Influence of implant connection type on the biomechanical environment of immediately placed implants - CT-based nonlinear, three-dimensional finite element analysis.

Roberto S Pessoa1, Luiza Muraru, Elcio Marcantonio Júnior, Luis Geraldo Vaz, Jos Vander Sloten, Joke Duyck, Siegfried V N Jaecques.   

Abstract

PURPOSE: The purpose of the present study was to evaluate the biomechanical environment of immediately placed implants, before and after osseointegration, by comparing three different implant-abutment connection types.
MATERIALS AND METHODS: A computer tomography-based finite element model of an upper central incisor extraction socket was constructed containing implants with either external hex, internal hex, or Morse-taper connection. Frictional contact elements were used in the bone, implant, abutment, and abutment screw interfaces in the immediately placed simulations. In osseointegrated simulations, the repair of bone alveolar defect and a glued bone-to-implant interface were assumed. By analysis of variance, the influence was assessed of connection type, clinical situation, and loading magnitude on the peak equivalent strain in the bone, peak von Mises stress in the abutment screw, bone-to-implant relative displacement, and abutment gap.
RESULTS: The loading magnitudes had a significant contribution, regardless of the assessed variable. However, the critical clinical situation of an immediately placed implant itself was the main factor affecting the peak equivalent strain in the bone and bone-to-implant displacement. The largest influence of the connection type in this protocol was seen on the peak equivalent stress in the abutment screw. On the other hand, a higher influence of the various connection types on bone stress/strain could be noted in osseointegrated simulations.
CONCLUSIONS: The implant-abutment connection design did not significantly influence the biomechanical environment of immediately placed implants. Avoiding implant overloading and ensuring a sufficient initial intraosseous stability are the most relevant parameters for the promotion of a safe biomechanical environment in this protocol.

Entities:  

Mesh:

Year:  2009        PMID: 19438946     DOI: 10.1111/j.1708-8208.2009.00155.x

Source DB:  PubMed          Journal:  Clin Implant Dent Relat Res        ISSN: 1523-0899            Impact factor:   3.932


  15 in total

1.  Biomechanical effects of the implant material and implant-abutment interface in immediately loaded small-diameter implants.

Authors:  Aaron Yu-Jen Wu; Heng-Li Huang; Jui-Ting Hsu; Winston Chee
Journal:  Clin Oral Investig       Date:  2013-09-03       Impact factor: 3.573

2.  A retrospective study of implant-abutment connections on crestal bone level.

Authors:  M-I Lin; Y-W Shen; H-L Huang; J-T Hsu; L-J Fuh
Journal:  J Dent Res       Date:  2013-10-24       Impact factor: 6.116

Review 3.  Finite element analysis: A boon to dentistry.

Authors:  Shilpa Trivedi
Journal:  J Oral Biol Craniofac Res       Date:  2014-12-04

4.  A Finite Element Stress Analysis of a Concical Triangular Connection in Implants: A New Proposal.

Authors:  Romy Angeles Maslucan; John Alexis Dominguez
Journal:  Materials (Basel)       Date:  2022-05-20       Impact factor: 3.748

Review 5.  Is the internal connection more efficient than external connection in mechanical, biological, and esthetical point of views? A systematic review.

Authors:  Marcelo Coelho Goiato; Eduardo Piza Pellizzer; Emily Vivianne Freitas da Silva; Liliane da Rocha Bonatto; Daniela Micheline dos Santos
Journal:  Oral Maxillofac Surg       Date:  2015-04-25

6.  Nonlinear finite element analysis of three implant- abutment interface designs.

Authors:  Chun-Bo Tang; Si-Yu Liul; Guo-Xing Zhou; Jin-Hua Yu; Guang-Dong Zhang; Yi-Dong Bao; Qiu-Ju Wang
Journal:  Int J Oral Sci       Date:  2012-06       Impact factor: 6.344

7.  Correlation between microleakage and screw loosening at implant-abutment connection.

Authors:  Cem Sahin; Simel Ayyildiz
Journal:  J Adv Prosthodont       Date:  2014-02-14       Impact factor: 1.904

8.  Effects of Implant-Abutment Connection Type and Inter-Implant Distance on Inter-Implant Bone Stress and Microgap: Three-Dimensional Finite Element Analysis.

Authors:  Takashi Matsuoka; Tamaki Nakano; Satoshi Yamaguchi; Shinji Ono; Shota Watanabe; Takumi Sato; Hirofumi Yatani
Journal:  Materials (Basel)       Date:  2021-05-06       Impact factor: 3.623

9.  A three-dimensional finite element study on the stress distribution pattern of two prosthetic abutments for external hexagon implants.

Authors:  Wagner Moreira; Caio Hermann; Jucélio Tomás Pereira; Jean Anacleto Balbinoti; Rodrigo Tiossi
Journal:  Eur J Dent       Date:  2013-10

10.  Long-term cumulative survival and mechanical complications of single-tooth Ankylos Implants: focus on the abutment neck fractures.

Authors:  Hye Won Shim; Byoung-Eun Yang
Journal:  J Adv Prosthodont       Date:  2015-12-30       Impact factor: 1.904

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.