MRI of the knee at 3T: first clinical results with an isotropic PDfs-weighted 3D-TSE-sequence.

Mike Notohamiprodjo1, Annie Horng, Matthias F Pietschmann, Peter E Müller, Wilhelm Horger, Jaeseok Park, Alexander Crispin, José Raya Garcia del Olmo, Sabine Weckbach, Karin A Herrmann, Maximilian F Reiser, Christian Glaser.   

Abstract

PURPOSE: To clinically evaluate MRI of the knee using a highly resolved isotropic fat-saturated (fs) proton-density weighted 3D-TSE-sequence (SPACE) at 3T.
MATERIALS AND METHODS: Imaging was performed on a 3T-scanner (Magnetom TRIO). For technical evaluation, sagittally orientated SPACE-datasets (repetition-time [TR], 1200 milliseconds/[TE], 30 milliseconds/voxel-size, 0.5 mm3/acquisition time, 10:35 minutes) were acquired from the dominant knee of 10 healthy volunteers. In the 3 major anatomic planes, 0.5, 1, and 2 mm thick reconstructions were performed. Signal-to-noise (SNR), SNR-efficiency, contrast-to-noise (CNR) ratios, and anatomic detail visualization were compared with a state-of-the-art 2D-TSE-sequence in 3 imaging planes (TR, 3200 milliseconds/TE, 30 milliseconds/acquisition time, 12:34 minutes). Sixty patients with cartilage and meniscus pathologies were examined with these techniques. Patient SPACE-datasets were assessed in 1-mm thick reconstructions. Arthroscopical correlation was available for 18 patients. Lesion detection and diagnostic confidence were assessed by 2 radiologists independently. Statistical analysis was performed using 95% confidence intervals, Wilcoxon signed rank tests, and Weighted-kappa.
RESULTS: SNR-efficiency of SPACE was 4 to 5 times higher than for 2D-TSE-sequences. SNR and CNR of 1-mm thick SPACE-reconstructions were comparable to 2D-TSE-sequences and provided superior visualization of small structures such as meniscal roots.Correlation with arthroscopy did not show significant differences between 2D- and 3D-sequences. One reader detected significantly more cartilage abnormalities with the 2D-TSE-sequence (131 vs. 151, P = 0.04), probably because of an unfamiliar fluid/cartilage contrast. Diagnostic confidence was significantly higher for meniscus abnormalities for SPACE for 1 reader. Intersequence-correlation was excellent (kappa = 0.82-0.92). Interreader-correlation was good to excellent (kappa = 0.71-0.80), intrareader-correlation was excellent (kappa = 0.90-0.92) for both sequences.
CONCLUSIONS: Time-efficient 3D-TSE-imaging of the knee at 3T is feasible with adequate SNR and CNR and excellent anatomic detail visualization. Detection and visualization of meniscus and cartilage pathologies is comparable to standard 2D-TSE-sequences. 3D-TSE-sequences with consecutive multiplanar reconstruction may become a valuable component of future knee-MRI protocols.

Entities:  

Mesh:

Year:  2009        PMID: 19668001     DOI: 10.1097/RLI.0b013e3181b4c1a1

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  29 in total

1.  Magnetic resonance diagnosis of posterior horn tears of the lateral meniscus using a thin axial plane: the zip sign--a preliminary study.

Authors:  P Y Savoye; J N Ravey; C Dubois; L Pittet Barbier; A Courvoisier; D Saragaglia; G Ferretti
Journal:  Eur Radiol       Date:  2010-07-20       Impact factor: 5.315

2.  Chronic anterior cruciate ligament tears and associated meniscal and traumatic cartilage lesions: evaluation with morphological sequences at 3.0 T.

Authors:  Marianna Vlychou; Michalis Hantes; Sotirios Michalitsis; Aspasia Tsezou; Ioannis V Fezoulidis; Konstantinos Malizos
Journal:  Skeletal Radiol       Date:  2010-10-27       Impact factor: 2.199

3.  Highlights of the European Society of Musculoskeletal Radiology (ESSR) annual meeting 2009.

Authors:  Andrew J Grainger
Journal:  Skeletal Radiol       Date:  2010-01       Impact factor: 2.199

4.  Optimizing isotropic three-dimensional fast spin-echo methods for imaging the knee.

Authors:  Charles Q Li; Weitian Chen; Jarrett K Rosenberg; Philip J Beatty; Richard Kijowski; Brian A Hargreaves; Reed F Busse; Garry E Gold
Journal:  J Magn Reson Imaging       Date:  2014-06       Impact factor: 4.813

5.  Diagnostic performance of wrist magnetic resonance (MR) arthrography: comparison of three-dimensional isotropic T1 weighted fast spin-echo MR arthrography and two-dimensional MR arthrography.

Authors:  Eunsun Oh; Young C Yoon; Min J Park
Journal:  Br J Radiol       Date:  2017-02-15       Impact factor: 3.039

6.  MR arthrography of the hip: evaluation of isotropic 3D intermediate-weighted FSE and hybrid GRE T1-weighted sequences.

Authors:  Giovanni Foti; Antonio Campacci; Michele Conati; Mirko Trentadue; Claudio Zorzi; Giovanni Carbognin
Journal:  Radiol Med       Date:  2017-06-09       Impact factor: 3.469

7.  CAIPIRINHA accelerated SPACE enables 10-min isotropic 3D TSE MRI of the ankle for optimized visualization of curved and oblique ligaments and tendons.

Authors:  Vivek Kalia; Benjamin Fritz; Rory Johnson; Wesley D Gilson; Esther Raithel; Jan Fritz
Journal:  Eur Radiol       Date:  2017-01-23       Impact factor: 5.315

8.  CAIPIRINHA-accelerated 10-min 3D TSE MRI of the ankle for the diagnosis of painful ankle conditions: Performance evaluation in 70 patients.

Authors:  Benjamin Fritz; Susanne Bensler; Gaurav K Thawait; Esther Raithel; Steven E Stern; Jan Fritz
Journal:  Eur Radiol       Date:  2018-07-23       Impact factor: 5.315

9.  Can a single isotropic 3D fast spin echo sequence replace three-plane standard proton density fat-saturated knee MRI at 1.5 T?

Authors:  B Pass; P Robinson; R Hodgson; A J Grainger
Journal:  Br J Radiol       Date:  2015-06-12       Impact factor: 3.039

10.  SLAP tears: diagnosis using 3-T shoulder MR arthrography with the 3D isotropic turbo spin-echo space sequence versus conventional 2D sequences.

Authors:  Joon-Yong Jung; Won-Hee Jee; Michael Yong Park; So-Yeon Lee; Yang-Soo Kim
Journal:  Eur Radiol       Date:  2012-08-04       Impact factor: 5.315