Evaluation of a patient-specific Monte Carlo software for CT dosimetry.

Marios Myronakis1, Kostas Perisinakis, Antonis Tzedakis, Sofia Gourtsoyianni, John Damilakis.   

Abstract

The aim was to validate the ImpactMC computed tomography (CT) dosimetry software that allows patient-specific dose determination. Measured values of head- and body-weighted CT dose index (CTDI(w)) were compared with corresponding values derived using ImpactMC software. A physical anthropomorphic phantom simulating the average adult was employed to study the effect of exposure parameters used to produce the input image set on a normalised dose output and the relationship between exposure parameters selected for simulation on the dose output. The difference between CTDI(w) values obtained through measurements and simulations were found to be up to 12.8 and 18.3% for head and body phantoms, respectively. Exposure parameters of the image set used as input were found to have a minor impact on the normalised dose output. Simulations confirmed the expected linear relationship between dose and tube load and the power law relationship between dose and tube potential. Results demonstrate that ImpactMC may be capable of providing reliable CT dose estimates.

Entities:  

Mesh:

Year:  2009        PMID: 19329513     DOI: 10.1093/rpd/ncp051

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  6 in total

1.  Radiation dose reduction in computed tomography: techniques and future perspective.

Authors:  Lifeng Yu; Xin Liu; Shuai Leng; James M Kofler; Juan C Ramirez-Giraldo; Mingliang Qu; Jodie Christner; Joel G Fletcher; Cynthia H McCollough
Journal:  Imaging Med       Date:  2009-10

2.  The effect of iodine uptake on radiation dose absorbed by patient tissues in contrast enhanced CT imaging: Implications for CT dosimetry.

Authors:  Kostas Perisinakis; Antonis Tzedakis; Kostas Spanakis; Antonios E Papadakis; Adam Hatzidakis; John Damilakis
Journal:  Eur Radiol       Date:  2017-07-14       Impact factor: 5.315

Review 3.  Quality assurance of imaging techniques used in the clinical management of osteoporosis.

Authors:  G Guglielmi; J Damilakis; G Solomou; A Bazzocchi
Journal:  Radiol Med       Date:  2012-10-22       Impact factor: 3.469

Review 4.  Radiation risks associated with serial imaging in colorectal cancer patients: should we worry?

Authors:  Jeong Suk Oh; Jonathan B Koea
Journal:  World J Gastroenterol       Date:  2014-01-07       Impact factor: 5.742

5.  Assessment of CT dose to the fetus and pregnant female patient using patient-specific computational models.

Authors:  Tianwu Xie; Pierre-Alexandre Poletti; Alexandra Platon; Christoph D Becker; Habib Zaidi
Journal:  Eur Radiol       Date:  2017-09-08       Impact factor: 5.315

6.  The presence of contrast agent increases organ radiation dose in contrast-enhanced CT.

Authors:  Mahta Mazloumi; Gert Van Gompel; Veerle Kersemans; Johan de Mey; Nico Buls
Journal:  Eur Radiol       Date:  2021-03-30       Impact factor: 5.315

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.