An Image-Guided Radiotherapy-capable linear accelerator (Elekta Synergy) was installed at our hospital, which is equipped with a kV x-ray volume imaging (XVI) system and electronic portal imaging device (iViewGT). The objective of this presentation is to describe the results of commissioning measurements carried out on the XVI facility to verify the manufacturer's specifications and also to evolve a QA schedule which can be used to test its performance routinely.The QA program consists of a series of tests (safety features, geometric accuracy, and image quality). These tests were found to be useful to assess the performance of the XVI system and also proved that XVI system is very suitable for image-guided high-precision radiation therapy.Entities:
Keywords: Contrast visibility; X-Ray Volume Imaging system; image-guided radiation therapy; spatial resolution
Year: 2008 PMID: 19893694 PMCID: PMC2772033 DOI: 10.4103/0971-6203.41276
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Figure 1Catphan phantom 503-setup in Linear accelerator
Figure 2Measuring the distance between the two inserts
Figure 3Ball-bearing phantom
Figure 4Registraton on transverse, coronal and saggital views
Results of volume view registration accuracy tests
| Lateral (Y) | −0.9 |
| Longitudinal (X) | −1.9 |
| Height (Z) | −0.4 |
Results of registration accuracy tests
| Lateral (Y) | 0.2 | 0.36 | 0.2 |
| Longitudinal (X) | −0.2 | −0.73 | 0.5 |
| Height (Z) | 0.2 | −0.35 | 0.5 |
Results of 2-D geometric accuracy test
| X | Y | X | Y | X | Y | ||
|---|---|---|---|---|---|---|---|
| 1 | 0 | 505 | 521 | 504 | 518 | +1 | +3 |
| 2 | 90 | 507 | 524 | 509 | 520 | −2 | +4 |
| 3 | 180 | 501 | 519 | 503 | 515 | −2 | +4 |
| 4 | 270 | 509 | 520 | 509 | 516 | 0 | +4 |
Figure 53-D low contrast visibility from Catphan phantom CTP-503
Results of 3-D low-contrast visibility test
| Polystyrene | 1085.57 | 17.8 |
| LDPE | 827.27 | 18.65 |
Figure 6Spatial resolution-transverse
Figure 73-D uniformity
Mean pixel values in different locations
| Mean pixel value (center) | 646.34 |
| Location 1 | 646.21 |
| Location 2 | 645.90 |
| Location 3 | 646.44 |
Figure 8Resolution grid and number of spatial frequency groups