Literature DB >> 19952641

Comparative healing of surgical incisions created by the PEAK PlasmaBlade, conventional electrosurgery, and a scalpel.

Shang A Loh1, Grace A Carlson, Edward I Chang, Eric Huang, Daniel Palanker, Geoffrey C Gurtner.   

Abstract

BACKGROUND: The PEAK PlasmaBlade is a new electrosurgical device that uses pulsed radiofrequency to generate a plasma-mediated discharge along the exposed rim of an insulated blade, creating an effective cutting edge while the blade stays near body temperature.
METHODS: Full-thickness incisions were made on the dorsums of pigs with the PlasmaBlade, a conventional electrosurgical device, and a scalpel, and blood loss was quantified. Wounds were harvested at designated time points, tested for wound tensile strength, and examined histologically for scar formation and tissue damage.
RESULTS: Bleeding was reduced significantly (59 percent) in PlasmaBlade incisions compared with scalpel incisions, and acute thermal damage from the PlasmaBlade (66 +/- 5 microm) was significantly less than both cut and coagulation mode electrosurgical incisions (456 +/- 35 microm and 615 +/- 22 microm, respectively). Histologic scoring for injury and wound strength was equivalent between the PlasmaBlade and scalpel incisions. By 6 weeks, the healed PlasmaBlade and scalpel incisions were approximately three times stronger, and scar cosmetic appearance was significantly better compared with electrosurgical incisions.
CONCLUSIONS: The PlasmaBlade is a promising new surgical instrument that provides atraumatic, scalpel-like cutting precision and electrosurgical-like hemostasis, resulting in minimal bleeding, tissue injury, and scar formation.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19952641     DOI: 10.1097/PRS.0b013e3181bcee87

Source DB:  PubMed          Journal:  Plast Reconstr Surg        ISSN: 0032-1052            Impact factor:   4.730


  34 in total

1.  Comparative analysis on the effect of low-thermal plasma dissection device (PEAK PlasmaBlade) versus conventional electro surgery in post-bariatric body-contouring procedures: A retrospective randomised clinical study.

Authors:  Torsten Schlosshauer; Marcus Kiehlmann; Pablo Ramirez; Marc-Oliver Riener; Gabriel Djedovic; Ulrich M Rieger
Journal:  Int Wound J       Date:  2019-04-01       Impact factor: 3.315

2.  Comparison of the effects of surgical dissection devices on the rabbit liver.

Authors:  Joel D MacDonald; Christian A Bowers; Steven S Chin; Greg Burns
Journal:  Surg Today       Date:  2013-09-05       Impact factor: 2.549

3.  Bilateral reduction mammaplasty with pulsed electron avalanche knife PlasmaBlade™ and conventional electrosurgical surgery: A retrospective, randomised controlled clinical trial.

Authors:  Torsten Schlosshauer; Marcus Kiehlmann; Jens Rothenberger; Robert Sader; Ulrich M Rieger
Journal:  Int Wound J       Date:  2020-07-09       Impact factor: 3.315

4.  Effect of low-thermal dissection device versus conventional electrocautery in mastectomy for female-to-male transgender patients.

Authors:  Torsten Schlosshauer; Marcus Kiehlmann; Marc-Oliver Riener; Jens Rothenberger; Robert Sader; Ulrich M Rieger
Journal:  Int Wound J       Date:  2020-05-05       Impact factor: 3.315

5.  Comparative analysis on the effect of low-thermal plasma dissection device (PEAK PlasmaBlade) vs conventional electrosurgery in post-bariatric abdominoplasty: A retrospective randomised clinical study.

Authors:  Torsten Schlosshauer; Marcus Kiehlmann; Marc-Oliver Riener; Robert Sader; Ulrich M Rieger
Journal:  Int Wound J       Date:  2019-09-17       Impact factor: 3.315

6.  Incidence and characteristics of vaginal cuff dehiscence in robotic-assisted and traditional total laparoscopic hysterectomy.

Authors:  Erin Dauterive; George Morris
Journal:  J Robot Surg       Date:  2011-09-13

Review 7.  Scalpel versus electrosurgery for major abdominal incisions.

Authors:  Kittipat Charoenkwan; Zipporah Iheozor-Ejiofor; Kittipan Rerkasem; Elizabeth Matovinovic
Journal:  Cochrane Database Syst Rev       Date:  2017-06-14

8.  Plasmakinetic vaporization versus plasmakinetic resection to treat benign prostatic hyperplasia: A prospective randomized trial with 1 year follow-up.

Authors:  Mert Ali Karadag; Kursat Cecen; Aslan Demir; Ramazan Kocaaslan; Fatih Altunrende
Journal:  Can Urol Assoc J       Date:  2014-09       Impact factor: 1.862

9.  An electrical plasma dissection tool for surgical treatment of chronic ulcers: Results of a prospective randomised trial.

Authors:  Giovanni F Marangi; Tiziano Pallara; Daniela Lamberti; Eleonora Perrella; Raffaele Serra; Francesco Stilo; Giovanni De Caridi; Andrea Onetti Muda; Paolo Persichetti
Journal:  Int Wound J       Date:  2018-03-30       Impact factor: 3.315

10.  The use of plasmakinetic cautery compared to conventional electrocautery for dissection of abdominal free flap for breast reconstruction: single-centre, randomized controlled study.

Authors:  Whitney T H Chow; Georgette Oni; Venkat V Ramakrishnan; Mat Griffiths
Journal:  Gland Surg       Date:  2019-06
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.