Comparative study of two image space noise reduction methods for computed tomography: bilateral filter and nonlocal means.

Juan C Ramirez Giraldo1, Zachary S Kelm, Luis S Guimaraes, Lifeng Yu, Joel G Fletcher, Bradley J Erickson, Cynthia H McCollough.   

Abstract

Optimal noise control is important for improving image quality and reducing radiation dose in computed tomography. Here we investigated two image space based nonlinear filters for noise reduction: the bilateral filter (BF) and the nonlocal means (NLM) algorithm. Images from both methods were compared against those from a commercially available weighted filtered backprojection (WFBP) method. A standard phantom for quality assurance testing was used to quantitatively compare noise and spatial resolution, as well as low contrast detectability (LCD). Additionally, an image dataset from a patient's abdominal CT exam was used to assess the effectiveness of the filters on full dose and simulated half dose acquisitions. We found that both the BF and NLM methods improve the tradeoff between noise and high contrast spatial resolution with no significant difference in LCD. Results from the patient dataset demonstrated the potential of dose reduction with the denoising methods. Care must be taken when choosing the NLM parameters in order to minimize the generation of artifacts that could possibly compromise diagnostic value.

Entities:  

Mesh:

Year:  2009        PMID: 19964998     DOI: 10.1109/IEMBS.2009.5334714

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X


  7 in total

1.  Statistical image reconstruction for low-dose CT using nonlocal means-based regularization. Part II: An adaptive approach.

Authors:  Hao Zhang; Jianhua Ma; Jing Wang; Yan Liu; Hao Han; Hongbing Lu; William Moore; Zhengrong Liang
Journal:  Comput Med Imaging Graph       Date:  2015-03-06       Impact factor: 4.790

2.  Assessment of bilateral filter on 1/2-dose chest-pelvis CT views.

Authors:  Abdel Razzak Al-Hinnawi; Mohammed Daear; Said Huwaijah
Journal:  Radiol Phys Technol       Date:  2013-04-20

Review 3.  Applications of nonlocal means algorithm in low-dose X-ray CT image processing and reconstruction: A review.

Authors:  Hao Zhang; Dong Zeng; Hua Zhang; Jing Wang; Zhengrong Liang; Jianhua Ma
Journal:  Med Phys       Date:  2017-03       Impact factor: 4.071

4.  Trainable joint bilateral filters for enhanced prediction stability in low-dose CT.

Authors:  Fabian Wagner; Mareike Thies; Felix Denzinger; Mingxuan Gu; Mayank Patwari; Stefan Ploner; Noah Maul; Laura Pfaff; Yixing Huang; Andreas Maier
Journal:  Sci Rep       Date:  2022-10-20       Impact factor: 4.996

5.  Statistical image reconstruction for low-dose CT using nonlocal means-based regularization.

Authors:  Hao Zhang; Jianhua Ma; Jing Wang; Yan Liu; Hongbing Lu; Zhengrong Liang
Journal:  Comput Med Imaging Graph       Date:  2014-05-14       Impact factor: 4.790

6.  Investigation of Low-Dose CT Image Denoising Using Unpaired Deep Learning Methods.

Authors:  Zeheng Li; Shiwei Zhou; Junzhou Huang; Lifeng Yu; Mingwu Jin
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2020-07-07

7.  Improving spatial adaptivity of nonlocal means in low-dosed CT imaging using pointwise fractal dimension.

Authors:  Xiuqing Zheng; Zhiwu Liao; Shaoxiang Hu; Ming Li; Jiliu Zhou
Journal:  Comput Math Methods Med       Date:  2013-03-31       Impact factor: 2.238

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.