Dosimetric characterization of round HDR 192Ir accuboost applicators for breast brachytherapy.

Mark J Rivard1, Christopher S Melhus, David E Wazer, Raymond J Bricault.   

Abstract

PURPOSE: The AccuBoost brachytherapy system applies HDR 192Ir beams peripherally to the breast using collimating applicators. The purpose of this study was to benchmark Monte Carlo simulations of the HDR 192Ir source, to dosimetrically characterize the round applicators using established Monte Carlo simulation and radiation measurement techniques and to gather data for clinical use.
METHODS: Dosimetric measurements were performed in a polystyrene phantom, while simulations estimated dose in air, liquid water, polystyrene and ICRU 44 breast tissue. Dose distribution characterization of the 4-8 cm diameter collimators was performed using radiochromic EBT film and air ionization chambers.
RESULTS: The central axis dose falloff was steeper for the 4 cm diameter applicator in comparison to the 8 cm diameter applicator, with surface to 3 cm depth-dose ratios of 3.65 and 2.44, respectively. These ratios did not considerably change when varying the phantom composition from breast tissue to polystyrene, phantom thickness from 4 to 8 cm, or phantom radius from 8 to 15 cm. Dose distributions on the central axis were fitted to sixth-order polynomials for clinical use in a hand calculation spreadsheet (i.e., nomogram). Dose uniformity within the useful applicator apertures decreased as depth-dose increased.
CONCLUSIONS: Monte Carlo benchmarking simulations of the HDR 192Ir source using the MCNP5 radiation transport code indicated agreement within 1% of the published results over the radial/angular region of interest. Changes in phantom size and radius did not cause noteworthy changes in the central axis depth-dose. Polynomial fit depth-dose curves provide a simple and accurate basis for a nomogram.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19994513      PMCID: PMC2773456          DOI: 10.1118/1.3232001

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations.

Authors:  Mark J Rivard; Bert M Coursey; Larry A DeWerd; William F Hanson; M Saiful Huq; Geoffrey S Ibbott; Michael G Mitch; Ravinder Nath; Jeffrey F Williamson
Journal:  Med Phys       Date:  2004-03       Impact factor: 4.071

2.  Equivalent phantom sizes and shapes for brachytherapy dosimetric studies of 192Ir and 137Cs.

Authors:  Domingo Granero; Jose Perez-Calatayud; M Carmen Pujades-Claumarchirant; Facundo Ballester; Christopher S Melhus; Mark J Rivard
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

3.  Evaluation of high-energy brachytherapy source electronic disequilibrium and dose from emitted electrons.

Authors:  Facundo Ballester; Domingo Granero; José Pérez-Calatayud; Christopher S Melhus; Mark J Rivard
Journal:  Med Phys       Date:  2009-09       Impact factor: 4.071

4.  Monte Carlo-aided dosimetry of a new high dose-rate brachytherapy source.

Authors:  G M Daskalov; E Löffler; J F Williamson
Journal:  Med Phys       Date:  1998-11       Impact factor: 4.071

5.  Dosimetry of interstitial brachytherapy sources: recommendations of the AAPM Radiation Therapy Committee Task Group No. 43. American Association of Physicists in Medicine.

Authors:  R Nath; L L Anderson; G Luxton; K A Weaver; J F Williamson; A S Meigooni
Journal:  Med Phys       Date:  1995-02       Impact factor: 4.071

6.  Brachytherapy dosimetry parameters calculated for a 131Cs source.

Authors:  Mark J Rivard
Journal:  Med Phys       Date:  2007-02       Impact factor: 4.071

7.  Monte Carlo simulations and radiation dosimetry measurements of peripherally applied HDR 192Ir breast brachytherapy D-shaped applicators.

Authors:  Yun Yang; Mark J Rivard
Journal:  Med Phys       Date:  2009-03       Impact factor: 4.071

  7 in total
  8 in total

1.  Treatment planning of a skin-sparing conical breast brachytherapy applicator using conventional brachytherapy software.

Authors:  Yun Yang; Christopher S Melhus; Shirin Sioshansi; Mark J Rivard
Journal:  Med Phys       Date:  2011-03       Impact factor: 4.071

2.  Evaluation of (101)Rh as a brachytherapy source.

Authors:  Delaram Pakravan; Mahdi Ghorbani; Ali Soleimani Meigooni
Journal:  J Contemp Brachytherapy       Date:  2015-04-01

3.  Multi-axis dose accumulation of noninvasive image-guided breast brachytherapy through biomechanical modeling of tissue deformation using the finite element method.

Authors:  Mark J Rivard; Hamid R Ghadyani; Adam D Bastien; Nicholas N Lutz; Jaroslaw T Hepel
Journal:  J Contemp Brachytherapy       Date:  2015-02-17

4.  Commissioning and quality assurance for the treatment delivery components of the AccuBoost system.

Authors:  Ileana Iftimia; Mike Talmadge; Ron Ladd; Per Halvorsen
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

5.  Automated position and size selection of round applicators for AccuBoost breast brachytherapy.

Authors:  Foster L West; Reshma Munbodh; John C Patrick; Mark J Rivard; Sean A Roles; Ziad H Saleh
Journal:  J Contemp Brachytherapy       Date:  2020-12-16

6.  Optically stimulated luminescent dosimetry for high dose rate brachytherapy.

Authors:  Christopher Jason Tien; Robert Ebeling; Jessica R Hiatt; Bruce Curran; Edward Sternick
Journal:  Front Oncol       Date:  2012-08-06       Impact factor: 6.244

Review 7.  Accelerated partial breast irradiation with brachytherapy: patient selection and technique considerations.

Authors:  Daniel M Trifiletti; Kara D Romano; Shayna L Showalter; Kelli A Reardon; Bruce Libby; Timothy N Showalter
Journal:  Breast Cancer (Dove Med Press)       Date:  2015-07-29

8.  Evaluation of BEBIG HDR (60)Co system for non-invasive image-guided breast brachytherapy.

Authors:  Mehdi Zehtabian; Sedigheh Sina; Mark J Rivard; Ali S Meigooni
Journal:  J Contemp Brachytherapy       Date:  2015-12-30
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.