| Literature DB >> 19703285 |
Rembert Koczulla1, Silvano Dragonieri, Robert Schot, Robert Bals, Stefanie A Gauw, Claus Vogelmeier, Klaus F Rabe, Peter J Sterk, Pieter S Hiemstra.
Abstract
BACKGROUND: Analysis of exhaled breath condensate (EBC) is a non-invasive method for studying the acidity (pH) of airway secretions in patients with inflammatory lung diseases. AIM: To assess the reproducibility of EBC pH for two commercially available devices (portable RTube and non-portable ECoScreen) in healthy controls, patients with asthma or COPD, and subjects suffering from an acute cold with lower-airway symptoms. In addition, we assessed the repeatability in healthy controls.Entities:
Mesh:
Year: 2009 PMID: 19703285 PMCID: PMC2747836 DOI: 10.1186/1465-9921-10-78
Source DB: PubMed Journal: Respir Res ISSN: 1465-9921
Clinical characteristics of the study population
| Healthy control | Asthma | COPD | Cold | |
|---|---|---|---|---|
| No. of patients | 10 | 10 | 10 | 10 |
| Age (y) | 34 ± 11.2 | 25.1 ± 5.9 | 60.8 ± 5.4 | 29.6 ± 6.4 |
| Sex (male/female) | 5/5 | 1/9 | 8/2 | 7/3 |
| Smoker/non smoker | 0/10 | 0/10 | 7/3* | 1/9 |
| FEV1 pre broncho. (% pred.) | ND | 99.9 ± 7.7 | 70 ± 14.8 | ND |
| FEV1 post broncho. (% pred.) | ND | 111.9 ± 9.2 | 8.5 ± 13.3 | ND |
| FVC (% pred.) | ND | 109 ± 8.1 | 68 ± 11.6 | ND |
* = ex smoker, ND = Not Determined; Values are expressed as means ± SDs
%pred. =% predicted, pre broncho = pre bronchodilator; post broncho = post bronchodilator
Figure 1Effect of duration of degasification on fresh and frozen EBC samples. EBC samples collected by RTube were obtained from 5 healthy controls, and either used fresh or after storage at -20°C (frozen). Prior to pH analysis, EBC samples were de-aerated using argon gas for various periods of time. Results show mean ± SD.
Figure 3Comparison of pH in EBC obtained by either RTube or ECoScreen in healthy controls (HC), asthma, COPD and cold patients (n = 40). The dashed line is the line of identity.
Figure 4Bland-Altman plot showing good agreement between pH values in EBC collected from 10 healthy subjects using ECoScreen and RTube on two different days. Horizontal lines show the mean and 95% confidence interval.
Figure 5Bland-Altman plot showing good agreement between pH values in EBC collected by ECoScreen and RTube from patients with asthma, COPD or a cold (n = 10 for each group). Horizontal lines show the mean and 95% confidence interval.
Figure 6Comparison of EBC pH values obtained by ECoScreen (left panel) or RTube (right panel) in healthy controls, and patients with asthma, COPD or a cold (n = 10 per group). Using both devices, pH values in patients with a cold were significantly lower than in healthy controls, whereas the other patient groups did not show a difference.