Literature DB >> 19885091

Electrochemical sensor array for glucose monitoring fabricated by rapid immobilization of active glucose oxidase within photochemically polymerized hydrogels.

Amos Mugweru1, Becky L Clark, Michael V Pishko.   

Abstract

BACKGROUND: Currently, monitoring blood glucose levels for diabetic patients is invasive and painful, involving pricking the finger to obtain a blood sample three to four times daily. The need for frequent tests and pain involved with testing leads to poor compliance. In order to raise compliance, we propose to create an implantable electrochemical sensor array that would monitor glucose levels continuously.
METHODS: Glucose sensor arrays were fabricated on gold electrodes on flexible polyimide sheets by photopolymerization of the biocompatible polymer poly(ethylene glycol) diacrylate (PEG-DA) to develop hydrogels and encapsulate the sensing elements. Using conventional silicon fabrication methods, arrays of five gold microdisk electrodes were fabricated using lift-off photolithography and sputtering techniques. A redox polymer was then attached electrostatically to the electrode, and glucose oxidase was entrapped inside the hydrogel on the array of electrodes by ultraviolet-initiated photopolymerization of PEG-DA.
RESULTS: When the array of fabricated sensors was sampled together the elements behaved like one large electrode with peak current equivalent to the sum of individual array elements. The enzyme, glucose oxidase, catalyzed the oxidation of glucose and then exchanged electrons with the redox polymer in the hydrogel. The entrapped glucose oxidase was found to respond linearly to increasing glucose concentrations (0-360 mg/dl), as determined using cyclic voltammetry.
CONCLUSION: The fabricated microarray sensors were individually addressable and showed no cross talk between adjacent array elements as assessed using cyclic voltammetry. We have fabricated an array of glucose sensors on flexible polyimide sheets that exhibits the desired linear response in the biological range.

Entities:  

Keywords:  biocompatible; cyclic voltammetry; glucose sensor; microarray

Year:  2007        PMID: 19885091      PMCID: PMC2769590          DOI: 10.1177/193229680700100308

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  26 in total

Review 1.  Biomaterial-microvasculature interactions.

Authors:  A L Sieminski; K J Gooch
Journal:  Biomaterials       Date:  2000-11       Impact factor: 12.479

2.  A microfluidic bioreactor based on hydrogel-entrapped E. coli: cell viability, lysis, and intracellular enzyme reactions.

Authors:  Jinseok Heo; K Joseph Thomas; Gi Hun Seong; Richard M Crooks
Journal:  Anal Chem       Date:  2003-01-01       Impact factor: 6.986

3.  Noninvasive glucose monitoring in diabetic patients: a preliminary evaluation.

Authors:  M R Robinson; R P Eaton; D M Haaland; G W Koepp; E V Thomas; B R Stallard; P L Robinson
Journal:  Clin Chem       Date:  1992-09       Impact factor: 8.327

4.  Design and in vitro studies of a needle-type glucose sensor for subcutaneous monitoring.

Authors:  D S Bindra; Y Zhang; G S Wilson; R Sternberg; D R Thévenot; D Moatti; G Reach
Journal:  Anal Chem       Date:  1991-09-01       Impact factor: 6.986

5.  Statistics for critical clinical decision making based on readings of pairs of implanted sensors.

Authors:  D W Schmidtke; M V Pishko; C P Quinn; A Heller
Journal:  Anal Chem       Date:  1996-09-01       Impact factor: 6.986

6.  Kinetics of glucose delivery to subcutaneous tissue in rats measured with 0.3-mm amperometric microsensors.

Authors:  C P Quinn; M V Pishko; D W Schmidtke; M Ishikawa; J G Wagner; P Raskin; J A Hubbell; A Heller
Journal:  Am J Physiol       Date:  1995-07

7.  Noninvasive blood glucose sensors based on near-infrared spectroscopy.

Authors:  H M Heise; R Marbach; T Koschinsky; F A Gries
Journal:  Artif Organs       Date:  1994-06       Impact factor: 3.094

Review 8.  Progress toward the development of an implantable sensor for glucose.

Authors:  G S Wilson; Y Zhang; G Reach; D Moatti-Sirat; V Poitout; D R Thévenot; F Lemonnier; J C Klein
Journal:  Clin Chem       Date:  1992-09       Impact factor: 8.327

9.  Fabrication of poly(ethylene glycol) hydrogel microstructures using photolithography.

Authors:  A Revzin; R J Russell; V K Yadavalli; W G Koh; C Deister; D D Hile; M B Mellott; M V Pishko
Journal:  Langmuir       Date:  2001-09-04       Impact factor: 3.882

10.  Photo-crosslinked copolymers of 2-hydroxyethyl methacrylate, poly(ethylene glycol) tetra-acrylate and ethylene dimethacrylate for improving biocompatibility of biosensors.

Authors:  C P Quinn; C P Pathak; A Heller; J A Hubbell
Journal:  Biomaterials       Date:  1995-03       Impact factor: 12.479

View more
  6 in total

1.  Lab-on-a-chip technology for continuous glucose monitoring.

Authors:  Peter Gravesen; Kristian Raaby Poulsen; Holger Dirac
Journal:  J Diabetes Sci Technol       Date:  2007-05

Review 2.  Electrochemical glucose sensors--developments using electrostatic assembly and carbon nanotubes for biosensor construction.

Authors:  Alice Harper; Mark R Anderson
Journal:  Sensors (Basel)       Date:  2010-09-02       Impact factor: 3.576

Review 3.  Modulating the foreign body response of implants for diabetes treatment.

Authors:  Bhushan N Kharbikar; Gauree S Chendke; Tejal A Desai
Journal:  Adv Drug Deliv Rev       Date:  2021-01-21       Impact factor: 17.873

Review 4.  Wearable Biosensors for Non-Invasive Sweat Diagnostics.

Authors:  Jing Xu; Yunsheng Fang; Jun Chen
Journal:  Biosensors (Basel)       Date:  2021-07-23

5.  Development of an Amperometric Glucose Biosensor Based on the Immobilization of Glucose Oxidase on the Se-MCM-41 Mesoporous Composite.

Authors:  Sabriye Yusan; Mokhlesur M Rahman; Nasir Mohamad; Tengku M Arrif; Ahmad Zubaidi A Latif; Mohd Aznan M A; Wan Sani B Wan Nik
Journal:  J Anal Methods Chem       Date:  2018-05-10       Impact factor: 2.193

Review 6.  Comprehensive Review on Wearable Sweat-Glucose Sensors for Continuous Glucose Monitoring.

Authors:  Hima Zafar; Asma Channa; Varun Jeoti; Goran M Stojanović
Journal:  Sensors (Basel)       Date:  2022-01-14       Impact factor: 3.576