Dosimetric characteristics of 6 and 10MV unflattened photon beams.

Gabriele Kragl1, Sacha af Wetterstedt, Barbara Knäusl, Mårten Lind, Patrick McCavana, Tommy Knöös, Brendan McClean, Dietmar Georg.   

Abstract

PURPOSE: To determine dosimetric properties of unflattened megavoltage photon beams.
MATERIALS AND METHODS: Dosimetric data including depth dose, profiles, output factors and phantom scatter factors from three different beam qualities provided by Elekta Precise linacs, operated with and without flattening filter were examined. Additional measurements of leaf transmission, leakage radiation and surface dose were performed. In flattening filter free (FFF) mode a 6-mm thick copper filter was placed into the beam to stabilize it.
RESULTS: Depths of dose maxima for flattened and unflattened beams did not deviate by more than 2mm and penumbral widths agreed within 1mm. In FFF mode the collimator exchange effect was found to be on average 0.3% for rectangular fields. Between maximum and minimum field size head scatter factors of unflattened beams showed on average 40% and 56% less variation for 6 and 10MV beams than conventional beams. Phantom scatter factors for FFF beams differed up to 4% from the published reference data. For field sizes smaller than 15cm, surface doses relative to the dose at d(max) increased for unflattened beams with maximum differences of 7% at 6MV and 25% at 10MV for a 5x5cm(2) field. For a 30x30cm(2) field, relative surface dose decreased by about 10% for FFF beams. Leaf transmission on the central axis was 0.3% and 0.4% lower for unflattened 6 and 10MV beams, respectively. Leakage radiation was reduced by 52% for 6MV and by 65% for 10MV unflattened beams.
CONCLUSIONS: The results of the study were independently confirmed at two radiotherapy centres. Phantom scatter reference data need to be reconsidered for medical accelerators operated without a flattening filter.

Entities:  

Mesh:

Year:  2009        PMID: 19592123     DOI: 10.1016/j.radonc.2009.06.008

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  60 in total

1.  Tomographic Cherenkov-excited luminescence scanned imaging with multiple pinhole beams recovered via back-projection reconstruction.

Authors:  Mengyu J Jia; Xu Cao; Jason R Gunn; Petr Bruza; Shudong Jiang; Brian W Pogue
Journal:  Opt Lett       Date:  2019-04-01       Impact factor: 3.776

2.  Comparison of surface dose delivered by 7 MV-unflattened and 6 MV-flattened photon beams.

Authors:  Ashokkumar Sigamani; Arunai Nambiraj
Journal:  Rep Pract Oncol Radiother       Date:  2017-04-25

3.  Dosimetric influence of filtered and flattening filter free photon beam on rapid arc (RA) radiotherapy planning in case of cervix carcinoma.

Authors:  Lalit Kumar; Girigesh Yadav; Kothanda Raman Samuvel; Manindra Bhushan; Pawan Kumar; Mahammood Suhail; Manoj Pal
Journal:  Rep Pract Oncol Radiother       Date:  2016-10-17

4.  Stereotactic ablative radiation therapy for brain metastases with volumetric modulated arc therapy and flattening filter free delivery: feasibility and early clinical results.

Authors:  Alba Fiorentino; Niccolò Giaj-Levra; Umberto Tebano; Rosario Mazzola; Francesco Ricchetti; Sergio Fersino; Gioacchino Di Paola; Dario Aiello; Ruggero Ruggieri; Filippo Alongi
Journal:  Radiol Med       Date:  2017-04-26       Impact factor: 3.469

5.  Analysis of physical parameters and determination of inflection point for Flattening Filter Free beams in medical linear accelerator.

Authors:  A Pichandi; Kadirampatti Mani Ganesh; Amalraj Jerin; Karunakaran Balaji; Gurunath Kilara
Journal:  Rep Pract Oncol Radiother       Date:  2014-07-15

6.  Initial experience of hypofractionated radiation retreatment with true beam and flattening filter free beam in selected case reports of recurrent nasopharyngeal carcinoma.

Authors:  Filippo Alongi; Elena Clerici; Sara Pentimalli; Pietro Mancosu; Marta Scorsetti
Journal:  Rep Pract Oncol Radiother       Date:  2012-08-10

7.  Technical Report: Evaluation of peripheral dose for flattening filter free photon beams.

Authors:  E L Covington; T A Ritter; J M Moran; A M Owrangi; J I Prisciandaro
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

8.  Dosimetric comparison of intracranial metastasis treatment using two radiosurgery systems: TrueBeam STx with VMAT and Gamma Knife Model 4C.

Authors:  Vance Keeling; Ozer Algan; Salahuddin Ahmad; Sabbir Hossain
Journal:  J Radiosurg SBRT       Date:  2016

9.  Optimizing the flattening filter free beam selection in RapidArc®-based stereotactic body radiotherapy for Stage I lung cancer.

Authors:  J-Y Lu; Z Lin; P-X Lin; B-T Huang
Journal:  Br J Radiol       Date:  2015-07-02       Impact factor: 3.039

10.  Radiosurgery with flattening-filter-free techniques in the treatment of brain metastases : Plan comparison and early clinical evaluation.

Authors:  J Rieber; E Tonndorf-Martini; O Schramm; B Rhein; S Stefanowicz; J Kappes; H Hoffmann; K Lindel; J Debus; S Rieken
Journal:  Strahlenther Onkol       Date:  2016-07-04       Impact factor: 3.621

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.