; Mauro Ferrari; Lidong Qin
Journal:  Trends Biotechnol       Date:  2014-02-11       Impact factor: 19.536

10.  Effect of biointerfacing linker chemistries on the sensitivity of silicon nanowires for protein detection.

Authors:  Brian Dorvel; Bobby Reddy; Rashid Bashir
Journal:  Anal Chem       Date:  2013-10-03       Impact factor: 6.986

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.