Blood metal ion concentrations after hip resurfacing arthroplasty: a comparative study of articular surface replacement and Birmingham Hip Resurfacing arthroplasties.

D J Langton1, A P Sprowson, T J Joyce, M Reed, I Carluke, P Partington, A V F Nargol.   

Abstract

There have been no large comparative studies of the blood levels of metal ions after implantation of commercially available hip resurfacing devices which have taken into account the effects of femoral size and inclination and anteversion of the acetabular component. We present the results in 90 patients with unilateral articular surface replacement (ASR) hip resurfacings (mean time to blood sampling 26 months) and 70 patients with unilateral Birmingham Hip Resurfacing (BHR) implants (mean time 47 months). The whole blood and serum chromium (Cr) and cobalt (Co) concentrations were inversely related to the size of the femoral component in both groups (p < 0.05). Cr and Co were more strongly influenced by the position of the acetabular component in the case of the ASR, with an increase in metal ions observed at inclinations > 45 degrees and anteversion angles of < 10 degrees and > 20 degrees. These levels were only increased in the BHR group when the acetabular component was implanted with an inclination > 55 degrees. A significant relationship was identified between the anteversion of the BHR acetabular component and the levels of Cr and Co (p < 0.05 for Co), with an increase observed at anteversion angles < 10 degrees and > 20 degrees. The median whole blood and serum Cr concentrations of the male ASR patients were significantly lower than those of the BHR men (p < 0.001). This indicates that reduced diametral clearance may equate to a reduction in metal ion concentrations in larger joints with satisfactory orientation of the acetabular component.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19794161     DOI: 10.1302/0301-620X.91B10.22308

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  60 in total

1.  [Durom™ hip resurfacing. Short- to midterm clinical and radiological outcome].

Authors:  J Goronzy; M Stiehler; S Kirschner; K-P Günther
Journal:  Orthopade       Date:  2010-09       Impact factor: 1.087

Review 2.  Survival of hard-on-hard bearings in total hip arthroplasty: a systematic review.

Authors:  Michael G Zywiel; Siraj A Sayeed; Aaron J Johnson; Thomas P Schmalzried; Michael A Mont
Journal:  Clin Orthop Relat Res       Date:  2011-06       Impact factor: 4.176

3.  Failure of a metal on metal hip prosthesis presenting as a destructive soft tissue mass due to ALVAL.

Authors:  Jonathan R B Hutt; Constant Busch; Rod A Hughes
Journal:  Rheumatol Int       Date:  2010-11-27       Impact factor: 2.631

Review 4.  Surgical management of osteoarthritis.

Authors:  Jeffrey N Katz; Brandon E Earp; Andreas H Gomoll
Journal:  Arthritis Care Res (Hoboken)       Date:  2010-09       Impact factor: 4.794

5.  Serum metal ion levels after rotating-hinge knee arthroplasty: comparison between a standard device and a megaprosthesis.

Authors:  Joerg Friesenbichler; Werner Maurer-Ertl; Patrick Sadoghi; Thomas Lovse; Reinhard Windhager; Andreas Leithner
Journal:  Int Orthop       Date:  2011-07-14       Impact factor: 3.075

6.  Surgical variables influence metal ion levels after hip resurfacing.

Authors:  Nicholas M Desy; Stephane G Bergeron; Alain Petit; Olga L Huk; John Antoniou
Journal:  Clin Orthop Relat Res       Date:  2011-06       Impact factor: 4.176

7.  The effects on bone cells of metal ions released from orthopaedic implants. A review.

Authors:  Valerio Sansone; Davide Pagani; Marco Melato
Journal:  Clin Cases Miner Bone Metab       Date:  2013-01

8.  Assessment of inter- and intra-observer reliability in the determination of radiographic version and inclination of the cup in metal-on-metal hip resurfacing.

Authors:  Aleksi Reito; Timo Puolakka; Antti Paakkala; Jorma Pajamäki
Journal:  Int Orthop       Date:  2011-08-12       Impact factor: 3.075

Review 9.  Management of metal-on-metal hip implant patients: Who, when and how to revise?

Authors:  Reshid Berber; John A Skinner; Alister J Hart
Journal:  World J Orthop       Date:  2016-05-18

10.  Histological features of pseudotumor-like tissues from metal-on-metal hips.

Authors:  Pat Campbell; Edward Ebramzadeh; Scott Nelson; Karren Takamura; Koen De Smet; Harlan C Amstutz
Journal:  Clin Orthop Relat Res       Date:  2010-09       Impact factor: 4.176

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.