Comparison of anterior chamber depth measurements using Galilei, HR Pentacam, and Orbscan II.

Ramin Salouti1, Mohammad H Nowroozzadeh, Mohammad Zamani, Maryam Ghoreyshi, Rosa Salouti.   

Abstract

BACKGROUND: Anterior chamber depth measurement is used frequently in cataract and refractive surgery for advanced intraocular lens (IOL) power calculation formulas and implantation of phakic IOLs. In this article, we show agreement in anterior chamber depth measurements in normal eyes among 3 noncontact systems, namely Galilei, HR Pentacam, and Orbscan II.
METHODS: The anterior chamber depth measurements were obtained with Galilei, HR Pentacam, and the Orbscan II in both eyes of 37 healthy subjects. Pearson's correlations were calculated, and mean anterior chamber depth measurements were compared. Bland-Altman plots were used to assess the difference between individual measurements for each patient.
RESULTS: The mean anterior chamber depth readings (+/- standard error of means) with Galilei, Pentacam, and Orbscan were 3.22 +/- 0.05, 3.25 +/- 0.05, and 3.54 +/- 0.07, respectively. The difference between Galilei and Pentacam was not statistically significant (P = 0.013). However, Orbscan measurements were larger than those of either Galilei or Pentacam (P < 0.001). The differences between Orbscan and either Galilei or Pentacam were increased with greater anterior chamber depths. The 95% limits of agreement between Orbscan and Galilei, Orbscan and Pentacam, and Pentacam and Galilei in different ranges of chamber depth were ([ < 3 mm, -0.43 to +0.71 mm]; [3 to 3.5 mm, -0.23 to +0.83]; [ > 3.5 mm, +0.41 to +0.53]), ([ < 3 mm, -0.43 to +0.71 mm]; [3 to 3.5 mm, -0.26 to +0.80]; [ > 3.5 mm, +0.38 to +0.54]), and ([ < 3 mm, -0.05 to +0.06 mm]; [3 to 3.5 mm, -0.06 to +0.10]; [ > 3.5 mm, -0.11 to +0.12]), respectively.
CONCLUSIONS: These data indicate that in the assessment of normal eyes, Orbscan gives consistently higher measurements for anterior chamber depth compared with Galilei and Pentacam. Because the differences between Orbscan and either Galilei or Pentacam were not within clinically acceptable levels, they are not interchangeable in every clinical situation. However, the differences among anterior chamber depth values measured with Galilei and Pentacam were within clinically acceptable levels. Thus, these 2 systems can be regarded as interchangeable.

Entities:  

Mesh:

Year:  2010        PMID: 20004876     DOI: 10.1016/j.optm.2009.04.100

Source DB:  PubMed          Journal:  Optometry        ISSN: 1558-1527


  13 in total

1.  Anterior chamber depth in normal subjects by rotating scheimpflug imaging.

Authors:  Matthew T Feng; Michael W Belin; Renato Ambrósio; Satinder P S Grewal; Wang Yan; Mohamed S Shaheen; Charles McGhee; Naoyuki Maeda; Tobias H Neuhann; H Burkhard Dick; Saleh A Alageel; Andreas Steinmueller
Journal:  Saudi J Ophthalmol       Date:  2011-04-30

2.  Difference in normal corneal thickness and curvature between Mongolian and Han nationalities.

Authors:  Hai-Xia Zhao; Li Zhang; Wen-Ying Guan
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

3.  Repeatability and comparability of anterior segment biometry obtained by the Sirius and the Pentacam analyzers.

Authors:  Paola De la Parra-Colín; Manuel Garza-León; Tonatiuh Barrientos-Gutierrez
Journal:  Int Ophthalmol       Date:  2013-04-25       Impact factor: 2.031

Review 4.  Device interchangeability on anterior chamber depth and white-to-white measurements: a thorough literature review.

Authors:  Alberto Domínguez-Vicent; Cari Pérez-Vives; Teresa Ferrer-Blasco; Santiago García-Lázaro; Robert Montés-Micó
Journal:  Int J Ophthalmol       Date:  2016-07-18       Impact factor: 1.779

Review 5.  Clinically relevant biometry.

Authors:  Afsun Sahin; Pedram Hamrah
Journal:  Curr Opin Ophthalmol       Date:  2012-01       Impact factor: 3.761

6.  Comparison of Anterior Chamber Depth Measurements from the Galilei Dual Scheimpflug Analyzer with IOLMaster.

Authors:  Roma P Patel; Rahul T Pandit
Journal:  J Ophthalmol       Date:  2012-01-19       Impact factor: 1.909

7.  Repeatability of ocular measurements with a dual-Scheimpflug analyzer in healthy eyes.

Authors:  Carmen Lopez de la Fuente; Ana Sanchez-Cano; Francisco Segura; Lorena Fuentes-Broto; Isabel Pinilla
Journal:  Biomed Res Int       Date:  2014-04-24       Impact factor: 3.411

8.  Comparison of anterior segment measurements obtained by three different devices in healthy eyes.

Authors:  Carmen Lopez de la Fuente; Ana Sanchez-Cano; Francisco Segura; Isabel Pinilla
Journal:  Biomed Res Int       Date:  2014-06-02       Impact factor: 3.411

9.  Comparison of Anterior Segment Measurements Obtained by Aladdin Optical Biometer and Sirius Corneal Topography.

Authors:  Onur Polat; Zeki Baysal; Serkan Özcan; Sibel İnan; Ümit Übeyt İnan
Journal:  Turk J Ophthalmol       Date:  2016-12-01

10.  Comparison between Two Scheimpflug Anterior Segment Analyzers.

Authors:  Alireza Baradaran-Rafii; Tahmineh Motevasseli; Forouzan Yazdizadeh; Farid Karimian; Sahba Fekri; Amir Baradaran-Rafii
Journal:  J Ophthalmic Vis Res       Date:  2017 Jan-Mar
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.