In vitro evaluation of a ball-and-socket cervical disc prosthesis with cranial geometric center.

Cédric Barrey1, Thomas Mosnier, Jérôme Jund, Gilles Perrin, Wafa Skalli.   

Abstract

OBJECT: Few biomechanical in vitro studies have reported the effects of disc replacement on motion and kinematics of the cervical spine. The purpose of this study was to analyze motion through 3D load-displacement curves before and after implantation of a ball-and-socket cervical disc prosthesis with cranial geometric center; special focus was placed on coupled motion, which is a well-known aspect of normal cervical spine kinematics.
METHODS: Six human cervical spines were studied. There were 3 male and 3 female cadaveric specimens (mean age at death 68.5 +/- 5 years [range 54-74 years]). The specimens were evaluated sequentially in 2 different conditions: first they were tested intact; then the spinal specimens were tested after implantation of a ball-and-socket cervical disc prosthesis, the Discocerv, at the C5-6 level. Pure moment loading was applied in flexion/extension, left and right axial rotation, and left and right lateral bending. All tests were performed under load control with a 3D measurement system.
RESULTS: No differences were found to be statistically significant after comparison of range of motion between intact and instrumented spines for all loading conditions. The mean range of motion for intact spines was 10.3 degrees in flexion/extension, 5.6 degrees in lateral bending, and 5.4 degrees in axial rotation; that for instrumented spines was 10.4, 5.2, and 4.8 degrees , respectively. No statistical difference was observed for the neutral zone nor stiffness between intact and instrumented spines. Finally, the coupled motions were also preserved during axial rotation and lateral bending, with no significant difference before and after implantation.
CONCLUSIONS: This study demonstrated that, under specific testing conditions, a ball-and-socket joint with cranial geometrical center can restore motion in the 3 planes after discectomy in the cervical spine while maintaining physiological coupled motions during axial rotation and lateral bending.

Entities:  

Mesh:

Year:  2009        PMID: 19929355     DOI: 10.3171/2009.6.SPINE0949

Source DB:  PubMed          Journal:  J Neurosurg Spine        ISSN: 1547-5646


  6 in total

1.  Biomechanical effects of cervical arthroplasty with U-shaped disc implant on segmental range of motion and loading of surrounding soft tissue.

Authors:  Zhong Jun Mo; Yan Bin Zhao; Li Zhen Wang; Yu Sun; Ming Zhang; Yu Bo Fan
Journal:  Eur Spine J       Date:  2013-10-24       Impact factor: 3.134

2.  Cervical disc prosthesis versus arthrodesis using one-level, hybrid and two-level constructs: an in vitro investigation.

Authors:  Cédric Barrey; Sophie Campana; Sylvain Persohn; Gilles Perrin; Wafa Skalli
Journal:  Eur Spine J       Date:  2011-08-11       Impact factor: 3.134

3.  Relevance of using a compressive preload in the cervical spine: an experimental and numerical simulating investigation.

Authors:  Cédric Barrey; Marc-Antoine Rousseau; Sylvain Persohn; Sophie Campana; Gilles Perrin; Wafa Skalli
Journal:  Eur J Orthop Surg Traumatol       Date:  2015-04-07

4.  Sagittal alignment and kinematics at instrumented and adjacent levels after total disc replacement in the cervical spine.

Authors:  Cédric Barrey; Sabina Champain; Sophie Campana; Aymen Ramadan; Gilles Perrin; Wafa Skalli
Journal:  Eur Spine J       Date:  2012-02-14       Impact factor: 3.134

5.  Clinical and radiological outcome 1-year after cervical total disc replacement using the Signus ROTAIO - Prosthesis.

Authors:  Anna Lang; Sara Lener; Lukas Grassner; Anto Abramovic; Claudius Thomé; Dennis Päsler; Jens Lehmberg; Ralph Schär; Sebastian Hartmann
Journal:  Eur Spine J       Date:  2022-10-11       Impact factor: 2.721

6.  In vivo analysis of cervical kinematics after implantation of a minimally constrained cervical artificial disc replacement.

Authors:  Heiko Koller; Oliver Meier; Juliane Zenner; Michael Mayer; Wolfgang Hitzl
Journal:  Eur Spine J       Date:  2012-11-24       Impact factor: 3.134

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.