Bruno P Soares1, Jan Willem Dankbaar, Joerg Bredno, SuChun Cheng, Sumail Bhogal, William P Dillon, Max Wintermark.
Abstract
INTRODUCTION: The purpose of this study is to compare the variability of PCT results obtained by automatic selection of the arterial input function (AIF), venous output function (VOF) and symmetry axis versus manual selection.Entities:
Mesh:
Year: 2009 PMID: 19274457 PMCID: PMC2694925 DOI: 10.1007/s00234-009-0516-9
Source DB: PubMed Journal: Neuroradiology ISSN: 0028-3940 Impact factor: 2.804
Fig. 1Axial PCT map depicting the area of infarct core (red) and tissue at risk (green) in the right MCA territory. Also demonstrated are the regions of interest (ROIs) that were systematically drawn for all post-processing rounds in the MCA territory of the ischemic (yellow) and contralateral hemisphere (blue). Mean values of CBV, CBF and MTT measured in these ROIs were recorded
Interobserver variability for the manual and automated post-processing methods for the volumes of the infarct core and of the total perfusion defect.
| Interobserver variability | |||
|---|---|---|---|
| Manual post-processing | Automated post-processing | ||
| Volume of infarct core (mL) | |||
| Mean ± SD | 16.89 ± 21.01 | 17.42 ± 20.84 | 0.739 |
| Repeatability coefficient | 5.23 | 2.73 | |
| Coefficient of variation | 31.0% | 15.7% | |
| Slope | 1.069 | 0.987 | |
| Intercept | −1.133 | 0.631 | |
| Correlation | 0.992 | 0.997 | |
| Volume of total perfusion defect (mL) | |||
| Mean ± SD | 68.11 ± 49.60 | 68.36 ± 49.87 | 0.998 |
| Repeatability coefficient | 8.33 | 3.99 | |
| Coefficient of variation | 12.2% | 5.8% | |
| Slope | 1.015 | 0.995 | |
| Intercept | −2.239 | 0.698 | |
| Correlation | 0.995 | 0.999 | |
Interobserver variability for the manual and automated post-processing methods for PCT values of cerebral blood volume (CBV), cerebral blood flow (CBF), mean transit time (MTT) and blood–brain barrier permeability (BBBP) in the ischemic brain parenchyma.
| ROI on ischemic hemisphere | Interobserver variability | ||
|---|---|---|---|
| Manual post-processing | Automated post-processing | ||
| CBV (mL/100 g) | |||
| Mean ± SD | 3.00 ± 0.95 | 2.89 ± 0.93 | 0.650 |
| Repeatability coefficient | 0.71 | 0.32 | |
| Coefficient of variation | 23.7% | 11.3% | |
| Slope | 0.724 | 0.849 | |
| Intercept | 0.766 | 0.376 | |
| Correlation | 0.921 | 0.930 | |
| CBF (mL/100 g/s) | |||
| Mean ± SD | 18.94 ± 11.45 | 17.67 ± 9.84 | 0.808 |
| Repeatability coefficient | 6.21 | 1.71 | |
| Coefficient of variation | 32.8% | 9.7% | |
| Slope | 0.960 | 1.004 | |
| Intercept | 1.030 | −0.070 | |
| Correlation | 0.948 | 0.994 | |
| MTT (s) | |||
| Mean ± SD | 15.23 ± 21.02 | 15.67 ± 21.93 | 0.952 |
| Repeatability coefficient | 2.55 | 1.49 | |
| Coefficient of variation | 16.7% | 9.5% | |
| Slope | 0.966 | 1.035 | |
| Intercept | −0.300 | 0.006 | |
| Correlation | 0.997 | 0.995 | |
| BBBP (mL/100 g/s) | |||
| Mean ± SD | 1.10 ± 0.61 | 0.98 ± 0.61 | 0.514 |
| Repeatability coefficient | 0.40 | 0.33 | |
| Coefficient of variation | 35.9% | 34.2% | |
| Slope | 1.180 | 0.812 | |
| Intercept | −0.184 | 0.166 | |
| Correlation | 0.945 | 0.953 | |
Interobserver variability for the manual and automated post-processing methods for PCT values of cerebral blood volume (CBV), cerebral blood flow (CBF), mean transit time (MTT) and blood–brain barrier permeability (BBBP) in the non-ischemic brain parenchyma.
| ROI on contralateral, non-ischemic hemisphere | Interobserver variability | ||
|---|---|---|---|
| Manual post-processing | Automated post-processing | ||
| CBV (mL/100 g) | |||
| Mean ± SD | 3.04 ± 0.67 | 2.94 ± 0.50 | 0.117 |
| Repeatability coefficient | 0.74 | 0.29 | |
| Coefficient of variation | 24.3% | 9.7% | |
| Slope | 0.734 | 0.921 | |
| Intercept | 0.696 | 0.238 | |
| Correlation | 0.805 | 0.942 | |
| CBF (mL/100 g/s) | |||
| Mean ± SD | 34.40 ± 10.86 | 32.17 ± 8.31 | 0.082 |
| Repeatability coefficient | 12.19 | 4.69 | |
| Coefficient of variation | 35.4% | 14.6% | |
| Slope | 0.811 | 0.951 | |
| Intercept | 6.863 | 1.632 | |
| Correlation | 0.779 | 0.942 | |
| MTT (s) | |||
| Mean ± SD | 5.59 ± 1.23 | 5.70 ± 1.16 | 0.307 |
| Repeatability coefficient | 1.67 | 0.45 | |
| Coefficient of variation | 29.8% | 8.0% | |
| Slope | 0.790 | 0.877 | |
| Intercept | 0.985 | 0.680 | |
| Correlation | 0.696 | 0.979 | |
| BBBP (mL/100 g/s) | |||
| Mean ± SD | 1.10 ± 0.38 | 0.96 ± 0.38 | 0.424 |
| Repeatability coefficient | 0.40 | 0.25 | |
| Coefficient of variation | 36.7% | 25.6% | |
| Slope | 0.923 | 0.927 | |
| Intercept | 0.092 | 0.083 | |
| Correlation | 0.811 | 0.929 | |