Dual-source versus 64-section CT coronary angiography at lower heart rates: comparison of accuracy and radiation dose.

Stephan Baumüller1, Sebastian Leschka, Lotus Desbiolles, Paul Stolzmann, Hans Scheffel, Burkhardt Seifert, Borut Marincek, Hatem Alkadhi.   

Abstract

PURPOSE: To compare the diagnostic performance and radiation doses of dual-source and 64-section computed tomographic (CT) coronary angiography for the diagnosis of significant coronary stenoses in patients with heart rates of 65 beats/min or less.
MATERIALS AND METHODS: This retrospective study had local ethics committee approval; all patients gave written informed consent. Two hundred patients with heart rates of 65 beats/min or less were enrolled; 100 underwent dual-source and 100 underwent 64-section CT coronary angiography. Two blinded observers independently assessed image quality of all coronary segments by using a four-point scale and searched for significant (>50%) stenoses in each segment. Catheter angiography was used as the reference standard. Image noise was measured in the ascending aorta. Radiation doses were calculated.
RESULTS: No significant differences were found regarding sex, age, body weight, cardiovascular risk profile, prevalence of stenosis, mean and variability of heart rate, Agatston score, and image noise (all P > .1) between patients in both CT coronary angiography groups. No significant difference was found in the rate of nonassessable coronary segments between dual-source (1.0%, 14 of 1405) and 64-section CT coronary angiography (1.8%, 25 of 1387; P = .08). Motion artifacts occurred significantly more often in 64-section (21 of 25) versus dual-source (five of 14, P = .004) CT coronary angiography. Segment-based accuracy and specificity were significantly higher for dual-source versus 64-section CT coronary angiography. There was no significant difference in accuracy parameters at the per-vessel and per-patient analyses. No significant difference (P = .13) was found between the effective doses of dual-source (mean +/- standard deviation, 10.9 mSv +/- 1.1) and 64-section CT (10.4 mSv +/- 1.7) coronary angiography.
CONCLUSION: In patients with heart rates of 65 beats/min or less, the higher temporal resolution of dual-source CT coronary angiography results in improved accuracy and specificity for the diagnosis of significant stenoses on a per-segment level at a similar radiation dose, but provides a comparable diagnostic accuracy on a patient-based level as does 64-section coronary angiography.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19587311     DOI: 10.1148/radiol.2531090065

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  20 in total

1.  Examination of the optimal temporal resolution required for computed tomography coronary angiography.

Authors:  Kazuya Ohashi; Katsuhiro Ichikawa; Masaki Hara; Tatsuya Kawai; Hiroshi Kunitomo; Ryo Higashide; Yuta Shibamoto
Journal:  Radiol Phys Technol       Date:  2013-05-26

2.  Indications, imaging technique, and reading of cardiac computed tomography: survey of clinical practice.

Authors:  M H Maurer; E Zimmermann; P Schlattmann; C Germershausen; B Hamm; Marc Dewey
Journal:  Eur Radiol       Date:  2011-08-16       Impact factor: 5.315

3.  Cardiac CT imaging in coronary artery disease: Current status and future directions.

Authors:  Zhonghua Sun
Journal:  Quant Imaging Med Surg       Date:  2012-06

4.  320-detector row CT coronary angiography: effects of heart rate and heart rate variability on image quality, diagnostic accuracy and radiation exposure.

Authors:  G Sun; M Li; X-S Jiang; L Li; Z-H Peng; G-Y Li; L Xu
Journal:  Br J Radiol       Date:  2012-02-28       Impact factor: 3.039

5.  Comparison of radiation dose and image quality: 320-MDCT versus 64-MDCT coronary angiography.

Authors:  Atif Khan; Faisal Khosa; Khurram Nasir; Aya Yassin; Melvin E Clouse
Journal:  AJR Am J Roentgenol       Date:  2011-07       Impact factor: 3.959

6.  The culprit lesion and its consequences: combined visualization of the coronary arteries and delayed myocardial enhancement in dual-source CT: a pilot study.

Authors:  Andreas H Mahnken; Philipp Bruners; Ola Friman; Anja Hennemuth
Journal:  Eur Radiol       Date:  2010-06-23       Impact factor: 5.315

7.  Low-dose CT and cardiac MR for the diagnosis of coronary artery disease: accuracy of single and combined approaches.

Authors:  Hans Scheffel; Paul Stolzmann; Hatem Alkadhi; Naim Azemaj; André Plass; Stephan Baumueller; Lotus Desbiolles; Sebastian Leschka; Sebastian Kozerke; Volkmar Falk; Peter Boesiger; Christophe Wyss; Borut Marincek; Olivio F Donati
Journal:  Int J Cardiovasc Imaging       Date:  2010-02-10       Impact factor: 2.357

8.  Dual-source CT coronary angiography: prospective versus retrospective acquisition technique.

Authors:  C N De Cecco; V Buffa; S Fedeli; A Vallone; R Ruopoli; M Luzietti; V Miele; M Maurizi Enrici; F Musumeci; V David
Journal:  Radiol Med       Date:  2010-10-27       Impact factor: 3.469

9.  Myocardial bridging of the left anterior descending coronary artery: depiction rate and morphologic features by dual-source CT coronary angiography.

Authors:  Jin Ho Hwang; Sung Min Ko; Hong Gee Roh; Meong Gun Song; Je Kyoun Shin; Hyun Kun Chee; Joon Suk Kim
Journal:  Korean J Radiol       Date:  2010-08-27       Impact factor: 3.500

10.  Diagnostic quality of dual-source coronary CT examinations performed without heart rate control: importance of obesity and heart rate on image quality.

Authors:  Stefan L Zimmerman; Brian G Kral; Elliot K Fishman
Journal:  J Comput Assist Tomogr       Date:  2014 Nov-Dec       Impact factor: 1.826

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.