Nimalie Jacintha Perera1, Lynda Molyneaux, Maria Ines Constantino, Marg McGill, Dennis Koon-See Yue, Stephen Morris Twigg, Glynis Pauline Ross.
Abstract
OBJECTIVE: The objective of this study was to evaluate the performance of blood glucose meters in diabetes associated with pregnancy (DP). RESEARCH DESIGN AND METHODS: Finger-prick blood glucose levels measured using six different glucose meters on 102 patients with DP attending an antenatal clinic were compared with laboratory plasma glucose results. HbA(1c) and hematocrit were also measured.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21216849 PMCID: PMC3024344 DOI: 10.2337/dc10-1457
Source DB: PubMed Journal: Diabetes Care ISSN: 0149-5992 Impact factor: 19.112
Difference between blood glucose meters and plasma blood glucose
| Meters | Mean difference | SD (mmol/L) | 95% Limits of agreement ± 2 SD | Mean bias (%) | Mean total analytical error
%bias +1.96 | Coefficient of variation between meters (%) | |
|---|---|---|---|---|---|---|---|
| ACA | 0.361 | 0.55 | −0.749 | 1.470 | 8.99 | 14.87 | −0.220 |
| OPT20s | 0.572 | 0.66 | −0.747 | 1.890 | 13.08 | 26.21 | 2.597 |
| ACP | 0.385 | 0.57 | −0.758 | 1.528 | 9.04 | 15.65 | −0.299 |
| OPT5s | 0.725 | 0.62 | −0.523 | 1.972 | 15.76 | 32.14 | −0.796 |
| FSL | 0.232 | 0.69 | −1.157 | 1.622 | 6.37 | 17.07 | 1.416 |
| STAT | 0.256 | 0.54 | −0.817 | 1.329 | 6.10 | 12.29 | −0.623 |
ADA stipulates the most stringent goals, aiming for an analytical error <5% or total error (user plus analytical) <10%. All meters were calibrated to provide plasma glucose results, except one that reported capillary whole blood glucose, and adjusted plasma values were calculated as whole blood glucose = plasma glucose × [1 − (0.0024 × HCT)] for the Accu-Chek Advantage-II meter (13).
ACA, Accu-Chek Advantage-II (Roche Diagnostics, Mannheim, Germany) strips Lot No.450808; OPT20s, OptiumXceed 20 s strips Lot No. 54277 (Abbott Diabetes Care, Alameda, CA); ACP, Accu-Chek Performa (Roche Diagnostics) strips Lot No. 320151; OPT5s, OptiumXceed 5 s strips Lot No. 99884–5 (Abbott Diabetes Care); FSL, FreeStyle Lite (Abbott Diabetes Care) strips Lot No. 0820633; STAT, Stat-Strip (Nova Biomedical, Waltham, MA) strips Lot No. 0308273249.
*Mean difference = (mean of the meters − plasma glucose).
†[(meter − plasma glucose/plasma glucose)] × 100.