The incidence of physiological radiolucency following Oxford unicompartmental knee replacement and its relationship to outcome.

A Gulati1, R Chau, H G Pandit, H Gray, A J Price, C A F Dodd, D W Murray.   

Abstract

Narrow, well-defined radiolucent lines commonly observed at the bone-implant interface of unicompartmental knee replacement tibial components have been referred to as physiological radiolucencies. These should be distinguished from pathological radiolucencies, which are poorly defined, wide and progressive, and associated with loosening and infection. We studied the incidence and clinical significance of tibial radiolucent lines in 161 Oxford unicondylar knee replacements five years after surgery. All the radiographs were aligned with fluoroscopic control to obtain views parallel to the tibial tray to reveal the tibial bone-implant interface. We found that 49 knees (30%) had complete, 52 (32%) had partial and 60 (37%) had no radiolucent lines. There was no relationship between the incidence of radiolucent lines and patient factors such as gender, body mass index and activity, or operative factors including the status of the anterior cruciate ligament and residual varus deformity. Nor was any statistical relationship established between the presence of radiolucent lines and clinical outcome, particularly pain, assessed by the Oxford Knee score and the American Knee Society score. We conclude that radiolucent lines are common after Oxford unicompartmental knee replacement but that their aetiology remains unclear. Radiolucent lines were not a source of adverse symptoms or pain. Therefore, when attempting to identify a source of postoperative pain after Oxford unicompartmental knee replacement the presence of a physiological radiolucency should be ignored.

Entities:  

Mesh:

Year:  2009        PMID: 19567853     DOI: 10.1302/0301-620X.91B7.21914

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  42 in total

1.  Early migration of the cemented tibial component of unicompartmental knee arthroplasty: a radiostereometry study.

Authors:  Andrea Ensini; Paolo Barbadoro; Alberto Leardini; Fabio Catani; Sandro Giannini
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-06-04       Impact factor: 4.342

2.  The Oxford unicompartmental knee fails at a high rate in a high-volume knee practice.

Authors:  William C Schroer; C Lowry Barnes; Paul Diesfeld; Angela LeMarr; Rachel Ingrassia; Diane J Morton; Mary Reedy
Journal:  Clin Orthop Relat Res       Date:  2013-08-02       Impact factor: 4.176

3.  CORR Insights ®: The Oxford unicompartmental knee fails at a high rate in a high-volume knee practice.

Authors:  John M Clark
Journal:  Clin Orthop Relat Res       Date:  2013-08-24       Impact factor: 4.176

4.  Medial femoral head border is a reliable and reproducible reference for axis determination for femoral component of unicompartmental knee arthroplasty.

Authors:  T Hiranaka; H Pandit; H S Gill; Y Hida; H Uemoto; M Doita; M Tsuji; D W Murray
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-09-28       Impact factor: 4.342

5.  A semi-automated measurement technique for the assessment of radiolucency.

Authors:  E C Pegg; B J L Kendrick; H G Pandit; H S Gill; D W Murray
Journal:  J R Soc Interface       Date:  2014-04-23       Impact factor: 4.118

Review 6.  Cementless fixation in medial unicompartmental knee arthroplasty: a systematic review.

Authors:  S Campi; H G Pandit; C A F Dodd; D W Murray
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-07-19       Impact factor: 4.342

7.  Cementless versus cemented Oxford unicompartmental knee arthroplasty: early results of a non-designer user group.

Authors:  B Kerens; M G M Schotanus; B Boonen; P Boog; P J Emans; H Lacroix; N P Kort
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-05-09       Impact factor: 4.342

8.  Proximal tibial bone density is preserved after unicompartmental knee arthroplasty.

Authors:  Bradley I Richmond; Simon V Hadlow; Tim G Lynskey; Cameron G Walker; Jacob T Munro
Journal:  Clin Orthop Relat Res       Date:  2013-01-17       Impact factor: 4.176

9.  Assessment of radiolucent lines in patients with lateral unicompartmental knee arthroplasty and the relationship between these lines and the outcome.

Authors:  Long Xue; Huaming Xue; Tao Wen; Mengying Guan; Tao Yang; Tong Ma; Yihui Tu
Journal:  Int Orthop       Date:  2021-06-26       Impact factor: 3.075

10.  Five-year experience of cementless Oxford unicompartmental knee replacement.

Authors:  H G Pandit; S Campi; T W Hamilton; O D Dada; S Pollalis; C Jenkins; C A F Dodd; D W Murray
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2015-11-26       Impact factor: 4.342

View more