Literature DB >> 19365851

Short echo spectroscopic imaging of the human brain at 7T using transceiver arrays.

N I Avdievich1, J W Pan, J M Baehring, D D Spencer, H P Hetherington.   

Abstract

Recent advances in magnet technology have enabled the construction of ultrahigh-field magnets (7T and higher) that can accommodate the human head and body. Despite the intrinsic advantages of performing spectroscopic imaging at 7T, increased signal-to-noise ratio (SNR), and spectral resolution, few studies have been reported to date. This limitation is largely due to increased power deposition and B(1) inhomogeneity. To overcome these limitations, we used an 8-channel transceiver array with a short TE (15 ms) spectroscopic imaging sequence. Utilizing phase and amplitude mapping and optimization schemes, the 8-element transceiver array provided both improved efficiency (17% less power for equivalent peak B(1)) and homogeneity (SD(B(1)) = +/-10% versus +/-22%) in comparison to a transverse electromagnetic (TEM) volume coil. To minimize the echo time to measure J-modulating compounds such as glutamate, we developed a short TE sequence utilizing a single-slice selective excitation pulse followed by a broadband semiselective refocusing pulse. Extracerebral lipid resonances were suppressed with an inversion recovery pulse and delay. The short TE sequence enabled visualization of a variety of resonances, including glutamate, in both a control subject and a patient with a Grade II oligodendroglioma. (c) 2009 Wiley-Liss, Inc.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19365851      PMCID: PMC3675773          DOI: 10.1002/mrm.21970

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  23 in total

1.  In vivo 1H NMR spectroscopy of the human brain at 7 T.

Authors:  I Tkác; P Andersen; G Adriany; H Merkle; K Ugurbil; R Gruetter
Journal:  Magn Reson Med       Date:  2001-09       Impact factor: 4.668

2.  Transmit and receive transmission line arrays for 7 Tesla parallel imaging.

Authors:  Gregor Adriany; Pierre-Francois Van de Moortele; Florian Wiesinger; Steen Moeller; John P Strupp; Peter Andersen; Carl Snyder; Xiaoliang Zhang; Wei Chen; Klaas P Pruessmann; Peter Boesiger; Tommy Vaughan; Kāmil Uğurbil
Journal:  Magn Reson Med       Date:  2005-02       Impact factor: 4.668

3.  Exploring the limits of RF shimming for high-field MRI of the human head.

Authors:  Weihua Mao; Michael B Smith; Christopher M Collins
Journal:  Magn Reson Med       Date:  2006-10       Impact factor: 4.668

4.  Spectroscopic imaging with volume selection by unpaired adiabatic pi pulses: theory and application.

Authors:  Julien Valette; Jang-Yeon Park; Olli Gröhn; Kāmil Uğurbil; Michael Garwood; Pierre-Gilles Henry
Journal:  J Magn Reson       Date:  2007-08-24       Impact factor: 2.229

5.  Mutual Inductance in the Bird-Cage Resonator

Authors: 
Journal:  J Magn Reson       Date:  1997-05       Impact factor: 2.229

6.  4 T Actively detuneable double-tuned 1H/31P head volume coil and four-channel 31P phased array for human brain spectroscopy.

Authors:  N I Avdievich; H P Hetherington
Journal:  J Magn Reson       Date:  2007-03-03       Impact factor: 2.229

7.  Radiative losses of a birdcage resonator.

Authors:  M D Harpen
Journal:  Magn Reson Med       Date:  1993-05       Impact factor: 4.668

8.  Human magnetic resonance imaging at 8 T.

Authors:  P M Robitaille; A M Abduljalil; A Kangarlu; X Zhang; Y Yu; R Burgess; S Bair; P Noa; L Yang; H Zhu; B Palmer; Z Jiang; D M Chakeres; D Spigos
Journal:  NMR Biomed       Date:  1998-10       Impact factor: 4.044

9.  Interleaved narrow-band PRESS sequence with adiabatic spatial-spectral refocusing pulses for 1H MRSI at 7T.

Authors:  Priti Balchandani; John Pauly; Daniel Spielman
Journal:  Magn Reson Med       Date:  2008-05       Impact factor: 4.668

10.  Fat suppression for 1H MRSI at 7T using spectrally selective adiabatic inversion recovery.

Authors:  Priti Balchandani; Daniel Spielman
Journal:  Magn Reson Med       Date:  2008-05       Impact factor: 4.668

View more
  34 in total

1.  Multislice ¹H MRSI of the human brain at 7 T using dynamic B₀ and B₁ shimming.

Authors:  Vincent O Boer; Dennis W J Klomp; Christoph Juchem; Peter R Luijten; Robin A de Graaf
Journal:  Magn Reson Med       Date:  2011-12-12       Impact factor: 4.668

2.  J-refocused coherence transfer spectroscopic imaging at 7 T in human brain.

Authors:  J W Pan; N Avdievich; H P Hetherington
Journal:  Magn Reson Med       Date:  2010-11       Impact factor: 4.668

Review 3.  Magnetic resonance imaging at ultrahigh fields.

Authors:  Kamil Ugurbil
Journal:  IEEE Trans Biomed Eng       Date:  2014-03-25       Impact factor: 4.538

4.  Transmit B1-field correction at 7 T using actively tuned coupled inner elements.

Authors:  Hellmut Merkle; Joseph Murphy-Boesch; Peter van Gelderen; Shumin Wang; Tie-Qiang Li; Alan P Koretsky; Josef H Duyn
Journal:  Magn Reson Med       Date:  2011-03-24       Impact factor: 4.668

5.  Improved homogeneity of the transmit field by simultaneous transmission with phased array and volume coil.

Authors:  Nikolai I Avdievich; Sukhoon Oh; Hoby P Hetherington; Christopher M Collins
Journal:  J Magn Reson Imaging       Date:  2010-08       Impact factor: 4.813

6.  Role of very high order and degree B0 shimming for spectroscopic imaging of the human brain at 7 tesla.

Authors:  Jullie W Pan; Kai-Ming Lo; Hoby P Hetherington
Journal:  Magn Reson Med       Date:  2011-12-28       Impact factor: 4.668

Review 7.  Utility of magnetic resonance spectroscopic imaging for human epilepsy.

Authors:  Jullie W Pan; Ruben I Kuzniecky
Journal:  Quant Imaging Med Surg       Date:  2015-04

Review 8.  Imaging at ultrahigh magnetic fields: History, challenges, and solutions.

Authors:  Kamil Uğurbil
Journal:  Neuroimage       Date:  2017-07-08       Impact factor: 6.556

Review 9.  Strategies for rapid in vivo 1H and hyperpolarized 13C MR spectroscopic imaging.

Authors:  Sarah J Nelson; Eugene Ozhinsky; Yan Li; Il woo Park; Jason Crane
Journal:  J Magn Reson       Date:  2013-02-08       Impact factor: 2.229

10.  Resonant inductive decoupling (RID) for transceiver arrays to compensate for both reactive and resistive components of the mutual impedance.

Authors:  Nikolai I Avdievich; Jullie W Pan; Hoby P Hetherington
Journal:  NMR Biomed       Date:  2013-06-18       Impact factor: 4.044

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.