Influence of selecting EPI readout-encoding bandwidths on arterial spin labeling perfusion MRI.

Geon-Ho Jahng1, Norbert Schuff.   

Abstract

OBJECT: The objective of this study was to investigate effects of varying readout bandwidths on the arterial spin labeling (ASL)-perfusion MRI measurements at a high magnetic field MRI system.
MATERIALS AND METHODS: Brain perfusion studies were performed on nine volunteers (four males, five females) using flow sensitive alternating inversion recovery (FAIR) ASL single-shot echo-planar imaging (EPI)-MRI. To investigate EPI bandwidth effects on the time-series perfusion-weighted imaging (PWI) data, two regions-of-interest (ROI) were placed outside the brain to determine the level of noise and another ROI inside the brain to determine the level of signal. Coefficients of variations (CoV) were calculated for the time-series PWI data. One-way analysis of variance (ANOVA) was used to investigate voxel-wise differences in the time-series PWI data between two different bandwidth values.
RESULTS: At the level of ROI, there was no significant effect of changing EPI bandwidths on the time-series PWI data in any of the volunteers (P > 0.031). In contrast, CoV values over the dynamic PWI data varied with depending on selecting EPI bandwidths and voxel-based tests showed that N2 ghosting, modulated by EPI bandwidth, can appear in some brain regions, especially in areas that overlap with the spatial distribution of N2 ghosting artifacts.
CONCLUSIONS: Although N2 ghosting can be reduced by adjusting the bandwidth of EPI on the time-series of PWI data, the effects cannot be entirely eliminated. In particular, N2 ghosting can bias CBF quantification if EPI control scans to determine the equilibrium-state signal are confounded by N2 ghosting. Therefore, careful tuning of the bandwidth of EPI is necessary to avoid artifacts in the ASL signal from N2-ghosting.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19578915     DOI: 10.1007/s10334-009-0174-2

Source DB:  PubMed          Journal:  MAGMA        ISSN: 0968-5243            Impact factor:   2.310


  44 in total

1.  Investigation of acoustic noise on 15 MRI scanners from 0.2 T to 3 T.

Authors:  D L Price; J P De Wilde; A M Papadaki; J S Curran; R I Kitney
Journal:  J Magn Reson Imaging       Date:  2001-02       Impact factor: 4.813

2.  MRI receiver frequency response as a contributor to Nyquist ghosting in echo planar imaging.

Authors:  Ioannis Delakis; Krystallia Petala; Janet P De Wilde
Journal:  J Magn Reson Imaging       Date:  2005-08       Impact factor: 4.813

3.  Influence of gradient acoustic noise on fMRI response in the human visual cortex.

Authors:  Nanyin Zhang; Xiao-Hong Zhu; Wei Chen
Journal:  Magn Reson Med       Date:  2005-08       Impact factor: 4.668

4.  Slice profile optimization in arterial spin labeling using presaturation and optimized RF pulses.

Authors:  David Alberg Holm; Karam Sidaros
Journal:  Magn Reson Imaging       Date:  2006-09-25       Impact factor: 2.546

5.  Empirical optimization of ASL data analysis using an ASL data processing toolbox: ASLtbx.

Authors:  Ze Wang; Geoffrey K Aguirre; Hengyi Rao; Jiongjiong Wang; María A Fernández-Seara; Anna R Childress; John A Detre
Journal:  Magn Reson Imaging       Date:  2007-09-10       Impact factor: 2.546

6.  Quantification of relative cerebral blood flow change by flow-sensitive alternating inversion recovery (FAIR) technique: application to functional mapping.

Authors:  S G Kim
Journal:  Magn Reson Med       Date:  1995-09       Impact factor: 4.668

7.  Artifact reduction in EPI with phase-encoded reference scan.

Authors:  X Hu; T H Le
Journal:  Magn Reson Med       Date:  1996-07       Impact factor: 4.668

8.  Accurate T1 determination from inversion recovery images: application to human brain at 4 Tesla.

Authors:  S G Kim; X Hu; K Uğurbil
Journal:  Magn Reson Med       Date:  1994-04       Impact factor: 4.668

9.  Optimized EPI for fMRI studies of the orbitofrontal cortex: compensation of susceptibility-induced gradients in the readout direction.

Authors:  Nikolaus Weiskopf; Chloe Hutton; Oliver Josephs; Robert Turner; Ralf Deichmann
Journal:  MAGMA       Date:  2007-02-01       Impact factor: 2.310

10.  Susceptibility-matched envelope for the correction of EPI artifacts.

Authors:  Arnon Neufeld; Yaniv Assaf; Moshe Graif; Talma Hendler; Gil Navon
Journal:  Magn Reson Imaging       Date:  2005-10-28       Impact factor: 2.546

View more
  2 in total

1.  Magnetic resonance perfusion imaging evaluation in perfusion abnormalities of the cerebellum after supratentorial unilateral hyperacute cerebral infarction.

Authors:  Pan Liang; Yunjun Yang; Weijian Chen; Yuxia Duan; Hongqing Wang; Xiaotong Wang
Journal:  Neural Regen Res       Date:  2012-04-25       Impact factor: 5.135

2.  Characterization of Nyquist ghost in EPI-fMRI acquisition sequences implemented on two clinical 1.5 T MR scanner systems: effect of readout bandwidth and echo spacing.

Authors:  Marco Giannelli; Stefano Diciotti; Carlo Tessa; Mario Mascalchi
Journal:  J Appl Clin Med Phys       Date:  2010-07-12       Impact factor: 2.102