Measurements of bone mineral density in the lumbar spine and proximal femur using lunar prodigy and the new pencil-beam dual-energy X-ray absorptiometry.

Dongil Choi1, Deog-Yoon Kim, Chung Soo Han, Seonwoo Kim, Hae Sook Bok, Wooseong Huh, Jae-Wook Ko, Sung Hwa Hong.   

Abstract

OBJECTIVE: We evaluated the correlation of the absolute bone mineral density (BMD) values of the lumbar spine and standard sites of the proximal femur obtained from a Lunar Prodigy and the newly developed pencil-beam dual-energy X-ray absorptiometry (Dexxum).
MATERIALS AND METHODS: Between June 2008 and December 2008, 79 Korean volunteers were enrolled. Measurements were obtained on the same day using both densitometers. The absolute BMD values (g/cm(2)) from the two densitometers were evaluated using Pearson's correlation analysis with Bonferroni's correction for the three clinically important sites. In order to evaluate precision, we performed duplicate Dexxum measurements, and calculated the within-subject coefficient of variation (WSCV).
RESULTS: The Pearson's correlation coefficient (r) of BMD values for the total proximal femur, femoral neck, and lumbar spine by the two densitometers were 0.926, 0.948, and 0.955 respectively, and the null hypotheses of r = 0.8 were all rejected (p < 0.001 by one-sided Z-test with Fisher's z-transformation for each site). The T-scores (r ≧ 0.842) and Z-scores (r ≧ 0.709) also showed strong positive correlations. The duplicate BMD values of Dexxum showed a high level of precision (WSCV ≦ 4.27%).
CONCLUSION: Dexxum measurements of BMD, T-scores, and Z-scores showed a strong linear correlation with those measured on Lunar Prodigy.

Mesh:

Year:  2009        PMID: 19924413     DOI: 10.1007/s00256-009-0828-1

Source DB:  PubMed          Journal:  Skeletal Radiol        ISSN: 0364-2348            Impact factor:   2.199


  22 in total

1.  Assessment of spinal and femoral bone density by dual X-ray absorptiometry: comparison of lunar and hologic instruments.

Authors:  N A Pocock; P N Sambrook; T Nguyen; P Kelly; J Freund; J A Eisman
Journal:  J Bone Miner Res       Date:  1992-09       Impact factor: 6.741

2.  Comparative assessment of dual-photon absorptiometry and dual-energy radiography.

Authors:  C C Glüer; P Steiger; R Selvidge; K Elliesen-Kliefoth; C Hayashi; H K Genant
Journal:  Radiology       Date:  1990-01       Impact factor: 11.105

3.  A comparison of longitudinal measurements in the spine and proximal femur using lunar and hologic instruments.

Authors:  N A Pocock; K A Noakes; M Griffiths; N Bhalerao; P N Sambrook; J A Eisman; J Freund
Journal:  J Bone Miner Res       Date:  1997-12       Impact factor: 6.741

4.  Standardization of proximal femur BMD measurements. International Committee for Standards in Bone Measurement.

Authors:  J Hanson
Journal:  Osteoporos Int       Date:  1997       Impact factor: 4.507

5.  Universal standardization of bone density measurements: a method with optimal properties for calibration among several instruments.

Authors:  S L Hui; S Gao; X H Zhou; C C Johnston; Y Lu; C C Glüer; S Grampp; H Genant
Journal:  J Bone Miner Res       Date:  1997-09       Impact factor: 6.741

6.  The diagnosis of osteoporosis.

Authors:  J A Kanis; L J Melton; C Christiansen; C C Johnston; N Khaltaev
Journal:  J Bone Miner Res       Date:  1994-08       Impact factor: 6.741

Review 7.  Consensus of an international panel on the clinical utility of bone mass measurements in the detection of low bone mass in the adult population.

Authors:  P D Miller; S L Bonnick; C J Rosen
Journal:  Calcif Tissue Int       Date:  1996-04       Impact factor: 4.333

8.  Discrepancies in normative data between Lunar and Hologic DXA systems.

Authors:  K G Faulkner; L A Roberts; M R McClung
Journal:  Osteoporos Int       Date:  1996       Impact factor: 4.507

9.  Standardization of bone mineral density at femoral neck, trochanter and Ward's triangle.

Authors:  Y Lu; T Fuerst; S Hui; H K Genant
Journal:  Osteoporos Int       Date:  2001       Impact factor: 4.507

10.  Universal standardization for dual x-ray absorptiometry: patient and phantom cross-calibration results.

Authors:  H K Genant; S Grampp; C C Glüer; K G Faulkner; M Jergas; K Engelke; S Hagiwara; C Van Kuijk
Journal:  J Bone Miner Res       Date:  1994-10       Impact factor: 6.741

View more
  1 in total

1.  Natural changes of traumatic vertebral compression fractures during the first 6 months in patients visiting for disability certificates: A retrospective observational study.

Authors:  Jin Seok Bae; InHyuk Suh; Jong Keun Kim; Yong Sung Jeong; Jong Youb Lim
Journal:  Medicine (Baltimore)       Date:  2021-01-08       Impact factor: 1.817

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.