Correlation between microleakage and screw loosening at implant-abutment connection.

Cem Sahin1, Simel Ayyildiz2.   

Abstract

PURPOSE: This study aimed to evaluate the correlation between microleakage and screw loosening at different types of implant-abutment connections and/or geometries measuring the torque values before and after the leakage tests.
MATERIALS AND METHODS: Three different abutment types (Intenal hex titanium, internal hex zirconium, morse tapered titaniuım) with different geometries were connected to its own implant fixture. All the abutments were tightened with a standard torque value then the composition was connected to the modified fluid filtration system. After the measurements of leakage removal torque values were re-measured. Kruskal-wallis test was performed for non-parametric and one-way ANOVA was performed for parametric data. The correlation was evaluated using Spearman Correlation Test (α=0.05).
RESULTS: Significantly higher microleakage was found at the connection of implant-internal hex zirconium abutment. Observed mean torque value loss was also significantly higher than other connection geometries. Spearman tests revealed a significant correlation between microleakage and screw loosening.
CONCLUSION: Microleakage may provoke screw loosening. Removing torque values rationally decrease with the increase of microleakage.

Entities:  

Keywords:  Implant-abutment connection; Microleakage; Screw loosening

Year:  2014        PMID: 24605204      PMCID: PMC3942525          DOI: 10.4047/jap.2014.6.1.35

Source DB:  PubMed          Journal:  J Adv Prosthodont        ISSN: 2005-7806            Impact factor:   1.904


INTRODUCTION

Developments in dental implantology evidently changed the treatment modalities over the last 25 years. However micro-gap formation between the surfaces of implant fixture and abutment is still one of the major problems at the connection area which may lead to mechanical and biological failures such as screw loosening and periimplantitis. The screw, used to bind the implant and the abutment assisted by external, internal or morse tapered geometries, is tightened with a certain torque represented in Ncm according to manufacturers' instructions.1 This force is transferred along the interface of the abutment, screw thread surfaces and implant thread surfaces. During the first screwing process, the torque energy is expended in smoothing the implant and abutment mating surfaces. Though this induces material fatigue according to superficial composition and/or metallurgical properties, this also provides a better harmony of the opposing surfaces which may avoid microleakage at the connection area.2 The mismatch of the implant and the abutment surfaces can cause rapid stress which ends up with loosening of screw3 and also microleakage. Especially in the single crowns screw loosening is still the most frequently seen complication.4,5 It is claimed that greater forces are required to loosen the screws with greater preloads.6 However, even after performing essential tightening torque values, some of the researchers claimed that micro spaces still exist between these surfaces and they provoke undesirable movements, resulting in screw related mechanical failures.7-9 These gaps also provide suitable areas for microorganism survival. Their toxins and metabolites spread into peri implant tissues. Consequently because of such microbiologic activities, a slippery environmeent altering the loosing of the screw may take shape. Eventually periimplantitis with the peak of inflammatory cell content especially around the implant-abutment interface related with the misfit of these surfaces cannot be impeded.10 Several studies have been carried out examining screw loosening, microgap formation or mechanical and biological failures of dental implants.2,10-13 But, to date little or no data exist about the relationship between the screw loosening at connections of different types of implant-abutment and microleakage. This study aimed to evaluate the correlation between microleakage and screw loosening at connections of different types of implant-abutment and/or geometries measuring the torque values before and after the leakage tests.

MATERIALS AND METHODS

Table 1 shows the implant and abutment types used in this study. Three different abutment types with different geometries were connected to its own implant fixture according to manufacturer's recommendation in a laminar flow to avoid any contamination. All the abutments were tightened with a standard torque value of 25 Nm using a digital wrench (TQ-680; Instrutherm Mea. Ins., Sao Paulo, SP, Brasil). The implant fixtures were hold with a special device to mimic its stabilization in the bone. Then a transparent hydraulic plastic tube with 6 × 8 scales was binded to the specimen as shown in the Fig. 1 and this composition was connected to the modified fluid filtration system to perform microleakage tests.
Table 1

Shows the implant and abutment types used in this study

Fig. 1

Schematic view of microleakage throuh the screw threads and implant-abutment interface.

To measure the leakage at implant-abutment interface of implant-internal hex zirconium (I-IhZ), implant-internal hex titanium (I-IhT) and implant-morse tapered titanium (I-MTtT), a modified fluid filtration method was used.14 The parts of the filtration model (tubes, micropipettes, buffer, and etc.) were all filled with deionized water. The specimens were connected to the system using a hydraulic plastic tube. Regulated air from the pressure tank at 121.6 Kpa (1240 cm H2O) was applied. The water in the buffer was forced to move through the interface from the threads of implant fixture and screws shown in the Fig. 1, providing to test the tightness and/or geometric harmony of implantabutment connection. To pass in a tiny air bubble to the system a special designed microsyringe was used. Then the air bubble and the system were stabilized before measurements. Linear travelling of the air bubble through the 100 µL micropipette was then followed at a period of 20 minutes to determine the leakage quantitatively. The values were expressed as mL/min/cm H2O. After the measurements of leakage at the connections at the certain period of time, each sample was immediately disconnected from the system and plastic tube and dried with blotting paper. To obtain removal torque values, samples were re-inserted to the special holder. Data were used to calculate torque values using RTV/25 × 100 to express as percentages. Kruskal-wallis test was performed for non-parametric microleakage data, and one-way ANOVA was performed for parametric torque value data (PASW Statistics for Windows, version 18.0, SPSS, Chicago, Il, USA). Then Tukey test was used to obtain the statistical differences among the test groups. The correlation between microleakage and torque values were evaluated using Spearman Correlation Test. P value less than .05 was considered as significant for all the test methods.

RESULTS

The highest microleakage was found at the connection (I-IhZ), (Table 2). This result was significantly different from I-IhT and I-MtT groups. Microleakage at the connection of I-IhT was slightly higher than I-MtT connection, but this result was not statistically significant. Mean torque value loss, observed at the connection of I-IhZ was significantly higher than other connection geometries (Table 2). There were no statistical differences between losses of mean torque values at the connections of I-MtT and I-IhT.
Table 2

Microleakage and loss of torque values

Tightening torque = 25 Nm (Same letters with no statistical differences)

According to the results of Spearman Correlation Test, the coefficient was found to be 0.65 which means that as the microleakage increase, the potential of screw loosening increase. This result was statistically significant at the given P value.

DISCUSSION

According to the results of our study it can be speculated that microleakage, through the threads of the screw and the implant and also the surfaces of implant-abutment connection, may provoke screw loosening. The tightening torque before the microleakage tests was specified as 25 Nm. However after performing the tests for definitive period mean removal torque value was degreased at least 9.4%. According to the results of our study these two components were highly correlated (0.65) which means that as microleakage between the surfaces increase this may induce loosening of the screw. Microleakage was found to be minimum between implant and morse tapered titanium abutment surfaces. This result was parallel to the study of Pessoa et al.15 They found that morse tapered geometries of connections present a better harmony and stabilization which may avoid extreme deformation of mating surfaces and microleakage. Yet none of the 2 component implant systems today can prevent fluid diffusion.11,16 Besides, during the tightening process, opposing contacts may flatten, providing smoothened surfaces which may reduce microleakage. However this phenomenon may turn into a disadvantage when repeated tightening and removing cycles are performed by breaking down the frictional stabilization.17 Within the limitations in our study, we measured the removal torque value once after microleakage test. Cardoso et al.12 concluded that screw loosening may pass over 20% with repeated tightening and removing cycles. Consequently detorque data may change with repetations. The other major determinants are lubricants such as saliva, blood or microbial structures like extracellular matrix and/or slime layer that may aggravate microorganism proliferation at the connection area. These factors may alter detorque values in-vivo creating slippery environment. Despite there are several ways of determining microleakage at the connection area16,18-20 we used a modified filtration method in-vitro to mimic a lubricated surface using deionized water with pressure.14 Detorque value is expected to be the same as tightening value at perfect connections. However, it may be difficult to achieve according to many researchers.13,17,21 This is compatible with our study that removing torque values decreased more than 9% of tightening value. Under the same test conditions the decreased percentage was higher at implant-ceramic abutment connection possibly due to the worse mating surfaces compared with titanium. Biocompatibility and aesthetic are the primary reasons for choosing ceramic materials in dental practice. Nevertheless as microleakage at I-IhZ connection is about 5 times greater than the titanium connection in certain in-vitro study, a different point of view can arise. Grater values may be due to the milled surface of the zirconium abutments rougher than the titanium which avoids mating of the surfaces as discussed. It is known that such rough surfaces offer suitable areas for microorganism adhesion and/or colonization.22,23 Afterward oral microorganisms may easily invade through the periimplant areas with the aid of their own extracellular products, forming slippery environment as discussed. This process can cause failure of the implant along with soft and/or hard tissue loss. Consequently this may be commented as a deficit of ceramic material.

CONCLUSION

It may be speculated that microleakage provokes screw loosening therefore removing torque values rationally decrease with the increase of microleakage. The rougher surface of ceramic abutments compared with titanium abutment may turn the aesthetic advantageinto a deficit due to lubricants and ease of adhesion of microganisms. Removing torque values only reached up to 91% of the tightening torque values.
  23 in total

1.  Peri-implant inflammation defined by the implant-abutment interface.

Authors:  N Broggini; L M McManus; J S Hermann; R Medina; R K Schenk; D Buser; D L Cochran
Journal:  J Dent Res       Date:  2006-05       Impact factor: 6.116

2.  Effect of lateral cyclic loading on abutment screw loosening of an external hexagon implant system.

Authors:  Ameen Khraisat; Akihiko Hashimoto; Shuichi Nomura; Osamu Miyakawa
Journal:  J Prosthet Dent       Date:  2004-04       Impact factor: 3.426

3.  In vitro permeability of etch-and-rinse and self-etch adhesives used for immediate dentin sealing.

Authors:  Cem Sahin; Zafer C Cehreli; Muhittin Yenigul; Bulent Dayangac
Journal:  Dent Mater J       Date:  2012-05-14       Impact factor: 2.102

4.  Preloads generated with repeated tightening in three types of screws used in dental implant assemblies.

Authors:  Declan Byrne; Stuart Jacobs; Brian O'Connell; Frank Houston; Noel Claffey
Journal:  J Prosthodont       Date:  2006 May-Jun       Impact factor: 2.752

5.  The effect of implant/abutment hexagonal misfit on screw joint stability.

Authors:  P P Binon
Journal:  Int J Prosthodont       Date:  1996 Mar-Apr       Impact factor: 1.681

6.  Retention of oral microorganisms on cobalt-chromium alloy and dental acrylic resin with different surface finishes.

Authors:  R Taylor; C Maryan; J Verran
Journal:  J Prosthet Dent       Date:  1998-11       Impact factor: 3.426

7.  Torque removal evaluation of prosthetic screws after tightening and loosening cycles: an in vitro study.

Authors:  Mayra Cardoso; Marcelo Ferreira Torres; Eduardo José Veras Lourenço; Daniel de Moraes Telles; Renata Cristina Silveira Rodrigues; Ricardo Faria Ribeiro
Journal:  Clin Oral Implants Res       Date:  2011-04-04       Impact factor: 5.977

8.  Nonlinear contact analysis of preload in dental implant screws.

Authors:  R L Sakaguchi; S E Borgersen
Journal:  Int J Oral Maxillofac Implants       Date:  1995 May-Jun       Impact factor: 2.804

Review 9.  Comparison of survival and complication rates of tooth-supported fixed dental prostheses (FDPs) and implant-supported FDPs and single crowns (SCs).

Authors:  Bjarni E Pjetursson; Urs Brägger; Niklaus P Lang; Marcel Zwahlen
Journal:  Clin Oral Implants Res       Date:  2007-06       Impact factor: 5.977

10.  In vitro evaluation of the implant abutment connection sealing capability of different implant systems.

Authors:  P G Coelho; P Sudack; M Suzuki; K S Kurtz; G E Romanos; N R F A Silva
Journal:  J Oral Rehabil       Date:  2008-10-29       Impact factor: 3.837

View more
  9 in total

1.  Removal torque evaluation of three different abutment screws for single implant restorations after mechanical cyclic loading.

Authors:  T Paepoemsin; P A Reichart; P Chaijareenont; F P Strietzel; P Khongkhunthian
Journal:  Oral Implantol (Rome)       Date:  2016-11-16

Review 2.  Microleakage at the Different Implant Abutment Interface: A Systematic Review.

Authors:  Sunil Kumar Mishra; Ramesh Chowdhary; Shail Kumari
Journal:  J Clin Diagn Res       Date:  2017-06-01

3.  Evaluation the loosening of abutment screws in fluid contamination: an in vitro study.

Authors:  Fei Sun; Wei Cheng; Bao-Hong Zhao; Gui-Qiu Song; Zeng Lin
Journal:  Sci Rep       Date:  2022-06-24       Impact factor: 4.996

4.  A Comparative 3D Finite Element Computational Study of Three Connections.

Authors:  Davide Farronato; Mattia Manfredini; Andrea Stevanello; Veronica Campana; Lorenzo Azzi; Marco Farronato
Journal:  Materials (Basel)       Date:  2019-09-26       Impact factor: 3.623

5.  Analysis of Torque Maintenance and Fracture Resistance after Fatigue in Retention Screws Made of Different Metals for Screw-Retained Implant-Borne Prosthesis Joints.

Authors:  Maria Beatriz Bello Taborda; Gabriela Sumie Yaguinuma Gonçalves; Cecília Alves de Sousa; Wirley Gonçalves Assunção
Journal:  Int J Dent       Date:  2021-11-18

6.  Influence of Anodizing Stages on the Preload Force of Implant-Abutment Screws and Their Benefits Regarding the Concept of Immediate Implant Placement-An In Vitro Study.

Authors:  Florian Rathe; Paul Weigl; Jan Wasiak; Christoph Ratka; Holger Zipprich
Journal:  Materials (Basel)       Date:  2022-01-20       Impact factor: 3.623

7.  Evaluation of sealing efficacy and removal convenience of sealing materials for implant abutment screw access holes.

Authors:  Huangjun Zhou; Sixian Ye; Xingyu Lyu; Hao Feng; Min Liu; Cai Wen
Journal:  BMC Oral Health       Date:  2022-08-25       Impact factor: 3.747

8.  Comparative study of abutment screw loosening with or without adhesive material.

Authors:  Mahnaz Arshad; Gholamreza Shirani; Sina Refoua; Mohammadreza Rahimi Yeganeh
Journal:  J Adv Prosthodont       Date:  2017-04-06       Impact factor: 1.904

9.  Comparison of the Effect of Four Different Abutment Screw Torques on Screw Loosening in Single Implant-Supported Prosthesis after the Application of Mechanical Loading.

Authors:  Behnaz Ebadian; Amirhossein Fathi; Saba Khodadad
Journal:  Int J Dent       Date:  2021-07-19
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.