Literature DB >> 19891717

Influence of the shape of artificial canals on the fatigue resistance of NiTi rotary instruments.

G Plotino1, N M Grande, M Cordaro, L Testarelli, G Gambarini.   

Abstract

AIM: To investigate the influence of the trajectory of NiTi rotary instruments on the outcome of cyclic fatigue tests.
METHODOLOGY: Ten ProFile and Mtwo instruments tip size 20, taper 0.06 and tip size 25, taper 0.06 were tested in two simulated root canals with an angle of curvature of 60 degrees and radius of curvature of 5 mm but with different shape. Geometrical analysis of the angle and radius of the curvature that each instrument followed inside the two different artificial canals was performed on digital images. The instruments were then rotated until fracture at a constant speed of 300 rpm to calculate the number of cycles to failure (NCF) and the length of the fractured fragment. Mean values were calculated and analysed using two different multivariate linear regression models and an independent sample t-test.
RESULTS: The shape of the artificial root canal used in cyclic fatigue studies influenced the trajectory of the instrument. This difference is reflected by the NCF measured for the same instrument in the different artificial root canals and by the impact of the type of canal on both the NCF (St.beta = 0.514) and fragment length (St.beta = -0.920).
CONCLUSIONS: Small variations in the geometrical parameters of the curvature of an instrument subjected to flexural fatigue could have a significant influence on the results of fatigue tests.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 19891717     DOI: 10.1111/j.1365-2591.2009.01641.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  5 in total

1.  Deformations and cyclic fatigue resistance of nickel-titanium instruments inside a sequence.

Authors:  Gianluca Gambarini; Gianluca Plotino; Lucila Piasecki; Dina Al-Sudani; Luca Testarelli; Gianpaolo Sannino
Journal:  Ann Stomatol (Roma)       Date:  2015-05-18

2.  Fatigue resistance of rotary instruments manufactured using different nickel-titanium alloys: a comparative study.

Authors:  Gianluca Plotino; Luca Testarelli; Dina Al-Sudani; Gianlcarlo Pongione; Nicola M Grande; Gianluca Gambarini
Journal:  Odontology       Date:  2012-10-18       Impact factor: 2.634

3.  Flexibility and resistance to cyclic fatigue of endodontic instruments made with different nickel-titanium alloys: a comparative test.

Authors:  Giancarlo Pongione; Giorgio Pompa; Valerio Milana; Stefano Di Carlo; Alessio Giansiracusa; Emanuele Nicolini; Francesca De Angelis
Journal:  Ann Stomatol (Roma)       Date:  2012-01-14

4.  Cyclic fatigue resistance of Mtwo NiTi rotary instruments used by experienced and novice operators--an in vivo and in vitro study.

Authors:  Gianluca Plotino; Dina Al-Sudani; Stefano Pulino; Nicola M Grande; Piero Alessandro Marcoli; Silvia Pizzi; Luca Testarelli; Gianluca Gambarini
Journal:  Med Sci Monit       Date:  2012-06

5.  Comparing the Fracture Rate of Hero 642, FlexMaster and Mtwo in the Simulated Canals.

Authors:  Hosein Labaf; Roza Haghgoo; Kiumars Nazarimoghadam; Mahshid Mohamadibasir
Journal:  Iran Endod J       Date:  2014-03-08
  5 in total