| Literature DB >> 19893700 |
Abstract
In this study, absorbed doses were measured and compared for high-energy electrons (6, 9, 12, 16, and 20 MeV) using International Atomic Energy Agency (IAEA), Technical Reports Series No. 277 (TRS), TRS 381, and TRS 398 dosimetry protocols. Absolute dose measurements were carried out using FC65-G Farmer chamber and Nordic Association of Clinical Physicists (NACP) parallel plate chamber with DOSE1 electrometer in WP1-D water phantom for reference field size of 15 x 15 cm(2) at 100 cm source-to-surface distance. The results show that the difference between TRS 398 and TRS 381 was about 0.24% to 1.3% depending upon the energy, and the maximum difference between TRS 398 and TRS 277 was 1.5%. The use of cylindrical chamber in electron beam gives the maximum dose difference between the TRS 398 and TRS 277 in the order of 1.4% for energies above 10 MeV (R(50) > 4 g/cm(2)). It was observed that the accuracy of dose estimation was better with the protocols based on the water calibration procedures, as no conversion quantities are involved for conversion of dose from air to water. The cross-calibration procedure of parallel plate chamber with high-energy electron beams is recommended as it avoids p(wall) correction factor entering into the determination of k(Q,Qo).Entities:
Keywords: Absorbed dose; chamber; cross-calibration; protocol
Year: 2008 PMID: 19893700 PMCID: PMC2772041 DOI: 10.4103/0971-6203.42754
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Figure 1Experimental setup
Measurement depth zref used in various protocols
| 6 MeV | 1.33 | 1.33 | 1.33 |
| 9 MeV | 2.11 | 2.11 | 2.05 |
| 12 MeV | 2.85 | 2.85 | 2.89 |
| 16 MeV | 3.23 | 3.23 | 3.85 |
| 20 MeV | 2.31 | 2.31 | 4.88 |
Calibration factors for the chambers used in this study
| NACP Plane parallel chamber | 1.4422 × 108 Gy / C | 1.634 × 108 Gy / C | 1.4591 × 108 Gy / C |
| FC65-G Farmer chamber | 4.2409 × 107 Gy / C | 4.805 × 107 Gy / C | —— |
Various correction factors as a function of energy and chamber
| Energy | NACP Plane parallel chamber | FC65-G Farmer chamber | ||||||
|---|---|---|---|---|---|---|---|---|
| kQ,Q0 | Ion recombination (ks) | Polarity effect (kpol) | Stopping power ratio (sw,air) | kQ,Q0 | Ion recombination (ks) | Polarity effect (kpol) | Stopping power ratio (sw,air) | |
| 6 MeV | 0.926 | 1.008 | 0.998 | 1.079 | — | — | — | — |
| 9 MeV | 0.911 | 1.009 | 0.999 | 1.050 | — | — | — | – |
| 12 MeV | 0.899 | 1.009 | 1.000 | 1.037 | 0.916 | 1.013 | 1.001 | 1.037 |
| 16 MeV | 0.888 | 1.001 | 1.003 | 1.010 | 0.911 | 1.012 | 1.000 | 1.010 |
| 20 MeV | 0.879 | 1.010 | 1.000 | 0.977 | 0.906 | 1.013 | 1.000 | 0.977 |
Calibration of chambers for dosimetry with high-energy electron beams according to the IAEA technical reports
| Report | Calibration | Phantom material | Beam | Remarks |
|---|---|---|---|---|
| TRS 277 (Update 1997) | Absorbed dose to air | Air | Co-60 | Calibration laboratory |
| TRS 277 (Update 1997) | Absorbed dose to water | Water | Co-60 | Calibration laboratory |
| TRS 381 (1997) | Absorbed dose to air | Air | Co-60 | Only for plane-parallel chambers, Calibration laboratory |
| TRS 381 (1997) | Absorbed dose to air | Plastic | Co-60 | Only for plane-parallel chambers, Calibration laboratory or User |
| TRS 381 (1997) | Absorbed dose to air | Water | Electron beam | Only for plane-parallel chambers, Calibration laboratory or User |
| TRS 381 (1997) | Absorbed dose to water | Water | Co-60 | Only for plane-parallel chambers, Calibration laboratory or User |
| TRS 398 (2000) | Absorbed dose to water | Water | Co-60 | Calibration laboratory |
| TRS 398 (2000) | Absorbed dose to water | Water | Electron beam | Only for plane-parallel chambers, “cross-calibration” user |
Figure 2Experimental comparison of dose ratios of IAEA TRS protocols in electron beams, at the depth of maximum dose for plane-parallel ion chamber having ND,w calibrations in 60Co
Figure 3Experimental comparison of dose ratios TRS 381 / TRS 398 in electron beams, at the depth of maximum dose for plane-parallel ion chamber cross calibrated in high energy electron against Farmer type ion chamber
Figure 4Experimental comparison of dose ratios TRS 277 / TRS 398 in electron beams, at the depth of maximum dose for cylindrical ion chamber having ND,w calibrations in 60Co