Literature DB >> 19362249

Clinical accuracy of the respiratory tumor tracking system of the cyberknife: assessment by analysis of log files.

Mischa Hoogeman1, Jean-Briac Prévost, Joost Nuyttens, Johan Pöll, Peter Levendag, Ben Heijmen.   

Abstract

PURPOSE: To quantify the clinical accuracy of the respiratory motion tracking system of the CyberKnife treatment device. METHODS AND MATERIALS: Data in log files of 44 lung cancer patients treated with tumor tracking were analyzed. Errors in the correlation model, which relates the internal target motion with the external breathing motion, were quantified. The correlation model error was compared with the geometric error obtained when no respiratory tracking was used. Errors in the prediction method were calculated by subtracting the predicted position from the actual measured position after 192.5 ms (the time lag to prediction in our current system). The prediction error was also measured for a time lag of 115 ms and a new prediction method.
RESULTS: The mean correlation model errors were less than 0.3 mm. Standard deviations describing intrafraction variations around the whole-fraction mean error were 0.2 to 1.9 mm for cranio-caudal, 0.1 to 1.9 mm for left-right, and 0.2 to 2.5 mm for anterior-posterior directions. Without the use of respiratory tracking, these variations would have been 0.2 to 8.1 mm, 0.2 to 5.5 mm, and 0.2 to 4.4 mm. The overall mean prediction error was small (0.0 +/- 0.0 mm) for all directions. The intrafraction standard deviation ranged from 0.0 to 2.9 mm for a time delay of 192.5 ms but was halved by using the new prediction method.
CONCLUSIONS: Analyses of the log files of real clinical cases have shown that the geometric error caused by respiratory motion is substantially reduced by the application of respiratory motion tracking.

Entities:  

Mesh:

Year:  2009        PMID: 19362249     DOI: 10.1016/j.ijrobp.2008.12.041

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  80 in total

1.  Online monitoring and error detection of real-time tumor displacement prediction accuracy using control limits on respiratory surrogate statistics.

Authors:  Kathleen Malinowski; Thomas J McAvoy; Rohini George; Sonja Dieterich; Warren D D'Souza
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

2.  Correlation and prediction uncertainties in the cyberknife synchrony respiratory tracking system.

Authors:  Eric W Pepin; Huanmei Wu; Yuenian Zhang; Bryce Lord
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

3.  Factors affecting the accuracy of respiratory tracking of the image-guided robotic radiosurgery system.

Authors:  Mitsuhiro Inoue; Kohei Okawa; Junichi Taguchi; Yoshifumi Hirota; Yohei Yanagiya; Chie Kikuchi; Michio Iwabuchi; Taro Murai; Hiromitsu Iwata; Hiroya Shiomi; Izumi Koike; Koshi Tatewaki; Seiji Ohta
Journal:  Jpn J Radiol       Date:  2019-07-31       Impact factor: 2.374

4.  Real-time tumor tracking using sequential kV imaging combined with respiratory monitoring: a general framework applicable to commonly used IGRT systems.

Authors:  Byungchul Cho; Per Rugaard Poulsen; Paul J Keall
Journal:  Phys Med Biol       Date:  2010-05-18       Impact factor: 3.609

5.  A Study on Stereoscopic X-ray Imaging Data Set on the Accuracy of Real-Time Tumor Tracking in External Beam Radiotherapy.

Authors:  Ahmad Esmaili Torshabi; Leila Ghorbanzadeh
Journal:  Technol Cancer Res Treat       Date:  2016-07-08

6.  A single reference measurement can predict liver tumor motion during respiration.

Authors:  Jakub Cvek; Lukas Knybel; Lukas Molenda; Bretislav Otahal; Tomas Jonszta; Daniel Czerny; David Feltl
Journal:  Rep Pract Oncol Radiother       Date:  2015-12-29

7.  Real-time prediction of tumor motion using a dynamic neural network.

Authors:  Majid Mafi; Saeed Montazeri Moghadam
Journal:  Med Biol Eng Comput       Date:  2020-01-08       Impact factor: 2.602

8.  Experimental investigation of a general real-time 3D target localization method using sequential kV imaging combined with respiratory monitoring.

Authors:  Byungchul Cho; Per Poulsen; Dan Ruan; Amit Sawant; Paul J Keall
Journal:  Phys Med Biol       Date:  2012-10-24       Impact factor: 3.609

9.  CyberKnife radiosurgery for inoperable stage IA non-small cell lung cancer: 18F-fluorodeoxyglucose positron emission tomography/computed tomography serial tumor response assessment.

Authors:  Saloomeh Vahdat; Eric K Oermann; Sean P Collins; Xia Yu; Malak Abedalthagafi; Pedro Debrito; Simeng Suy; Shadi Yousefi; Constanza J Gutierrez; Thomas Chang; Filip Banovac; Eric D Anderson; Giuseppe Esposito; Brian T Collins
Journal:  J Hematol Oncol       Date:  2010-02-04       Impact factor: 17.388

10.  Couch and multileaf collimator tracking: A clinical feasibility study for pancreas and liver treatment.

Authors:  Lei Zhang; Thomas LoSasso; Pengpeng Zhang; Margie Hunt; Gig Mageras; Grace Tang
Journal:  Med Phys       Date:  2020-09-11       Impact factor: 4.071

View more