BACKGROUND: This study investigated the non-invasive induction of spherical aberration (SA) levels consistent with complication-free wavefront-guided (WFG) refractive surgery using custom WFG soft contact lenses and quantified the resulting impact on visual performance.
METHODS: Twelve healthy individuals of typical military age (mean of 26.08 +/- 1.92 years) meeting the inclusion criteria of emmetropia (less than +/- 0.50 DS and less than +/- 0.50 DC) as measured by subjective refraction served as subjects. Five plano lenses were manufactured to induce a range of SA typical of those encountered after refractive surgery. The measured SA values over a 6 mm artificial pupil in these five lenses were -0.224 microm, 0.074 microm, 0.214 microm, 0.495 microm and 0.621 microm. For each subject, the level of total ocular aberrations through 10 Zernike radial orders over a 6 mm pupil was measured with a custom Shack-Hartmann wavefront sensor, while wearing each lens. Visual performance was evaluated using high contrast visual acuity (HCVA) measured through a 6 mm artificial pupil.
RESULTS: Custom WFG soft contact lenses can be manufactured to include the range of SA that occurs with complication-free refractive surgery (-0.20 microm to +0.60 microm) and produces a change in SA within a similar range when worn on the eye. High contrast visual performance with these lenses varied, depending on the amount of SA in each lens. Visual performance worsened with greater amounts of positive or negative SA and a quadratic fit to the data peaked at +0.209 microm.
CONCLUSIONS: Defined levels of SA can be manufactured and induced (non-invasively) with WFG soft contact lenses and their effect on logMAR high contrast visual performance can be measured. Results show that subjects' best logMAR HCVA occurs with the presence of positive residual SA. When designing the actual test lens, the targeted change in aberrations for each subject will likely be better achieved by first measuring the aberrations of a template lens on the eye and then compensating for unique subject dependent eye-lens interactions.

Entities:  

Mesh:

Year:  2009        PMID: 19351378      PMCID: PMC2773130          DOI: 10.1111/j.1444-0938.2009.00375.x

Source DB:  PubMed          Journal:  Clin Exp Optom        ISSN: 0816-4622            Impact factor:   2.742


  13 in total

1.  Ocular aberrations before and after myopic corneal refractive surgery: LASIK-induced changes measured with laser ray tracing.

Authors:  E Moreno-Barriuso; J M Lloves; S Marcos; R Navarro; L Llorente; S Barbero
Journal:  Invest Ophthalmol Vis Sci       Date:  2001-05       Impact factor: 4.799

2.  Visual acuity as a function of Zernike mode and level of root mean square error.

Authors:  Raymond A Applegate; Charles Ballentine; Hillery Gross; Edwin J Sarver; Charlene A Sarver
Journal:  Optom Vis Sci       Date:  2003-02       Impact factor: 1.973

3.  Three-dimensional relationship between high-order root-mean-square wavefront error, pupil diameter, and aging.

Authors:  Raymond A Applegate; William J Donnelly; Jason D Marsack; Darren E Koenig; Konrad Pesudovs
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2007-03       Impact factor: 2.129

4.  The Psychophysics Toolbox.

Authors:  D H Brainard
Journal:  Spat Vis       Date:  1997

5.  Measuring visual acuity--mesopic or photopic conditions, and high or low contrast letters?

Authors:  Konrad Pesudovs; Jason D Marsack; William J Donnelly; Larry N Thibos; Raymond A Applegate
Journal:  J Refract Surg       Date:  2004 Sep-Oct       Impact factor: 3.573

6.  Image metrics for predicting subjective image quality.

Authors:  Li Chen; Ben Singer; Antonio Guirao; Jason Porter; David R Williams
Journal:  Optom Vis Sci       Date:  2005-05       Impact factor: 1.973

7.  Personalized correction of spherical aberration in cataract surgery.

Authors:  George H H Beiko
Journal:  J Cataract Refract Surg       Date:  2007-08       Impact factor: 3.351

8.  Validation of an off-eye contact lens Shack-Hartmann wavefront aberrometer.

Authors:  Pete Kollbaum; Meredith Jansen; Larry Thibos; Arthur Bradley
Journal:  Optom Vis Sci       Date:  2008-09       Impact factor: 1.973

9.  Customized LASIK treatment for myopia: relationship between preoperative higher order aberrations and refractive outcome.

Authors:  Manoj V Subbaram; ScottM MacRae; Stephen G Slade; Daniel S Durrie
Journal:  J Refract Surg       Date:  2006-10       Impact factor: 3.573

10.  Are all aberrations equal?

Authors:  Raymond A Applegate; Edwin J Sarver; Vic Khemsara
Journal:  J Refract Surg       Date:  2002 Sep-Oct       Impact factor: 3.573

View more
  2 in total

1.  Visual impact of Zernike and Seidel forms of monochromatic aberrations.

Authors:  Xu Cheng; Arthur Bradley; Sowmya Ravikumar; Larry N Thibos
Journal:  Optom Vis Sci       Date:  2010-05       Impact factor: 1.973

2.  Comparing the optical properties of soft contact lenses on and off the eye.

Authors:  Pete S Kollbaum; Arthur Bradley; Larry N Thibos
Journal:  Optom Vis Sci       Date:  2013-09       Impact factor: 1.973

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.