Literature DB >> 19673219

Field size dependence of the output factor in passively scattered proton therapy: influence of range, modulation, air gap, and machine settings.

J Daartz1, M Engelsman, Harald Paganetti, M R Bussière.   

Abstract

At the Francis H. Burr Proton Therapy Center field specific output factors (i.e., dose per monitor unit) for patient treatments were modeled for all beamlines (two gantries, fixed stereotactic, and fixed eye beamline). The authors evaluated the accuracy of dose calculation and output model for small fields. Measurements in a water phantom were performed in three of our beamlines quantifying the dependency of the output factor on the field size for a variety of proton ranges. The influence of snout size, air gap, modulation, and second scatterer was investigated. The impact of field size on output depends strongly on the depth of interest. The air gap has a notable influence on small field outputs. A field size specific correction factor to the output is necessary if the latter was modeled or measured without the custom hardware in place. The output was shown to be field size dependent even for large fields, indicating an effect beyond charged particle disequilibrium caused by lateral scatter.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19673219     DOI: 10.1118/1.3152111

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  12 in total

1.  Dosimetric accuracy of planning and delivering small proton therapy fields.

Authors:  Bryan Bednarz; Juliane Daartz; Harald Paganetti
Journal:  Phys Med Biol       Date:  2010-11-19       Impact factor: 3.609

2.  Commissioning a passive-scattering proton therapy nozzle for accurate SOBP delivery.

Authors:  M Engelsman; H M Lu; D Herrup; M Bussiere; H M Kooy
Journal:  Med Phys       Date:  2009-06       Impact factor: 4.071

3.  TOPAS: an innovative proton Monte Carlo platform for research and clinical applications.

Authors:  J Perl; J Shin; J Schumann; B Faddegon; H Paganetti
Journal:  Med Phys       Date:  2012-11       Impact factor: 4.071

4.  Learning-based CBCT correction using alternating random forest based on auto-context model.

Authors:  Yang Lei; Xiangyang Tang; Kristin Higgins; Jolinta Lin; Jiwoong Jeong; Tian Liu; Anees Dhabaan; Tonghe Wang; Xue Dong; Robert Press; Walter J Curran; Xiaofeng Yang
Journal:  Med Phys       Date:  2018-12-11       Impact factor: 4.071

5.  Assessing the Clinical Impact of Approximations in Analytical Dose Calculations for Proton Therapy.

Authors:  Jan Schuemann; Drosoula Giantsoudi; Clemens Grassberger; Maryam Moteabbed; Chul Hee Min; Harald Paganetti
Journal:  Int J Radiat Oncol Biol Phys       Date:  2015-04-08       Impact factor: 7.038

6.  Site-specific range uncertainties caused by dose calculation algorithms for proton therapy.

Authors:  J Schuemann; S Dowdell; C Grassberger; C H Min; H Paganetti
Journal:  Phys Med Biol       Date:  2014-07-03       Impact factor: 3.609

7.  Quantification of proton dose calculation accuracy in the lung.

Authors:  Clemens Grassberger; Juliane Daartz; Stephen Dowdell; Thomas Ruggieri; Greg Sharp; Harald Paganetti
Journal:  Int J Radiat Oncol Biol Phys       Date:  2014-04-11       Impact factor: 7.038

8.  Feasibility of using PRESAGE® for relative 3D dosimetry of small proton fields.

Authors:  Li Zhao; Joseph Newton; Mark Oldham; Indra J Das; Chee-Wai Cheng; John Adamovics
Journal:  Phys Med Biol       Date:  2012-10-26       Impact factor: 3.609

9.  Recommendations for the referral of patients for proton-beam therapy, an Alberta Health Services report: a model for Canada?

Authors:  S Patel; X Kostaras; M Parliament; I A Olivotto; R Nordal; K Aronyk; N Hagen
Journal:  Curr Oncol       Date:  2014-10       Impact factor: 3.677

10.  Implementation of an improved dose-per-MU model for double-scattered proton beams to address interbeamline modulation width variability.

Authors:  Liyong Lin; JiaJian Shen; Christopher G Ainsley; Timothy D Solberg; James E McDonough
Journal:  J Appl Clin Med Phys       Date:  2014-05-08       Impact factor: 2.102