A comparison of the performance of 20 pulse oximeters under conditions of poor perfusion.

D G Clayton1, R K Webb, A C Ralston, D Duthie, W B Runciman.   

Abstract

The performance of 20 pulse oximeters with finger probes was evaluated by comparison of their readings with directly measured arterial blood oxygen saturations. The samples were taken from patients who had undergone cardiac surgery under hypothermic cardiopulmonary bypass and had poor peripheral perfusion. The mean difference (bias, accuracy), standard deviation (precision) and drop-out rate for each pulse oximeter was determined. An overall ranking of performance of each pulse oximeter was calculated using five criteria (accuracy, precision, number of readings within 3% of standard, percentage of readings given within 3% of standard, expected overread limit in 95% of cases). Two pulse oximeters achieved a combination of accuracy and precision such that 95% of measurements would be expected to be within 4% of the co-oximeter value; these two also had the lowest drop-out rate.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1996749     DOI: 10.1111/j.1365-2044.1991.tb09303.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  10 in total

Review 1.  Noninvasive blood gas monitoring: a review for use in the adult critical care unit. Technology Subcommittee of the Working Group on Critical Care, Ontario Ministry of Health.

Authors: 
Journal:  CMAJ       Date:  1992-03-01       Impact factor: 8.262

2.  Analysis of the ear pulse oximeter waveform.

Authors:  Aymen A Awad; Robert G Stout; M Ashraf M Ghobashy; Hoda A Rezkanna; David G Silverman; Kirk H Shelley
Journal:  J Clin Monit Comput       Date:  2006-04-13       Impact factor: 2.502

Review 3.  Pulse oximetry: a practical review.

Authors:  C D Hanning; J M Alexander-Williams
Journal:  BMJ       Date:  1995-08-05

4.  Sleep disordered breathing does not predict acute severe pain episodes in children with sickle cell anemia.

Authors:  Shaina M Willen; Mark Rodeghier; Carol L Rosen; Michael R DeBaun
Journal:  Am J Hematol       Date:  2018-01-25       Impact factor: 10.047

5.  Pulse oximeter signal at various blood flow conditions in an in vitro model.

Authors:  L G Lindberg; M Vegfors; C Lennmarken; P A Oberg
Journal:  Med Biol Eng Comput       Date:  1995-01       Impact factor: 2.602

6.  Misleading information from pulse oximetry and the usefulness of continuous blood gas monitoring in a post cardiac surgery patient.

Authors:  J P Benson; B Venkatesh; V Patla
Journal:  Intensive Care Med       Date:  1995-05       Impact factor: 17.440

7.  A system for investigating oesophageal photoplethysmographic signals in anaesthetised patients.

Authors:  P A Kyriacou; A R Moye; A Gregg; D M Choi; R M Langford; D P Jones
Journal:  Med Biol Eng Comput       Date:  1999-09       Impact factor: 2.602

8.  Failure rate of pulse oximetry in the postanesthesia care unit.

Authors:  B S Gillies; K Posner; P Freund; F Cheney
Journal:  J Clin Monit       Date:  1993-11

9.  Accuracy of pulse oximetry and capnography in healthy and compromised horses during spontaneous and controlled ventilation.

Authors:  Judith Koenig; Wayne McDonell; Alexander Valverde
Journal:  Can J Vet Res       Date:  2003-07       Impact factor: 1.310

10.  Pulse oximetry.

Authors: 
Journal:  Crit Care       Date:  1999       Impact factor: 9.097

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.