| Literature DB >> 35054424 |
Fawzia E M Elbashir1, Wassim Ksouri2, Mohamed Hassan Eisa3, Sitah Alanazi3, Farouk Habbani4, Abdelmoneim Sulieman5, David A Bradley6,7, Ibrahim I Suliman3,8.
Abstract
This paper presents guidelines for the calibration of radiation beams that were issued by the International Atomic Energy Agency (IAEA TRS 398), the American Association of Physicists in Medicine (AAPM TG 51) and the German task group (DIN 6800-2). These protocols are based on the use of an ionization chamber calibrated in terms of absorbed dose to water in a standard laboratory's reference quality beam, where the previous protocols were based on air kerma standards. This study aims to determine uncertainties in dosimetry for electron beam radiotherapy using internationally established high-energy radiotherapy beam calibration standards.Entities:
Keywords: absorbed dose standards; ionisation chamber; medical LINAC; radiation dosimetry; radiotherapy
Year: 2021 PMID: 35054424 PMCID: PMC8781094 DOI: 10.3390/life12010031
Source DB: PubMed Journal: Life (Basel) ISSN: 2075-1729
Summary of different annotations used in the three dosimetry protocols.
| Factor | Annotations | ||
|---|---|---|---|
| IAEA TRS 398 | AAPM TG 51 | DIN 6800-2 | |
| Reference radiation beam quality (cobalt-60) |
| 60 | - |
| Hospital radiation beam quality |
|
| - |
| absorbed dose to water |
|
|
|
| absorbed dose to water calibration factor |
|
|
|
| beam quality correction factor |
|
| K |
| the corrected ion chamber reading |
|
|
|
Physical characteristics of plane-parallel and cylindrical ionization chambers.
| Chamber | Type | Entrance Window | Cavity Wall | Cavity | Waterproof (Y/N) | Preference | ||
|---|---|---|---|---|---|---|---|---|
| Material | Material | Thickness (mm) | ||||||
| NACP-02 (A) | parallel plate | 0.17 mm | 7 | NA | NA | NA | no | Absolute dosimetry |
| NACP-02 (B) | parallel plate | 0.17 mm | 7 | NA | NA | NA | no | Absolute dosimetry |
| IC15 | cylindrical | NA | NA | C552 | 0.4 | 0.13 | yes | Relative dosimetry |
| CC13 | cylindrical | NA | NA | C552 | 0.4 | 0.13 | yes | Relative dosimetry |
| FC65-G | cylindrical | NA | NA | Graphite | 0.4 | 0.65 | yes | Absolute dosimetry |
Characteristics of the electron beams used.
| Beam Energy (MeV) | ||||
|---|---|---|---|---|
| TG 51, DIN 6800-2 | TRS 398 | |||
| 6 | 2.3 | 1.3 | 1.3 | 1.4 |
| 12 | 4.9 | 2.9 | 2.8 | 3.0 |
| 18 | 7.6 | 2.5 | 4.5 | 4.6 |
Beam quality correction factors used to calculate absorbed dose to water using the TRS 398, TG 51 and DIN 6800-2 absorbed dose-based protocols.
| Chamber | Beam Energy |
|
| ||
|---|---|---|---|---|---|
| TRS 398 | TG -51 | DIN 6800-2 | |||
| NACP-02 (13505) | 6 | 1.748 × 108 | 0.9268 | 0.9302 | 0.9293 |
| 12 | 0.8990 | 0.9059 | 0.9023 | ||
| 18 | 0.8824 | 0.8881 | 0.8845 | ||
| NACP-02 (13703) | 6 | 1.58 × 108 | 0.9268 | 0.9302 | 0.9293 |
| 12 | 0.8990 | 0.9059 | 0.9023 | ||
| 18 | 0.8824 | 0.8881 | 0.8845 | ||
| IC15 | 6 | 1.3633 × 108 | 0.9350 | 0.9203 | 0.9395 |
| 12 | 0.9172 | 0.9093 | 0.9138 | ||
| 18 | 0.9082 | 0.9005 | 0.9060 | ||
| FC65-G | 6 | 4.738 × 107 | 0.9350 | 0.9218 | 0.9376 |
| 12 | 0.9164 | 0.9092 | 0.9128 | ||
| 18 | 0.9082 | 0.8984 | 0.9051 | ||
Polarity effect (k), ion recombination (k) and temperature and pressure correction factors (k).
| Correction Factor | 6 MeV | 12 MeV | 18 MeV | ||||||
|---|---|---|---|---|---|---|---|---|---|
| TRS | TG 51 | DIN 6800-2 | TRS | TG 51 | 6800-2 | TRS | TG 51 | 6800-2 | |
| NACP | |||||||||
|
| 0.9994 | 0.9994 | 0.9996 | 0.9993 | 0.9993 | 0.9990 | 0.9981 | 0.9981 | 0.9989 |
|
| 1.0096 | 1.0099 | 1.0089 | 1.0107 | 1.0110 | 1.0028 | 1.0094 | 1.0096 | 1.0113 |
|
| 1.0169 | 1.0100 | 1.0169 | 1.0169 | 1.0100 | 1.0169 | 1.0169 | 1.0100 | 1.0169 |
| CC13 | |||||||||
|
| 1.0025 | 0.9922 | 1.0027 | 1.0009 | 1.0008 | 1.0010 | 1.0010 | 1.0005 | 1.0010 |
|
| 1.0244 | 1.0365 | 1.0252 | 1.0246 | 1.0247 | 1.0255 | 1.0256 | 1.0252 | 1.0259 |
|
| 1.0168 | 1.0099 | 1.0168 | 1.0168 | 1.0099 | 1.0168 | 1.0169 | 1.0099 | 1.0168 |
| FC65-G | |||||||||
|
| 1.0014 | 1.0013 | 1.0011 | 1.0001 | 1.0001 | 0.9999 | 1.0001 | 0.9988 | 0.9992 |
|
| 1.0175 | 1.0181 | 1.0182 | 1.0177 | 1.0181 | 1.0180 | 1.0184 | 1.0188 | 1.0197 |
|
| 1.0291 | 1.0221 | 1.0291 | 1.0291 | 1.0221 | 1.0291 | 1.0291 | 1.0221 | 1.0291 |
Conversion factors for absorbed dose to water between the TRS 398, TG 51 and DIN 6800-2 dosimetry protocols using NACP-02, CC13, IC15 and FC65-G ionisation chambers.
| Ionisation Chamber | Beam Energy | IAEA TRS 398/ | IAEA TRS 398/ | AAPM TG 51/ |
|---|---|---|---|---|
| NACP-02 | 6 | 1.0021 | 1.0029 | 0.9992 |
| 12 | 1.0037 | 1.0138 | 0.9901 | |
| 18 | 1.0162 | 0.9895 | 1.0270 | |
| CC13 | 6 | 0.9866 | 1.0040 | 0.9826 |
| 12 | 0.9885 | 0.9952 | 0.9933 | |
| 18 | 0.9886 | 0.9726 | 1.0164 | |
| IC15 | 6 | 0.9917 | 1.0126 | 0.9793 |
| 12 | 0.9934 | 0.9987 | 0.9948 | |
| 18 | 1.0164 | 0.9963 | 1.0202 | |
| FC65-G | 6 | 1.0065 | 1.0020 | 1.0045 |
| 12 | 0.9917 | 0.9967 | 0.9950 | |
| 18 | 1.0127 | 0.9980 | 1.0148 |
Figure 1Ratio of the absorbed dose to water values obtained using five ionization chambers in three electron beams: (a) IAEA TRS 398/AAPM TG 51; (b) IAEA TRS 398/DIN 6800-2; (c) AAPM TG 51/DIN 6800-2.
Relative uncertainty (%) associated with measurements of the absorbed dose to water, Dw, in energy electron beam radiotherapy using the IAEA TRS 398 and DIN 6800-2 dosimetry protocols.
| Influence Quantities | Evaluation | Cylindrical Chamber | Plane-Parallel Chamber | |||
|---|---|---|---|---|---|---|
| TRS 398 | DIN 6800-2 | TRS 398 | DIN 6800-2 | |||
| ND, W | Chamber certificate | B | 0.55 | 0.55 | 0.55 | 0.55 |
| Measurement depth | Calculated | B | 0.46 | 0.46 | 0.46 | 0.46 |
|
| Calculated | A/B | 0.04 | 0.04 | 0.12 | 0.12 |
|
| Calculated | A/B | 0.01 | 0.01 | 0.01 | 0.01 |
|
| Calculated | B | 0.04 | 0.04 | 0.12 | 0.12 |
|
| IAEA | B | 1.2 | - | 1.7 | - |
|
| DIN 6800-2 | B | - | 1.2 | - | 1.3 |
| Meter stability | Dosimeter | B | 0.28 | 0.28 | 0.28 | 0.28 |
| Meter reading | Calculated | A | 0.2 | 0.2 | 0.2 | 0.2 |
| Combined standard uncertainty ( | 1.44 | 1.44 | 1.88 | 1.53 | ||