Differences in exhaled breath condensate pH measurements between samples obtained with two commercial devices.

L Prieto1, A Ferrer, J Palop, J Domenech, R Llusar, R Rojas.   

Abstract

BACKGROUND: The objective of this study was to determine differences in EBC pH between samples obtained by RTube and EcoScreen, and to identify the effect of storage at -80 degrees C on the pH values.
METHODS: Twenty-three nonsmoking subjects with asthma or allergic rhinitis or without respiratory disease performed two sequential exhaled breath condensate (EBC) collections, using the RTube collection system and the EcoScreen condenser. EBC pH was measured immediately after collection and after storage at -80 degrees C for 8 weeks, without deaeration and repeated following deaeration with Argon.
RESULTS: In EBC samples without deaeration, the EcoScreen pH values were significantly higher than the RTube pH values (mean difference, 0.32; 95% CI, 0.21-0.44, P < 0.0001). In deaerated EBC samples, the EcoScreen pH values were also significantly higher than the RTube pH values (mean difference, 0.12; 95% CI, 0.01-0.25, P=0.04). For both EBC collection systems, storage for 8 weeks had a significant influence on pH of nondeaerated samples.
CONCLUSIONS: The present study demonstrates that EBC pH value is dependent on the collection device used and that the storage for 8 weeks had a significant influence on the pH of samples analyzed without deaeration.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17481878     DOI: 10.1016/j.rmed.2007.02.023

Source DB:  PubMed          Journal:  Respir Med        ISSN: 0954-6111            Impact factor:   3.415


  11 in total

1.  Efficacy of two breath condensers.

Authors:  A Davidsson; B Schmekel
Journal:  J Clin Lab Anal       Date:  2010       Impact factor: 2.352

2.  Lipoxin A(4) and 8-isoprostane in the exhaled breath condensate of children hospitalized for status asthmaticus.

Authors:  Rashed A Hasan; Edmund OʼBrien; Peter Mancuso
Journal:  Pediatr Crit Care Med       Date:  2012-03       Impact factor: 3.624

Review 3.  Measures of asthma control.

Authors:  Christian Bime; Jessica Nguyen; Robert A Wise
Journal:  Curr Opin Pulm Med       Date:  2012-01       Impact factor: 3.155

4.  Biomarkers in exhaled breath condensate: a review of collection, processing and analysis.

Authors:  N M Grob; M Aytekin; R A Dweik
Journal:  J Breath Res       Date:  2008-09-08       Impact factor: 3.262

5.  Effect of allergen-specific immunotherapy with purified Alt a1 on AMP responsiveness, exhaled nitric oxide and exhaled breath condensate pH: a randomized double blind study.

Authors:  Luis Prieto; Ricardo Palacios; Dulce Aldana; Anna Ferrer; Carmen Perez-Frances; Victoria Lopez; Rocio Rojas
Journal:  Allergy Asthma Clin Immunol       Date:  2010-09-16       Impact factor: 3.406

6.  Analysis of nitrogen oxides (NOx) in the exhaled breath condensate (EBC) of subjects with asthma as a complement to exhaled nitric oxide (FeNO) measurements: a cross-sectional study.

Authors:  Nathalie Chérot-Kornobis; Sébastien Hulo; Jean-Louis Edmé; Virginie de Broucker; Régis Matran; Annie Sobaszek
Journal:  BMC Res Notes       Date:  2011-06-16

7.  Comparison of two devices and two breathing patterns for exhaled breath condensate sampling.

Authors:  Eva-Maria Hüttmann; Timm Greulich; Akira Hattesohl; Severin Schmid; Sarah Noeske; Christian Herr; Gerrit John; Rudolf A Jörres; Bernd Müller; Claus Vogelmeier; Andreas Rembert Koczulla
Journal:  PLoS One       Date:  2011-11-07       Impact factor: 3.240

8.  Comparison of exhaled breath condensate pH using two commercially available devices in healthy controls, asthma and COPD patients.

Authors:  Rembert Koczulla; Silvano Dragonieri; Robert Schot; Robert Bals; Stefanie A Gauw; Claus Vogelmeier; Klaus F Rabe; Peter J Sterk; Pieter S Hiemstra
Journal:  Respir Res       Date:  2009-08-24

9.  Comparative analysis of selected exhaled breath biomarkers obtained with two different temperature-controlled devices.

Authors:  Frank Hoffmeyer; Monika Raulf-Heimsoth; Volker Harth; Jürgen Bünger; Thomas Brüning
Journal:  BMC Pulm Med       Date:  2009-11-30       Impact factor: 3.317

10.  Exhaled breath marker in asthma patients with gastroesophageal reflux disease.

Authors:  Yasuo Shimizu; Kunio Dobashi; Masatomo Mori
Journal:  J Clin Biochem Nutr       Date:  2007-11       Impact factor: 3.114

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.