| Literature DB >> 19774485 |
Wisnumurti Kristanto1, Peter M van Ooijen, Riksta Dikkers, Marcel J Greuter, Felix Zijlstra, Matthijs Oudkerk.
Abstract
Multi detector-row CT (MDCT), the current preferred method for coronary artery disease assessment, is still affected by motion artefacts. To rule out motion artefacts, qualitative image analysis is usually performed. Our study aimed to develop a quantitative image analysis for motion artefacts detection as an added value to the qualitative analysis. An anthropomorphic moving heart phantom with adjustable heart-rate was scanned on 64-MDCT and dual-source-CT. A new software technique was developed which detected motion artefacts in the coronaries and also in the myocardium, where motion artefacts are more apparent; with direct association to the qualitative analysis. The new quantitative analysis managed to detect motion artefacts in phantom scans and relate them to artefact-induced vessel stenoses. Quantifying these artefacts at corresponding locations in the myocardium, artefact-induced vessel stenosis findings could be avoided. In conclusion, the quantitative analysis together with the qualitative analysis rules out artefact-induced stenosis.Entities:
Mesh:
Year: 2009 PMID: 19774485 PMCID: PMC2795151 DOI: 10.1007/s10554-009-9502-x
Source DB: PubMed Journal: Int J Cardiovasc Imaging ISSN: 1569-5794 Impact factor: 2.357
Fig. 1The phantom used for experiment. Images without motion artefacts (left) and with motion artefacts (right) are shown both in a volume reconstruction (top) and a sagittal reconstruction (bottom). Arrows no. 1 and 2 denote start and end location of vessel analysis, respectively. Dotted white lines on the top row indicate the location of the sagittal slices
Fig. 2Diagram of the quantitative image analysis methods for the detection of motion artefacts in coronary artery computed tomography
Fig. 3Illustration of myocardium analysis. From the sagittal cross section image of the phantom (left), the inner-lining of the myocardium was extracted (green line). The inner-lining was then analyzed for any discontinuities (right) by plotting the gradient along z-axis. Discontinuities were found at locations whose gradient deviates more than a certain threshold from the reference line
Fig. 4Illustration on vessel analysis. The vessel extraction algorithm (top images) was started by manual selection of the starting point (white arrow—top left) inside the vessel lumen at location 1 depicted at Fig. 1, from where an ROI (blue rectangle) was selected. Inside the ROI (top centre), the lumen boundary (bold blue line with centre point at blue circle) was detected using GVF snake algorithm. The vessel was constructed from the detected vessel boundaries and centre points (top right) along z-axis. Afterwards, the smoothness of vessel centreline (bottom left), the consistency of vessel lumen area (bottom centre) and the consistency of vessel lumen mean value (bottom right) along z-axis were analyzed
Qualitative [9] and quantitative motion artefact analysis on 64CT
| Heart rate (bpm) | 64CT | ||||||
|---|---|---|---|---|---|---|---|
| Qualitative analysisa | Quantitative analysis | ||||||
| Myocardium inner-lining discontinuitiesb | Lumen area | Lumen value | |||||
| Cumulative area differences (%) | Artefact-induced lumen area stenoses segmentsc | Lumen mean value change(s)d | |||||
| Medium | Large | Medium | Large | ||||
| 0 | 4.0 ± 0 | – | – | – | – | 3 | – |
| 50 | 3.7 ± 0.8 | – | −3.1 | 1 | – | 2 | – |
| 60 | 3.5 ± 0.5 | – | −3.4 | 1 | – | 4 | – |
| 70 | 2.3 ± 0.5 | 3 | −4.9 | 1 | – | 1 | 1 (1) |
| 80 | 3.8 ± 0.4 | – | −3.5 | 1 | – | 2 | – |
| 90 | 3.0 ± 1.3 | – | −4.1 | 1 | – | 1 | – |
| 100 | 1.3 ± 0.5 | 7 (4) | −10.3 | 1 (1) | 2 (2) | 2 (1) | 1 (1) |
| 110 | 2.0 ± 0.6 | 7 (2) | −5.1 | 3 (3) | – | – | 1 (1) |
| Overall | 3.0 ± 1.1 | 17 (6) | −4.9 | 9 (4) | 2 (2) | 15 (1) | 3 (3) |
aThe value was taken from previous publication [9]. The value was given based on criteria listed in Table 3
bAmount of myocardium inner-lining discontinuities found by the quantitative analysis. The values between brackets indicate findings categorized as severe by visual observation
cAmount of vessel stenoses segments found by the quantitative analysis. The values between brackets indicate the amount of stenoses segments that coincide with myocardium artefacts
dAmount of vessel lumen mean value changes found by the quantitative analysis. The values between brackets indicate the amount of the changes that coincide with myocardium artefacts
Qualitative [9] and quantitative motion artefact analysis on DSCT
| Heart rate (bpm) | DSCT | ||||||
|---|---|---|---|---|---|---|---|
| Qualitative analysisa | Quantitative analysis | ||||||
| Myocardium Inner-lining discontinuitiesb | Lumen area | Lumen value | |||||
| Cumulative area differences (%) | Artefact-induced lumen area stenoses segmentsc | Lumen mean value change(s)d | |||||
| Medium | Large | Medium | Large | ||||
| 0 | 4.3 ± 0.5 | – | – | – | – | 3 | – |
| 50 | 4.0 ± 0.6 | 2 | −3.1 | 1 | – | 4 | 1 |
| 60 | 4.5 ± 0.5 | 1 | 0.4 | – | – | 3 | – |
| 70 | 3.8 ± 0.4 | 2 (1) | −3.9 | 2 (1) | – | 3 | 1 |
| 80 | 4.3 ± 0.5 | – | −4.8 | 1 | – | 2 | – |
| 90 | 4.5 ± 0.5 | – | −3.4 | – | – | 1 | – |
| 100 | 4.3 ± 0.5 | 3 | −2.7 | 1 (1) | – | 3 (1) | 1 |
| 110 | 3.8 ± 0.8 | 4 | −6.8 | 2 (2) | – | 4 | 3 (2) |
| Overall | 4.2 ± 0.6 | 12 (1) | −3.5 | 7 (4) | – | 23 (1) | 6 (2) |
aThe value was taken from previous publication [9]. The value was given based on criteria listed in Table 3
bAmount of myocardium inner-lining discontinuities found by the quantitative analysis. The values between brackets indicate findings categorized as severe by visual observation
cAmount of vessel stenoses segments found by the quantitative analysis. The values between brackets indicate the amount of stenoses segments that coincide with myocardium artefacts
dAmount of vessel lumen mean value changes found by the quantitative analysis. The values between brackets indicate the amount of the changes that coincide with myocardium artefacts
Definition of image quality scores [9]
| Score | Definition of image quality |
|---|---|
| 1 | Image with step artefacts and/or stripes throughout the image limiting evaluation of the coronary artery and pericardium |
| 2 | Image with step artefacts and/or stripes in part of the image that result in limited evaluation of the coronary artery and pericardium |
| 3 | Image with step artefacts and/or stripes which have minor implication on the evaluation of the coronary artery and pericardium |
| 4 | Image with minor motion artefacts not hampering the evaluation of the coronary artery and pericardium |
| 5 | Excellent image quality without motion artefacts |
Fig. 5Combination of vessel (IIb) and myocardium (Ia) analysis. The location of the detected myocardium artefacts are indicated by arrows A to D in the three-dimensional volume reconstruction view (left) and by red vertical lines A to D in the vessel lumen area consistency graph (right)
Recommendation to interpret findings
| Type of findings | Meaning | |
|---|---|---|
| Vessel stenosis | Myocardial artefact | |
| − | − | Normal vessel |
| + | − | True |
| − | + | Suspicious area of motion artefact |
| + | + | Possible artefact-induced |