Towards a teleoperated needle driver robot with haptic feedback for RFA of breast tumors under continuous MRI.

Rebecca Kokes1, Kevin Lister, Rao Gullapalli, Bao Zhang, Alan MacMillan, Howard Richard, Jaydev P Desai.   

Abstract

OBJECTIVE: The purpose of this paper is to explore the feasibility of developing a MRI-compatible needle driver system for radiofrequency ablation (RFA) of breast tumors under continuous MRI imaging while being teleoperated by a haptic feedback device from outside the scanning room. The developed needle driver prototype was designed and tested for both tumor targeting capability as well as RFA.
METHODS: The single degree-of-freedom (DOF) prototype was interfaced with a PHANToM haptic device controlled from outside the scanning room. Experiments were performed to demonstrate MRI-compatibility and position control accuracy with hydraulic actuation, along with an experiment to determine the PHANToM's ability to guide the RFA tool to a tumor nodule within a phantom breast tissue model while continuously imaging within the MRI and receiving force feedback from the RFA tool.
RESULTS: Hydraulic actuation is shown to be a feasible actuation technique for operation in an MRI environment. The design is MRI-compatible in all aspects except for force sensing in the directions perpendicular to the direction of motion. Experiments confirm that the user is able to detect healthy vs. cancerous tissue in a phantom model when provided with both visual (imaging) feedback and haptic feedback.
CONCLUSION: The teleoperated 1-DOF needle driver system presented in this paper demonstrates the feasibility of implementing a MRI-compatible robot for RFA of breast tumors with haptic feedback capability.

Entities:  

Year:  2009        PMID: 19303805      PMCID: PMC2698937          DOI: 10.1016/j.media.2009.02.001

Source DB:  PubMed          Journal:  Med Image Anal        ISSN: 1361-8415            Impact factor:   8.545


  27 in total

1.  Force feedback plays a significant role in minimally invasive surgery: results and analysis.

Authors:  Gregory Tholey; Jaydev P Desai; Andres E Castellanos
Journal:  Ann Surg       Date:  2005-01       Impact factor: 12.969

Review 2.  Quantitative MRI-based temperature mapping based on the proton resonant frequency shift: review of validation studies.

Authors:  N McDannold
Journal:  Int J Hyperthermia       Date:  2005-09       Impact factor: 3.914

3.  A biplanar fluoroscopic approach for the measurement, modeling, and simulation of needle and soft-tissue interaction.

Authors:  James T Hing; Ari D Brooks; Jaydev P Desai
Journal:  Med Image Anal       Date:  2006-11-17       Impact factor: 8.545

4.  Performance of interventions with manipulator-driven real-time MR guidance: implementation and initial in vitro tests.

Authors:  Eftychios Christoforou; Erbil Akbudak; Alpay Ozcan; Menelaos Karanikolas; Nikolaos V Tsekos
Journal:  Magn Reson Imaging       Date:  2006-11-07       Impact factor: 2.546

5.  Radiofrequency ablation in patients with primary breast carcinoma: a pilot study in 26 patients.

Authors:  F Izzo; R Thomas; P Delrio; M Rinaldo; P Vallone; A DeChiara; G Botti; G D'Aiuto; P Cortino; S A Curley
Journal:  Cancer       Date:  2001-10-15       Impact factor: 6.860

6.  Radiofrequency ablation of small breast cancer followed by surgical resection.

Authors:  Masakuni Noguchi; Mitsuharu Earashi; Hisatake Fujii; Koichi Yokoyama; Ken-ichi Harada; Koichi Tsuneyama
Journal:  J Surg Oncol       Date:  2006-02-01       Impact factor: 3.454

7.  [MRI-guided percutaneous radiofrequency ablation of hepatic neoplasms--first technical and clinical experiences].

Authors:  P E Huppert; J Trübenbach; F Schick; P Pereira; C König; C D Claussen
Journal:  Rofo       Date:  2000-08

8.  Radiofrequency ablation of breast cancer.

Authors:  S Eva Singletary
Journal:  Am Surg       Date:  2003-01       Impact factor: 0.688

9.  MRI evaluation of the contralateral breast in women with recently diagnosed breast cancer.

Authors:  Constance D Lehman; Constantine Gatsonis; Christiane K Kuhl; R Edward Hendrick; Etta D Pisano; Lucy Hanna; Sue Peacock; Stanley F Smazal; Daniel D Maki; Thomas B Julian; Elizabeth R DePeri; David A Bluemke; Mitchell D Schnall
Journal:  N Engl J Med       Date:  2007-03-28       Impact factor: 91.245

Review 10.  Radiofrequency ablation for breast cancer: a review of the literature.

Authors:  I M C van der Ploeg; S van Esser; M A A J van den Bosch; W P Th M Mali; P J van Diest; I H M Borel Rinkes; R van Hillegersberg
Journal:  Eur J Surg Oncol       Date:  2007-03-21       Impact factor: 4.424

View more
  13 in total

1.  An MRI-Guided Telesurgery System Using a Fabry-Perot Interferometry Force Sensor and a Pneumatic Haptic Device.

Authors:  Hao Su; Weijian Shang; Gang Li; Niravkumar Patel; Gregory S Fischer
Journal:  Ann Biomed Eng       Date:  2017-04-26       Impact factor: 3.934

2.  Teleoperation System with Hybrid Pneumatic-Piezoelectric Actuation for MRI-Guided Needle Insertion with Haptic Feedback.

Authors:  Weijian Shang; Hao Su; Gang Li; Gregory S Fischer
Journal:  Rep U S       Date:  2013

3.  Design, Development, and Evaluation of a Master-Slave Surgical System for Breast Biopsy under Continuous MRI.

Authors:  Bo Yang; Steven Roys; U-Xuan Tan; Mathew Philip; Howard Richard; Rao Gullapalli; Jaydev P Desai
Journal:  Int J Rob Res       Date:  2014-04       Impact factor: 4.703

4.  Design and Control of a 1-DOF MRI Compatible Pneumatically Actuated Robot with Long Transmission Lines.

Authors:  Bo Yang; U-Xuan Tan; Alan McMillan; Rao Gullapalli; Jaydev P Desai
Journal:  IEEE ASME Trans Mechatron       Date:  2011-12-01       Impact factor: 5.303

5.  Tri-Axial MRI Compatible Fiber-optic Force Sensor.

Authors:  U-Xuan Tan; Bo Yang; Rao Gullapalli; Jaydev P Desai
Journal:  IEEE Trans Robot       Date:  2011-02-01       Impact factor: 5.567

6.  Design, development, and evaluation of an MRI-guided SMA spring-actuated neurosurgical robot.

Authors:  Mingyen Ho; Yeongjin Kim; Shing Shin Cheng; Rao Gullapalli; Jaydev P Desai
Journal:  Int J Rob Res       Date:  2015-05-18       Impact factor: 4.703

7.  Fiber Optic Force Sensors for MRI-Guided Interventions and Rehabilitation: A Review.

Authors:  Hao Su; Iulian I Iordachita; Junichi Tokuda; Nobuhiko Hata; Xuan Liu; Reza Seifabadi; Sheng Xu; Bradford Wood; Gregory S Fischer
Journal:  IEEE Sens J       Date:  2017-01-17       Impact factor: 3.301

8.  Robotic system for MRI-guided prostate biopsy: feasibility of teleoperated needle insertion and ex vivo phantom study.

Authors:  Reza Seifabadi; Sang-Eun Song; Axel Krieger; Nathan Bongjoon Cho; Junichi Tokuda; Gabor Fichtinger; Iulian Iordachita
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-06-23       Impact factor: 2.924

9.  MR-Compatible Haptic Display of Membrane Puncture in Robot-Assisted Needle Procedures.

Authors:  Amy Kyungwon Han; Jung Hwa Bae; Katerina C Gregoriou; Christopher J Ploch; Roger E Goldman; Gary H Glover; Bruce L Daniel; Mark R Cutkosky
Journal:  IEEE Trans Haptics       Date:  2018-03-19       Impact factor: 2.487

10.  Reaction force/torque sensing in a master-slave robot system without mechanical sensors.

Authors:  Tao Liu; Chunguang Li; Yoshio Inoue; Kyoko Shibata
Journal:  Sensors (Basel)       Date:  2010-07-29       Impact factor: 3.576

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.